3D-printed patient-specific guides improve screw accuracy in upper cervical spine surgery: a comparative study Authors: Guoqi Niu, Yalong Guo, Gong Zhou, Yang Gao, Weili Jiang, Hui Chen, Hu Nie, Tao Liu, Zheng Sun, Jiawei Cheng, Yi Yang, Jianzhong Bai Published: 2026 ABSTRACT ObjectiveTo evaluate the feasibility and clinical efficacy of patient-specific 3D-printed guide plates for pedicle screw placement in the upper cervical spine, compared with the conventional free-hand technique.MethodsA retrospective analysis was conducted on 39 patients who underwent upper cervical spine surgery between June 2009 and June 2022. Patients were divided into a guide plate group (n = 24) and a free-hand control group (n = 15). Surgical parameters (operative time, blood loss, fluoroscopy frequency), screw placement accuracy (assessed by Kawaguchi classification on postoperative CT), and clinical outcomes (VAS, JOA, and ASIA scores) were compared between groups.ResultsA total of 141 pedicle screws were placed. The guide plate group achieved a significantly higher rate of perfect screw placement (81.8 vs. 62.3%, P = 0.009) and fewer grade 2 cortical breaches. Operative time, blood loss, and fluoroscopy exposure were significantly reduced in the guide plate group (all P < 0.05). Both groups showed significant postoperative improvements in all clinical scores (P < 0.05), with no significant intergroup differences at final follow-up.ConclusionPatient-specific 3D-printed guide plates significantly improve screw placement accuracy and reduce operative time, blood loss, and radiation exposure in upper cervical spine surgery, while achieving clinical outcomes comparable to the free-hand technique. This method is a safe and effective alternative for upper cervical pedicle screw fixation. Introduction Upper cervical spine pathologies, including developmental anomalies, trauma, infection, and tumors, frequently result in spinal instability that can severely compromise neurological function and even be life-threatening (1, 2). Early surgical stabilization is crucial to restore spinal alignment and prevent further injury (3, 4). Among various posterior fixation techniques (5), pedicle screw constructs have gained increasing acceptance for the management of upper cervical disorders due to their superior biomechanical stability, high fusion rates, and reliable three-column fixation (6, 7). However, the complex and variable anatomy of the upper cervical vertebrae, with narrow pedicles and close proximity to the vertebral artery and spinal cord, makes screw placement exceptionally challenging and risks serious neurovascular injury (8). Achieving accurate and safe pedicle screw insertion remains a central objective in upper cervical spine surgery (2, 9). Conventional free-hand screw placement relies heavily on anatomical landmarks and intraoperative fluoroscopy (10). This technique is associated with considerable limitations: it often yields suboptimal accuracy, necessitates repeated fluoroscopic verification, and prolongs operative time, thereby increasing blood loss, infection risk, and radiation exposure (8). These drawbacks are particularly pronounced in patients with deformity or aberrant anatomy, where visual and fluoroscopic guidance is less reliable. Consequently, the free-hand method is gradually being supplemented by more precise, technology-assisted approaches. Modern surgical aids such as robotic systems and computer-assisted navigation have been developed to improve screw placement precision (11). Nevertheless, their high cost, operational complexity, and steep learning curve restrict widespread clinical adoption. In contrast, advances in digital orthopedic modeling and additive manufacturing have made patient-specific three-dimensional (3D)-printed guide plates a promising alternative (12–14). Based on reverse-engineering principles, these individualized guides offer navigational-level accuracy while being cost-effective, easy to use, and capable of reducing operative duration, blood loss, and fluoroscopy frequency (15). Despite these advantages, reported clinical applications of 3D-printed guides in the upper cervical spine remain relatively limited. To further evaluate the feasibility and clinical value of this technique, the present study designed and utilized individualized 3D-printed guide plates to assist pedicle screw placement in patients with upper cervical spine disorders. The clinical and radiological outcomes were compared with those of conventional free-hand insertion. The present study aims to provide evidence supporting the broader clinical application of this innovative surgical technique. Materials and methods Study design and patient