Pitfalls in 2HG Detection With TE‐Optimized MRS at 3T Authors: Seyma Alcicek, Dunja Simicic, Lindsay Blair, Max Saint‐Germain, Helge J. Zöllner, Christopher W. Davies‐Jenkins, Matthias Holdhoff, John Laterra, Chetan Bettegowda, Karisa C. Schreck, Doris D. Lin, Peter B. Barker, David O. Kamson, Georg Oeltzschner Published: 2026 ABSTRACT ABSTRACTBackground and PurposeIn vivo MRS of 2‐hydroxyglutarate (2HG) may provide diagnostic and monitoring biomarkers in isocitrate dehydrogenase (IDH)‐mutated glioma. A previous meta‐analysis has shown good diagnostic accuracy of TE‐optimized PRESS for IDH‐mutated glioma, but most studies feature IDH‐wildtype glioma as a comparison. However, when considering newly identified brain lesions that may mimic glioma, full characterization of its diagnostic utility should also consider the accuracy of 2HG measurement in nontumor tissue. Therefore, we tested how well TE‐optimized 2HG levels distinguish between IDH‐mutated glioma and nontumor tissue—here, normal‐appearing brain. We further examined the impact of different spectral modeling strategies (baseline stiffness, macromolecule inclusion, and basis set composition).MethodsForty‐eight patients with diagnosed/suspected IDH‐mutated glioma were enrolled. 3T MRS data were acquired from tumor and contralateral nontumor tissue with TE‐optimized PRESS and analyzed with “LCModel” software. ROC analysis evaluated 2HG estimates’ ability to distinguish IDH‐mutated glioma from nontumor tissue. Modeling interactions between 2HG and other metabolites were evaluated to identify reasons for potential false‐positive 2HG detection.ResultsTE‐optimized PRESS distinguished IDH‐mutated glioma from nontumor tissue with lower sensitivity (range 0.76–0.62) and specificity (0.85–0.78) than the literature suggests for IDH‐mutated versus IDH‐wildtype glioma. Strong negative correlations between gamma‐aminobutyric acid (GABA) and 2HG persisted across all modeling strategies and may lead to false‐positive 2HG detection in nontumor tissue. We further present a cautionary example from a patient on a ketogenic diet, showing that ketone body acetone can interfere with 2HG detection.ConclusionsSpectral overlap with GABA and acetone can lead to false‐positive 2HG detection in nontumor tissue. Introduction Metabolic reprogramming is a hallmark of glioma. Key somatic mutations occur in the genes encoding isocitrate dehydrogenase 1 and 2 (IDH1/2) [1]. Gliomas with IDH mutations (IDHmut) display distinct biological behaviors compared to IDH wildtype (IDHwt). The presence of IDH mutations is associated with a more favorable outcome in glioma, including longer progression‐free and overall survival, and increased sensitivity to treatment [2]. IDH mutation status has therefore become one of the most critical factors in classifying gliomas [3]. One of the key molecular alterations in IDHmut glioma is the production of the oncometabolite D‐2‐hydroxyglutarate (2HG) instead of α‐ketoglutarate. In healthy brain tissue, 2HG is an error product of normal metabolism and therefore only present at negligible levels [4, 5]. However, in IDHmut gliomas, 2HG can accumulate at millimolar levels. At these concentrations, 2HG profoundly influences tumor biology by modulating cell death, epigenome, and metabolic processes [6]. Consequently, targeting this neo‐enzymatic activity with IDH inhibitors has emerged as a novel therapeutic class, with IDH1/2 inhibitor vorasidenib gaining FDA approval in 2024 for the treatment of gliomas [7, 8]. At pathological levels, 2HG can be measured noninvasively with in vivo 1H magnetic resonance spectroscopy (MRS) [9]. MRS‐derived estimates of 2HG may serve as a useful biomarker for diagnosis and treatment monitoring in IDHmut glioma [9, 10, 11, 12, 13]. MRS may be especially valuable in cases where a biopsy is considered too risky or not feasible. 2HG maps derived from spectroscopic imaging may also contribute to the assessment of tumor infiltration beyond contrast‐enhancing margins for radiotherapy planning [14]. The key challenge in unequivocally measuring 2HG is the poor spectral dispersion at clinical field strength (i.e., 3T), which causes its three J‐coupled multiplet signals (1.9, 2.25, and 4.02 ppm) to overlap with signals from other metabolites, including glutamate (Glu), glutamine (Gln), and gamma‐aminobutyric acid (GABA). The most commonly used approach is an optimized timing of the point‐resolved spectroscopy (PRESS) localization sequence (echo time TE/TE1/TE2 = 97/32/65 ms), originally proposed by Choi et al. [9]. This setup is designed to achieve a “pseudo‐singlet” appearance of the key resonances of Glu and 2HG, which reduces overlap in the 2.25 ppm region. Advanced MRS techniques (e.g., edited MRS [15] and 2D correlation MRS [11]) have also been proposed for 2HG detection, but their lack of commercial availability, higher technical demands, vulnerability to measurement instabilities, and longer acquisition times have positioned conventional MRS sequences with optimized TE spacing as more viable for clinical translation [16], although they are currently applied only in clinical research [13]. A recent meta‐analysis concluded that TE‐optimized PRESS has high diagnostic specificity (97%) and sensitivity (92%) for IDHmut glioma. However, most studies that have assessed diagnostic performance to date have done so featuring IDHwt glioma as a comparison [13]. 2HG detection should ideally facilitate diagnosis of IDHmut glioma while excluding other (non‐IDHwt) conditions, that is, nonglial primary brain tumors, inflammatory diseases, infections, infarction, and brain metastases [16]. Notably, brain tumors exhibit very different spectra than normal brain and other brain diseases. Specifically, they typically contain lower levels of the key metabolites that overlap with 2HG, like Glu and GABA [17], which is likely to make unequivocal 2HG identification easier. To fully characterize the diagnostic accuracy of TE‐optimized PRESS, one must also consider its performance in nontumor tissue, particularly in the context of assessing tumor infiltration. Furthermore, although all studies included in the abovementioned meta‐analysis employed the widely used LCModel software for spectral analysis, they vary considerably in the details of the modeling strategies. It is established that important modeling decisions (such as baseline modeling and basis‐set composition) influence metabolite estimation [18, 19], but this phenomenon has not been thoroughly investigated in the context of 2HG detection. In this study, potential causes of false‐positive 2HG detections in nontumor tissue using TE‐optimized PRESS were investigated. The performance of TE‐optimized 2HG detection in differentiating IDHmut glioma from viable normal‐appearing brain tissue from the contralateral hemisphere was systematically evaluated, and the results showed that modeling interactions between GABA and 2HG signals decrease diagnostic performance against normal‐appearing tissue. Further, the effect of different spectral modeling strategies on performance was investigated, focusing on the impact of baseline modeling stiffness, macromolecule inclusion, and basis set composition. Finally, a cautionary example is presented that demonstrates that the presence of the ketone body acetone can interfere with 2HG detection. Methods Study Design The study protocol was approved by the local institutional review board (IRB protocol NA_00036313), and written informed consent was obtained from all subjects. Subjects with IDHmut glioma diagnosis or suspected low‐grade glioma at the time of the MRS scan, sufficient tumor volume (>8 mL) estimated from 3D T2‐weighted FLAIR images, and age >18 years were recruited. Datasets were only included in the subsequent analysis if IDHmut diagnosis was confirmed following biopsy/surgery. IDH mutation status was determined by immunostaining (Anti‐IDH1 R132H antibody) or next‐generation sequencing (Johns Hopkins Hospital NGS Solid tumor panel) [20]. Magnetic Resonance Acquisition Patients were scanned using an MRS protocol on a 3T Philips Ingenia Elition scanner (Philips Healthcare, the Netherlands) with a 32‐channel head coil. The protocol included 3D FLAIR, 3D T1‐weighted MPRAGE, a B 0 map for shimming up to second order [21], and 1H single‐voxel spectroscopy. In cases of technical failure of the shimming tool [21], vendor‐provided automatic shimming (second order, based on the FASTMAP approach) was used as an alternative. MR spectra were acquired using the vendor‐provided product PRESS sequence at two different TEs, 30 ms (128 transients) and 97 ms (256 transients), with repetition time TR = 2 s and the same B 0 shimming parameters. Echo spacing in PRESS at 97 ms was optimized to detect the 2HG signal at ∼2.25 ppm (TE1/TE2 = 32/65 ms) [9]. The spectroscopic voxel was placed in the T2‐hyperintense tumor area defined on fluid‐attenuated inversion recover (FLAIR) images. Voxel size was adjusted according to tumor size to reduce partial volume effects (8–27 mL). In the presence of a large tumor, data were acquired from multiple locations within the lesion in several participants. Control spectra were obtained from anatomically similar contralateral, normal‐appearing brain regions. Water‐unsuppressed spectra were acquired from the same volumes of interest at the same TR and TE (1 transient) for water‐referenced metabolite quantification and eddy‐current correction. Spectral Analysis MRS data were analyzed using the open‐source toolbox Osprey (v. 2.6.3; https://github.com/schorschinho/osprey/) [22], using the integrated LCModel binary for fitting [23]. Data had already been coil‐combined, averaged, and eddy‐current‐corrected on the scanner. Basis sets were simulated with MRSCloud [24], assuming ideal excitation, with sequence timings and refocusing RF pulse waveforms used in the actual sequence and effects of spatial localization considered (30 × 30 × 30 mm3 volume, 41 × 41 points). The basis set included 2HG, alanine (Ala), ascorbate (Asc), aspartate (Asp), citrate (Cit), creatine (Cr), negative creatine methylene (‐CrCH2), cystathionine (Cystat), ethanolamine (EA), GABA, glycerophosphocholine (GPC), glutathione (GSH), glucose (Glc), Gln, Glu, glycine (Gly), myo‐inositol (mI), lactate (Lac), N‐acetylaspartate (NAA), N‐acetylaspartylglutamate (NAAG), phosphocholine (PCh), phosphocreatine (PCr), phosphoethanolamine (PE), scyllo‐inositol (sI), serine (Ser), and taurine (Tau) [25]. If participants were following a ketogenic diet (see Section 4.4), acetoacetate (AcAc), acetone (Ace), and β‐hydroxybutyrate (β‐OHB) were included. The spectral fitting range was 0.5–4.1 ppm, including all 2HG resonances. Full width at half maximum (FWHM) estimates for the water peak were calculated by Osprey and used for spectral quality assessment; spectra with FWHM >12 Hz were excluded [26]. Modeling Strategies Spectral modeling was repeated with seven baseline stiffness settings (varying the LCModel spline knot spacing parameter “DKNTMN” ‐ minimum distance of knots ‐ from 0.15 to 5). Modeling was further performed with and without default LCModel‐simulated macromolecules and lipids (MM09, MM12, MM14, MM17, MM20, Lip09, Lip13a, Lip13b, Lip20). To evaluate the goodness of fit in the target spectral region (2.0–2.4 ppm), the fit quality number (FQN, recommended by expert consensus) was calculated by dividing the fit residual variance by the pure measurement noise variance [27]. We hypothesized that false‐positive 2HG modeling would lead to subtle modeling imperfections, which would result in higher fit residual and, consequently, be identifiable through elevated FQN. Metabolite Quantification Metabolite measurements were quantified relative to the internal tissue water signal. Water T2 relaxation correction was performed using T2 estimates derived by fitting a linear model to the logarithm of the water signal intensities against the two available TEs, 30 and 97 ms [28]. SPM12 [29] was used to perform tissue segmentation to yield tissue volume fractions of gray matter (GM), white matter (WM), and cerebrospinal fluid (CSF) in normal‐appearing brain, and water concentration was calculated using tissue‐specific water concentrations from the literature [30]. For the tumor tissue, we assumed an equal composition of gray and white matter and used a water concentration value of 42.3 M, as proposed by Choi et al. [9]. To evaluate the performance of dual‐TE T2 estimation, metabolite estimates in normal‐appearing brain tissue were recalculated using tissue‐specific water T2 given in the literature [30, 31]. Further, the magnetic field inhomogeneity within the voxel (ΔB 0) was estimated based on the difference between the T2 estimate and T2 * derived from FWHM. Statistical Analysis 2HG estimates in IDHmut tumor and contralateral normal‐appearing brain were compared using nonparametric Wilcoxon rank‐sum testing, following the Shapiro–Wilk test confirming nonnormality. Detection of 2HG in nontumor spectra obtained from normal‐appearing brain was considered a false positive, whereas nondetection was considered a true negative. To assess the impact of spectral linewidth on the accuracy of 2HG MRS detection, FWHM and ΔB 0 estimates calculated from contralateral datasets with false‐positive and true‐negative 2HG findings were compared using a Wilcoxon rank‐sum test, following confirmation of nonnormal distribution using the Shapiro–Wilk test. Two outcome measures were used to investigate the relationship between estimates of 2HG and other metabolites. First, LCModel coefficients of modeling covariance (CMCs) were extracted, a measure of interaction between two model parameters within a single dataset, derived from the Fisher information matrix. Second, correlations between 2HG and other metabolite estimates across contralateral MRS datasets were assessed by Spearman analysis, following Shapiro–Wilk testing confirming nonnormality. Receiver operating characteristic analysis was performed to evaluate 2HG estimates’ ability to distinguish IDHmut glioma and nontumor brain tissue. The optimal decision threshold for 2HG was calculated using the Youden index, which provides the threshold that maximizes the difference between the true positive rate (sensitivity) and the false positive rate (1 − specificity). To investigate the association between fitting goodness and false‐positive 2HG detection in control spectra, FQN and 2HG estimates were correlated using the Spearman correlation test. Furthermore, an additional ROC analysis was performed to assess the performance of FQN in distinguishing false‐positive 2HG detection. All analyses were performed using R v4.4.1 in RStudio v2024.04.2. Results were considered significant at p < 0.05. To validate the dual‐TE quantification approach, Spearman correlation analysis was performed between metabolite estimates in normal‐appearing brain tissue calculated using dual‐TE water T2 estimates and tissue‐specific literature‐based water T2 relaxation times. Water T2 estimates of IDHmut tumor and normal‐appearing contralateral tissue in MRS voxels were compared using the Wilcoxon rank‐sum test, following confirmation of nonnormality. Results Patient Characteristics Forty‐eight glioma patients were prospectively enrolled and