population This retrospective controlled study enrolled 39 patients who underwent upper cervical spine surgery for deformities (n = 19) or traumatic atlantoaxial fracture-dislocation (n = 20) between June 2009 and June 2022. Patient allocation was based on the clinical introduction date of the 3D-printed guide technique at our institution: patients treated between June 2009 and December 2014 received conventional free-hand screw placement (control group, n = 15), while patients treated from January 2015 to June 2022 received guide-assisted surgery (guide group, n = 24). This study was approved by the institutional ethics committee. Inclusion and exclusion criteria Inclusion criteria: (1) diagnosis of upper cervical deformity, fracture-dislocation, infection, or tumor requiring posterior fixation; (2) pedicle morphology suitable for screw insertion as assessed on preoperative computed tomography (CT); (3) physiological ability to tolerate surgery; (4) provision of informed consent for the use of patient-specific guides; and (5) availability of complete clinical data with a minimum one-year follow-up. Exclusion criteria: (1) anatomical contraindications, such as an atlantoaxial pedicle too narrow or an anomalous vertebral artery course precluding safe screw placement; (2) poor general medical condition posing a high surgical risk; and (3) psychiatric illness impairing compliance or unwillingness to participate. Guide plate design and fabrication 3D Anatomical model reconstruction Preoperative thin-slice cervical CT scans (slice thickness: 0.625 mm) were obtained for patients in the guide plate group. The DICOM data were imported into Mimics 17.0 software (Materialize, Belgium). The bony structures of C1 and C2 were segmented (Figures 1A–C) to generate a 3D surface model (Figures 1D–F), which was exported in STL format. (A-C) CT images of the atlantoaxial vertebrae in the coronal, axial, and sagittal planes; (D-F) 3D reconstruction model of the atlantoaxial joint. Virtual screw trajectory planning and guide design Within Mimics, a cylindrical model (diameter 3.5 mm) was used to simulate the pedicle screw. An experienced spine surgeon determined the optimal entry point and trajectory by visualizing the cylinder within the pedicle walls on multiplanar views (Figures 2A, B). A protective sleeve (inner diameter 2.0 mm) was designed along this trajectory. The guide plate was modeled as a semi-arched bridge structure with bilateral extensions to securely fit over the spinous process and contact the posterior elements (posterior arch, lamina, and partial facet joints) of C1 and C2. Using reverse engineering principles, a negative mold of the bone surface was created. Boolean operations merged this surface with the protective sleeve to form the integrated guide plate (Figures 2C–E). The final design was exported as an STL file. (A) Simulated screw trajectory from the lateral view after transparency rendering; (B) Simulated screw trajectory from the superior view after transparency rendering; (C) Superior view of the guide; (D) Posterior view of the guide; (E) Lateral view of the guide. 3D printing and preoperative validation The STL files were processed using IdeaMaker slicing software and printed on a fused deposition modeling (FDM) 3D printer with medical-grade polylactic acid (PLA) filament (layer thickness: 0.2 mm; infill density: 100%; Figures 3A–C). After printing, supports were removed and surfaces were smoothed. The fit and accuracy of the guide were verified on its corresponding 3D-printed anatomical model. (A) Superior view of the guide and model; (B) Posterior view of the guide and model; (C) Lateral view of the guide and model; (D) Lateral view of Kirschner wire insertion along the guide sleeve; (E) Superior view; (F) Lateral view of the model after simulated screw placement in vitro; (G) Posterior view; (H) Intraoperative fluoroscopy. Simulated screw placement In Vitro The anatomical model was securely fixed, and the sterilized guide plate was positioned on its posterior surface. A 2.0 mm K-wire was drilled through the guide sleeve to the pre-planned depth under fluoroscopic guidance (Figures 3D, E). Wire position and pedicle wall integrity were confirmed visually and radiographically. The tract was then tapped, and a 3.5 mm pedicle screw was inserted (Figures 3F–H). Following successful simulation, the guide plate was sterilized for surgical use. Surgical technique Guide plate-assisted surgery (guide group) Under general anesthesia, patients were positioned prone with the neck slightly flexed under skull traction (3 kg). A standard posterior midline incision exposed the occiput, C1 posterior arch, C2 lamina, and lateral masses. Soft tissue