underwent the MRI/MRS examination. Six patients with undefined histological status and two patients diagnosed with IDHwt glioblastoma were subsequently excluded. Twenty‐four of the remaining 40 patients were diagnosed with IDHmut astrocytoma and 16 with IDHmut oligodendroglioma. Thirty‐seven patients underwent partial resection and had measurable residual disease before the MRS examination. Thirteen patients had a history of radiotherapy and chemotherapy (no IDH inhibitor). The median interval between treatment and MRS acquisition ranged from 2.9 to 3.8 years depending on treatment modality. Patient characteristics are listed in Table 1. TABLE 1Patient characteristics. IDH = isocitrate dehydrogenase.CharacteristicsAll IDHmut patients, n = 40GeneralAge, years, median (range)40 (18–79)Female12HistologyOligodendroglioma, IDH‐mutant, 1p/19q‐codeleted, grade 212Oligodendroglioma, IDH‐mutant, 1p/19q‐codeleted, grade 34Astrocytoma, IDH‐mutant, grade 217Astrocytoma, IDH‐mutant, grade 37IDH mutation statusIDH1 R132H37IDH1 R132C2IDH2 R172K1Prior treatment (n; time to MRS acquisition, years, median [IQR])Resection37; 2.9 (0.5–8.0)Radiotherapy13; 3.8 (2.6–7.1)Chemotherapy13; 3.2 (1.5–6.5)John Wiley & Sons, Ltd. Patient characteristics. IDH = isocitrate dehydrogenase. In 28 cases, control spectra were obtained from the contralateral normal‐appearing brain as described above (Figure 1A,B). In six IDHmut cases, data were acquired from multiple locations (2–4) within the lesion. Based on data exclusion criteria (FWHM >12 Hz), one tumor spectrum and one control spectrum were excluded, leaving 47 IDHmut tumor spectra (mean FWHM = 5.01 ± 1.58 Hz) and 27 contralateral spectra (6.79 ± 1.49 Hz) for further analysis (Figure 2A). Example voxel locations in tumor (A) and contralateral hemisphere (B) in one subject overlaid on FLAIR MRI. In vivo MR spectra acquired from (C) IDHmut glioma (n = 46) and (D) contralateral brain (n = 26) using TE‐optimized PRESS at 3T for 2HG detection. One additional IDHmut tumor and contralateral spectra are shown in Figure 5 and are not included here. Mean spectra are displayed in black. The spectral regions with major peaks of glutamate (Glu) and 2HG are magnified and highlighted with green and gray boxes. Boxplots of (E) 2HG estimate and (F) spectral fitting uncertainty values (CRLB) obtained from LCModel fitting of IDHmut tumor and contralateral spectra. Thirteen data points (seven IDHmut, six contralateral) with CRLB >100% are not shown. (A) Boxplots of FWHM estimates of the water peak, calculated by Osprey, from spectra acquired in IDHmut tumors and contralateral brain regions. (B) Contralateral datasets were stratified into two groups based on 2HG detection: true‐negative (2HG = 0, 2HG−) and false‐positive (2HG > 0, 2HG+). FWHM estimates of the water peak were compared between these groups to evaluate the potential influence of spectral linewidth on false‐positive 2HG detection. 2HG Estimates in IDHmut Tumor and Normal‐Appearing Brain Tissue Characteristic patterns in the 2.1–2.5 ppm region highlight lower Glu and higher 2HG signals in IDHmut spectra than contralateral spectra (Figure 1C,D). Despite this clear visual distinction, there was no obvious 2HG cutoff that would perfectly separate the tumor and contralateral groups (Figure 1E), although median 2HG estimates were significantly higher in tumor (median 2HGtumor = 0.85 i.u. and median 2HGcontrol = 0.47 i.u., p = 0.001). As shown in Figure 1F, 2HG was detected in seven control spectra with high spectral fitting certainty (i.e., Cramér–Rao lower bound [CRLB] <25%). The correlation between 2HG estimates in the contralateral voxel calculated using dual‐TE T2 estimates and those calculated using tissue‐specific water T2 relaxation times (segmentation‐based) was nearly perfect (ρ = 0.99). Water T2 estimates in IDHmut tumor voxels (114.0 ± 32.8 ms) were significantly higher than those in normal‐appearing contralateral voxels (67.9 ± 3.61 ms) (p < 10−16), consistent with literature [32]. This confirmed that it is justified to use 2HG estimates in both tumor and contralateral spectra, calculated using dual‐TE T2 estimates. Tumor spectra exhibited significantly lower water peak FWHM estimates compared to contralateral spectra (p = 7.01 × 10−6), whereas calculated ΔB 0 values did not differ between groups (p = 0.91), suggesting that the broader water peaks in the contralateral region are primarily attributable to their relatively shorter water T2 relaxation times compared to the tumor region. Within contralateral nontumor regions, neither water peak FWHM nor ΔB 0 estimates differed significantly between false‐positive and true‐negative 2HG detection groups (p = 0.10 and p = 0.06, respectively). Interactions Between 2HG and Other Metabolites’ Spectral Models Strong spectral overlap of 2HG and GABA between 2.1 and 2.5 ppm (Figure 3A) resulted in large negative modelling covariance between the two metabolites in both tumor and nontumor spectra (median CMCtumor = −0.44 and median CMCcontrol = −0.49), suggesting it remains difficult for the modeling process to separate them reliably (Figure 3B). 