was meticulously cleared from the bone surfaces at the guide's contact points. The patient-specific guide was firmly seated onto the posterior elements of C1 and C2 (Figure 4A). A 2.0 mm K-wire was drilled through the guide sleeve to the predetermined depth, and its position was verified with C-arm fluoroscopy. After removing the guide and K-wire, the pedicle tract was palpated with a ball-tipped probe, tapped, and the appropriate pedicle screw was inserted (Figure 4B). Final screw positions were confirmed fluoroscopically (Figure 4C). Intraoperative use of the guide for screw placement. (A) Drilling after the guide is fitted to the bone surface during surgery; (B) Intraoperative screw placement; (C) Postoperative X-ray review. Free-hand surgery (control group) Surgery proceeded via the same posterior approach. The screw entry point and trajectory were determined based on preoperative imaging and anatomical landmarks. A hand drill was advanced incrementally along the planned path, with frequent palpation of the pedicle walls using a probe to ensure integrity. The tract was then tapped, and the screw was placed. Postoperative management Postoperative care was standardized for both groups. Patients received symptomatic management including steroids, analgesics, antibiotics, and gastrointestinal protectants. Early mobilization was encouraged: patients began log-rolling on postoperative day 1, progressed to assisted ambulation with a cervical collar by day 3, and underwent postoperative radiographic evaluation (X-ray and 3D-CT) on day 4 to assess screw placement accuracy. Outcome measures Primary outcome measures included: (1) Surgical parameters: operative time (specific to the pedicle screw insertion segment), intraoperative blood loss, and fluoroscopy frequency; (2) Radiological accuracy: assessed on postoperative 3D-CT scans by two independent spine surgeons who were not involved in the index surgeries and were blinded to group allocation. Inter-observer reliability was assessed using Cohen's kappa coefficient. Screw placement was graded according to the Kawaguchi classification (16), where Grade 0 (screw entirely within pedicle) and Grade 1 (cortical breach ≤ 2 mm without complications) were defined as clinically acceptable; Grade 2 (breach >2 mm) and Grade 3 (with neurovascular complication) were defined as unacceptable; (3) Clinical outcomes: evaluated using the JOA score (17), VAS, and ASIA scale (18) preoperatively, at 1 week postoperatively, and at the final follow-up. Fusion status was assessed at the final follow-up using CT (presence of continuous trabecular bone bridging across the atlantoaxial joint) and dynamic radiographs (< 3° of angular motion on flexion-extension views). Fusion was evaluated by two independent observers blinded to group allocation. Statistical analysis Statistical analysis was performed using SPSS 27.0 and R 4.2.0. Continuous data are presented as mean ± SD. Normality was assessed using the Shapiro-Wilk test and homogeneity of variances using Levene's test. Normally distributed data were analyzed using paired or independent t-tests; otherwise, Wilcoxon signed-rank or Mann-Whitney U tests were used. For repeated measures (JOA, VAS, and ASIA), linear mixed-effects models with patient-level random intercepts and fixed effects for time, group, and their interaction were applied, with pairwise comparisons adjusted by Bonferroni-Holm correction. Screw-level accuracy was analyzed using mixed-effects logistic regression with random intercepts per patient to account for within-patient clustering. Categorical data were compared using the Chi-square test. Post-hoc power analysis for the primary outcome was performed using a two-proportion z-test with conservative correction for clustering (ICC = 0.3). Two-sided P < 0.05 was considered significant. Results Baseline patient characteristics The two groups were comparable at baseline, with no statistically significant differences in demographic characteristics, diagnosis distribution, or preoperative functional scores (all P > 0.05). Detailed data are summarized in Table 1. Table 1Baseline characteristics and preoperative functional scores of patients.VariableControl (n = 15)Guide (n = 24)t/χ2P-valueAge (years, Mean±SD)40.07 ± 9.6847.33 ± 11.82−1.980.06Gender (male/female, n)8/715/90.310.58follow-up duration (months)56.2 ± 20.545.8 ± 18.31.820.074Diagnosis (n)Deformity8110.200.65Traumatic fracture/dislocation713JOA score9.07 ± 1.499.79 ± 1.98−1.220.23VAS score6.20 ± 1.276.38 ± 1.66−0.350.73ASIA sensory score198.67 ± 12.76201.16 ± 13.48−0.580.57ASIA motor score87.27 ± 8.4289.33 ± 10.63−0.640.53 Baseline characteristics and preoperative functional scores of patients. Surgical