2HG estimates interacted similarly, but more weakly, with tCr, Glu, tNAA, Glx, NAAG, MM20, PCr, and tCho. Interactions between MM20 (macromolecules at 1.95, 2.0, 2.25, and 3.0 ppm) and 2HG were more prominent in IDHmut tumor spectra than in contralateral spectra (median CMC: −0.12 and −0.03, respectively). (A) Spectral simulations of metabolites that have resonances in the same spectral region as 2HG (2.1–2.5 ppm). The spectral region with major peaks of 2HG is highlighted with a gray box. (B) Coefficients of modeling covariance (CMCs) between 2HG and other small metabolites and macromolecules, calculated by LCModel for IDHmut tumor and contralateral control spectra. Boxplots display the interquartile range (IQR) with whiskers extending 1.5 × IQR, along with median lines and outlier points. The boxplots are ordered by absolute median CMC values. The largest (negative) CMC values were observed with GABA for both spectrum types, which also exhibits the greatest spectral overlap with 2HG. The strong negative association between 2HG and GABA was again present in the correlation of their amplitude estimates across all contralateral datasets (p < 10−5, ρ = −0.8) (Figure 4). There was no significant correlation between 2HG and other metabolite estimates in the contralateral data. Correlation between estimates of 2HG concentration and other metabolites (which have high coefficients of modeling covariance (CMC); see Figure 2) in contralateral spectra. Presumed false‐positive detection of 2HG in normal‐appearing brain tissue is inversely correlated with GABA levels (p < 0.00001). Influence of Baseline and Macromolecule Modeling on 2HG Estimates The negative 2HG and GABA association persisted strongly across different modeling strategies, independent of the choice of baseline stiffness and inclusion of macromolecular profiles in the basis set (p < 10−4 to 10−7, ρ = −0.7 to −0.83) (Figure 5), although mean 2HG levels decreased considerably with increasing baseline flexibility across all datasets (mean 2HGtumor = 1.11–1.02 i.u.; mean 2HGcontrol = 0.51–0.33 i.u.), suggesting that flexible baseline models absorb signal from 2HG. Although MRS estimates of GABA are assumed to be higher in GM than in WM [33], no association was observed between contralateral brain tissue composition (i.e., GM content) and 2HG and GABA estimates. Correlation between GABA and 2HG estimates across various baseline stiffness (defined with DKNTMN ‐ minimum distance of knots ‐ control parameter) and with/without inclusion of macromolecule (MM) signals in LCModel spectral fitting. DKNTMN values correspond to the minimum spacing between spline knots. The inverse correlation between GABA and 2HG detection remained highly significant across all analyses (p < 0.001). The blue‐to‐gray gradient encodes DKNTMN. The area under the ROC curve (AUC), a measure of diagnostic accuracy, was highest (0.78) for 2HG estimates obtained with flexible baseline modeling (DKNTMN = 0.15) (Figure 6). The optimum cutoff 2HG level for this case was 0.44 i.u., with a sensitivity of 0.79 and specificity of 0.81. The range of AUCs across all modeling strategies was 0.78–0.69, with an optimum threshold for 2HG estimates between 0.42 and 0.72 i.u., sensitivity of 0.76–0.62, and specificity of 0.85–0.78. Receiver operating characteristic curves showing the performance of 2HG estimates in distinguishing IDH‐mutant glioma from nontumor tissue for various spectral modeling approaches (i.e., various baseline stiffness defined using the DKNTMN control parameter and with/without the inclusion of macromolecule, MM, signals in spectral fitting). Each plot includes the area under the curve (AUC), along with sensitivity and specificity values for the optimum threshold (Opt. Thresh.) for 2HG estimates. When the stiffest baseline modeling approach was applied, excluding macromolecule signal models (DKNTMN = 5, no MM), FQN for the 2–2.4 ppm region (targeting the 2HG peak at ∼2.25 ppm) showed good performance (AUC = 0.71) in identifying false‐positive 2HG detections in control spectra. The optimal cutoff for FQN was <1.4, yielding a sensitivity of 1 and a specificity of 0.57. False‐Positive 2HG Detection in a Participant on a Ketogenic Diet Finally, a strong overlap between 2HG and the acetone singlet at 2.22 ppm can cause false‐positive 2HG detection (Figure 7). Unusually high 2HG levels were quantified in both tumor and contralateral spectra of one IDHmut glioma patient who was on a ketogenic diet. Therefore, the dataset was re‐analyzed with the ketone bodies, acetone, acetoacetate, and β‐hydroxybutyrate included in the basis set. After inclusion, acetone and β‐hydroxybutyrate were fitted in both spectra, while 2HG was estimated to be substantially lower in the tumor and not at all in the contralateral. The high CMC (−0.69) between 2HG and acetone confirmed strong modeling interaction between these two compounds. Example of false‐positive high‐level 2HG detection in both the tumor and normal‐appearing contralateral brain tissue of an IDHmut glioma patient following a ketogenic diet. (A) Simulated spectra of acetone (Ace), acetoacetate (AcAc), β‐hydroxybutyrate (β‐OHB), and 2HG. Fitting of (B) IDHmut tumor and (C) contralateral spectra with/without ketone bodies (KBs), strongly influencing 2HG estimates. Spectral data are presented in black and fitting models in dark blue. Individual fitting models are shown for 2HG and ketone bodies. Discussion In vivo MRS is one of the most important candidates for clinical application, and it is important to point out specific potential failure modes of the method. TE‐Optimized MRS Can Give False‐Positive 2HG Findings in Nontumor Tissue The current study presents evidence that positive 2HG findings on single‐voxel TE‐optimized (TE = 97 ms) MRS at 3T do not mean an unequivocal IDHmut glioma diagnosis, since a 2HG signal was quantified in normal brain with low uncertainty values in 7 of 27 control spectra (e.g., false‐positive 2HG). On the other hand, accurate 2HG quantification in tumor, peritumoral, and nontumor tissue is essential for reliably mapping tumor extent [34]. We further speculate that false‐positive 2HG detection might also be found in other lesion types, such as infection, inflammation, or others, depending on their spectral overlapping GABA signal levels. Understanding limitations and failure modes of 2HG MRS is especially critical for cases where biopsy is not feasible and MRS would be the only means of diagnosis. Our data show that 2HG estimates from TE‐optimized PRESS can be artifactually higher than expected in spectra from viable nontumor tissue with normal MRI appearance. The performance of TE‐optimized PRESS in distinguishing IDHmut glioma from contralateral normal‐appearing tissue was, therefore, noticeably lower compared to its ability to differentiate IDHmut from IDHwt glioma [13]. This discrepancy can be attributed to low GABA concentrations in both IDHwt glioma and IDHmut glioma [35], potentially causing the 2HG–GABA association to be overlooked [36]. 2HG detection performance has not been systematically investigated in pathologies that may mimic glioma. This literature gap may give clinicians a false sense of diagnostic certainty of 2HG MRS, particularly if those lesions have normal or elevated GABA levels. Furthermore, the risk of false‐positive 2HG detection likely increases if the measurement volume contains appreciable amounts of normal brain tissue, for example, with small and/or infiltrating lesions. GABA Strongly Interferes With 2HG Estimation Independent of Modeling Strategy The primary cause appears to be spectral overlap not entirely eliminated by TE optimization. CMCs, a measure of interaction between model parameters during spectral fitting, revealed a strong anticorrelation between GABA and 2HG, supporting similar findings for sLASER localization at 3T and 7T [37]. This increases the likelihood of the GABA signal being erroneously modeled as 2HG. The strong negative correlation between 2HG and GABA estimates in contralateral tissue gives further validation. While this study used data from normal‐appearing contralateral brain, our findings suggest that 2HG MRS in lesions with normal (or high) GABA levels may lead to false‐positive 2HG detection, which may, in turn, lead to false diagnoses if MRS were considered in isolation. Notably, 26% of contralateral spectra exhibited low 2HG fitting uncertainty (CRLB < 25%), although very little 2HG signal is expected in normal brain. The practice of using high relative CRLB values to identify low detection certainty and exclude data is commonly criticized for inducing bias [38], but the current data conversely show that low CRLBs alone do not guarantee accurate 2HG detection. Nevertheless, CRLB cutoff criteria remain popular, and previous 2HG studies have proposed cutoffs for reliable 2HG detection as high as 50% [15, 18], which would include most contralateral datasets in our study. Although spectral linewidth plays a crucial role in resolving overlapping resonances such as GABA and 2HG, our dataset, characterized by overall good spectral quality with the mean FWHM approximately half of the cutoff value defined by expert consensus for poor spectral quality [26], showed no association between false‐positive 2HG detection and either spectral linewidth or magnetic field inhomogeneity within the voxel. Furthermore, we demonstrated that the narrower linewidth observed in tumor spectra compared to nontumor spectra is a result of the longer T2 relaxation time of IDHmut tumor tissue, consistent with literature [32]. It was also demonstrated that the strong GABA–2HG interaction persists across many different modeling strategies. Different modeling strategies can affect the estimation of metabolite levels, particularly for J‐coupled signals like Glu, GABA, and 2HG. Indeed, changes in baseline modeling in our work influenced 2HG estimates and, consequently, the overall performance, including the optimal threshold for 2HG estimation, as well as diagnostic sensitivity and specificity. The GABA–2HG interaction remained unanimously strong, suggesting that the overlap problem cannot be solved at the modeling level. Previous work demonstrated that the sensitivity of MRS for detecting 2HG depends on tumor volume, recommending 8 mL as a lower limit for reliable detection [16]. Conversely, a small brain lesion could also lead to partial volume effects, since normal‐appearing brain tissue included in the MRS voxel could increase GABA levels in the measurement volume and facilitate false‐positive 2HG detection, for example, in a non‐IDHmut glioma lesion, or lead to overestimation of 2HG concentration in IDHmut glioma. While this might plausibly lead to assuming that gray‐matter‐rich voxels are more likely to exhibit higher (falsely detected) 2HG levels due to their greater GABA content, we did not observe an association between false‐positive 2HG detection and the GM content of the contralateral volume of interest. In searching for a reliable indicator of false‐positive 2HG detection, a quantitative estimate of the modeling quality, that is, FQN specific to the 2HG region (2−2.4 ppm), was investigated. Indeed, the FQN was reasonably good in identifying false‐positive 2HG detection (AUC = 0.71), but only for the stiffest baseline approach without the inclusion of macromolecule models. Weaker Interference of Other Metabolites With 2HG Other signals interacting with 2HG include Glu, Gln, NAAG, and MM20. The interactions did not vary much between tumor and control spectra except for MM20, which includes resonances at 1.95, 2.08, 2.25, and 3 ppm and was more prominent in tumor spectra. Lipid and macromolecule resonances near 2.25 ppm have been reported to interfere with 2HG quantification, particularly in IDHwt tumors with high lipid contamination [39]. In our study, which consisted of IDHmut tumor and contralateral spectra, the observed correlations of 2HG with MM20 and Lip20 were low, indicating only a minor impact in this comparison. Although there is no direct overlap between 2HG and tCr, their high coefficient of modeling covariance is likely again mediated through GABA, since GABA overlaps with 2HG at 1.9 and 2.25 ppm, and with tCr at 3 ppm, causing indirect interaction. An alternative explanation for the observed high model correlation between 2HG and tCr may be the spectral proximity of the 2HG resonance at 4.05 ppm to the Cr peak at 3.9 ppm, although we did not observe a similar correlation with the “CrCH2” component modeled at 3.9 ppm in the basis set. Even though we observe low modeling covariance coefficients (i.e., low spectral overlap) between Glu and 2HG, the high Glu concentration may still interact with accurate 2HG estimation. These weaker modeling interactions did, however, not lead to negative correlations between 2HG and other metabolite estimates across all datasets in contralateral tissue. Strong Interference of Acetone With 2HG Warrants Caution in Patients on a Ketogenic Diet Finally, this study showed that the acetone singlet (2.22 ppm) can also be mistakenly modeled as 2HG. This is highly important, as nutritional interventions (fasting, ketogenic diets) are often explored as complementary treatment approaches in low‐grade glioma [40]. Ketone bodies (acetone, acetoacetate, β‐hydroxybutyrate) are typically undetectable via conventional MRS in healthy brain under a standard diet, and therefore are not commonly included in basis sets. However, in the context of nutritional interventions or specific dietary preferences, brain ketone concentrations can rise above the MRS detectability threshold, especially in brain tumors [41, 42]. Therefore, the basis set composition needs to be modified based on medical history. High covariance between 2HG and acetone models underscores the difficulty in reliably separating them. Including acetone in the basis set by default may lead to underfitting of the 2HG signal, as it may reasonably assume some of the real‐existing 2HG signal. Physiological specifics of the subject, such as dietary state, metabolic conditions, or underlying disease, need to be taken into account when selecting the basis set composition either manually [18] or with an automatic data‐driven approach [43]. In general, for study cohorts or clinical cases involving nonstandard diets, the TE‐optimized PRESS sequence may not be the ideal method for unequivocal 2HG quantification. Other 2HG MRS Detection Techniques J‐difference editing of 2HG may provide higher sensitivity and specificity than conventional protocols like the TE‐optimized technique used in this work [15]. However, spectral editing is more technically demanding, is vulnerable to experimental instabilities (motion or field drift), requires more nuanced postprocessing, and is less widely available on clinical MRI scanners. Conventional approaches are likely to remain relevant, making it important to understand their failure modes. sLASER localization (TE = 110 ms [44]) and triple‐refocusing [36] attempt to further reduce spectral overlap between 2HG and metabolites, including GABA, Glu, and Gln, but like TE optimization, they cannot remove it entirely, and likely experience the same problematic interaction with GABA. Two‐dimensional correlation MRS techniques have also been employed to detect 2HG signals [11]. However, like spectral editing, adoption in clinical practice is limited, owing to long acquisition times, demand for complex postprocessing, and limited availability on clinical scanners. Furthermore, 7T MRS provides higher sensitivity for 2HG detection compared to 3T due to increased spectral dispersion, although the GABA–2HG overlap is not entirely eliminated [37]. Due to the limited availability of 7T MR scanners in clinical settings, optimizing 2HG MRS detection at 3T remains relevant for clinical translation. Our study is limited by a small sample size (n = 40). However, it is the largest high‐quality TE‐optimized PRESS dataset to date to compare spectra from IDHmut gliomas and contralateral normal‐appearing brain. Prior treatment status is an important factor influencing metabolic and spectroscopic tumor characteristics. Due to the limited number of treatment‐naïve patients, we combined pre‐ and posttreatment gliomas in our analysis. Although the median interval between treatment and MRS acquisition exceeded 2.9 years for all treatment modalities, reducing the likelihood of acute treatment‐related effects, we cannot exclude persistent treatment‐associated alterations in tumor metabolism. We acknowledge that 2HG levels may decrease after treatment (including chemoradiation), which could lower detection sensitivity in posttreatment patients [45]. To reduce confounding effects such as partial volume, only lesions with sufficient tumor volume (>8 mL), estimated from 3D T2‐weighted FLAIR images, were included, ensuring reliable voxel placement and minimizing contamination from adjacent normal‐appearing tissue. Additionally, water T1 and metabolite T1 and T2 relaxation time corrections were not performed, as tumor‐specific relaxation times were unavailable. In particular, the lack of knowledge about 2HG relaxation times remains a concern that future studies ought to investigate to avoid biases in concentration estimates (and quantitative diagnostic cutoffs). Nevertheless, we assumed that the major variability in relaxation correction originates from water T2 relaxation and calculated a specific T2 estimate using a dual‐TE approach. Another limitation is that for the tumor tissue, we assumed an equal composition of gray and white matter and used a water concentration value of 42.3 M, as proposed by Choi et al. [9]. The actual water concentrations in the tumor might differ from this assumption, but subject‐specific measurement of water concentration requires time‐consuming additional measurements. In light of these limitations, we opted for the nomenclature “institutional units” instead of “mM.” It might be reasonable to separate basis functions for aspartyl and methyl signals of NAA and NAAG in medium‐to‐long TE spectra due to their different T2. This is not common practice [9], and additional modeling parameters may introduce variance and/or bias. The modeling covariance coefficients between NAA and 2HG appeared to be low (0.04 ± 0.06 in tumor, –0.01 ± 0.08 in contralateral), suggesting that the interaction between the two basis functions was weak enough to justify the conventional approach. TE‐optimized MRS at 3T can exhibit false‐positive 2HG detection in nontumor tissue. Signal overlap with GABA and acetone (if patients are on a ketogenic diet/fasting) is the primary reason. Spectral modeling strategies such as baseline handling, macromolecule modeling, and basis set composition can furthermore influence the diagnostic performance of TE‐optimized PRESS, although this influence is mild, and no modeling strategy was able to substantially eliminate the overlap. These findings should be interpreted in the context of the heterogeneous, largely posttreatment cohort investigated in this study. Overall, while TE‐optimized PRESS remains a robust approach for clinical 2HG detection, interpretation of 2HG estimates for determining IDH mutation status should be performed cautiously, considering potential spectral overlap with GABA and acetone as well as the patient's clinical and metabolic background. High methodological and technical rigor is necessary to ensure optimal quality of data acquisition and analysis. Funding This work was supported by National Institutes of Health grants (R01 EB035529, R21 EB033516, K99 AG080084, and P41 EB031771). This manuscript is the result of funding in whole or in part by the National Institutes of Health (NIH). It is subject to the NIH Public Access Policy. Through acceptance of this federal funding, NIH has been given a right to make this manuscript publicly available in PubMed Central upon the Official Date of Publication, as defined by NIH. S.A. was funded by the Mildred Scheel Career Center Frankfurt (Deutsche Krebshilfe). Conflicts of Interest C.B. is a consultant for Depuy‐Synthes, Bionaut Labs, Haystack Oncology, and Privo Technologies. C.B. is a co‐founder of OrisDx and Belay Diagnostics. G.O. is a consultant for Neurona Therapeutics. The authors declare no other conflicts of interest.