parameters All procedures were completed successfully by the same surgical team, with no intraoperative injuries to the vertebral artery, spinal cord, or nerve roots, and postoperative imaging confirmed satisfactory implant positioning without evidence of loosening, migration, or breakage. The guide plate group demonstrated significant advantages over the control group across all surgical parameters, including significantly shorter operative time (141.54 ± 50.09 min vs. 197.47 ± 61.12 min, P = 0.004), reduced intraoperative blood loss (179.83 ± 65.17 mL vs. 338.67 ± 126.55 mL, P < 0.001), and lower fluoroscopy frequency (13.42 ± 8.19 times vs. 29.87 ± 10.11 times, P < 0.001). Detailed results are presented in Table 2. Table 2Comparison of surgical indicators between the two groups of patients.GroupCasesOperation time (min)Blood loss (mL)Fluoroscopy (times)Control15197.47 ± 61.12338.67 ± 126.5529.87 ± 10.11Guide24141.54 ± 50.09179.83 ± 65.1713.42 ± 8.19t-value-3.124.505.57P-value-0.004< 0.001< 0.001 Comparison of surgical indicators between the two groups of patients. Accuracy of screw placement A total of 141 pedicle screws were placed (88 in the guide plate group and 53 in the control group). Inter-observer agreement for Kawaguchi grading was substantial (κ = 0.84, 95% CI 0.76–0.91). Postoperative 3D-CT evaluation using the Kawaguchi classification (Figure 5) revealed significantly higher accuracy in the guide plate group. The rate of clinically acceptable screw placement (Grade 0 or 1) was 97.7% (86/88) in the guide plate group compared to 81.1% (43/53) in the control group (OR = 6.2, 95% CI 1.6–24.3, P = 0.008), with perfectly placed screws (Grade 0) accounting for 81.8% (72/88) vs. 62.3% (33/53; OR = 2.8, 95% CI 1.3–6.1, P = 0.009), while unacceptable placement (Grade 2 or 3) occurred in only 2.3% (2/88) of screws in the guide plate group vs. 18.9% (10/53) in the control group. Detailed accuracy data are summarized in Table 3. Postoperative imaging studies. (A, B) Sagittal, (C) coronal, and (D) transverse views of the three-dimensional computed tomography (3D-CT) reconstruction of the cervical spine. Table 3Accuracy of screw placement: acceptability and grading.GroupCasesTotalGrade 0Grade 0,1Control15533343Guide24887286OR (95% CI)--2.8 (1.3–6.1)6.2 (1.6–24.3)P-value*--0.0090.008*P values were derived from mixed-effects logistic regression. Accuracy of screw placement: acceptability and grading. *P values were derived from mixed-effects logistic regression. Stratified analysis by vertebral level showed that the guide technique significantly improved accuracy at both C1 and C2 levels. For C1 screws, the acceptable placement rate was 97.2% (35/36) in the guide group vs. 76.2% (16/21) in the control group (P = 0.02). For C2 screws, the acceptable placement rate was 98.1% (51/52) vs. 84.4% (27/32), respectively (P = 0.03). Postoperative clinical outcomes Both surgical techniques led to significant and sustained improvements in all clinical outcome measures from preoperative baselines to the final follow-up (all P < 0.01 for within-group comparisons). JOA scores: Although the guide plate group showed a marginally higher mean score at 1 week postoperatively, no statistically significant differences were observed between the two groups at any evaluation time point (preoperative, 1-week, or final follow-up; Table 4). Table 4Comparison of JOA scores between the two groups.GroupCasesPre-op1 weekFinal follow-upt (Pre-op vs. 1 week)P (Pre-op vs. 1 week)t (Pre-op vs. final)P (Pre-op vs. final)Control159.07 ± 1.4911.53 ± 2.0315.13 ± 1.30−5.76< 0.001−15.10< 0.001Guide249.79 ± 1.9812.96 ± 2.2415.62 ± 1.47−7.68< 0.001−14.92< 0.001t-value-−1.22−1.92−1.02----P-value-0.230.060.31----P-values were derived from linear mixed-effects models with Bonferroni-Holm correction for multiple comparisons. Comparison of JOA scores between the two groups. P-values were derived from linear mixed-effects models with Bonferroni-Holm correction for multiple comparisons. VAS scores: Pain scores decreased significantly in both groups postoperatively. At the 1-week assessment, the guide plate group exhibited a lower mean VAS score (2.46 ± 0.88) than the control group (3.00 ± 0.76), although this difference approached but did not reach statistical significance (P = 0.057). By the final follow-up, both groups reported minimal pain with no significant intergroup difference (Table 5). Table 5Comparison of VAS scores between the two groups.GroupCasesPre-op1 weekFinal follow-upt (Pre-op vs. 1 week)P (Pre-op vs. 1 week)t (Pre-op vs. final)P (Pre-op vs. final)Control156.20 ± 1.273.00 ± 0.760.60 ± 0.7410.65< 0.00116.30< 0.001Guide246.38 ± 1.662.46 ± 0.880.38 ± 0.6512.55< 0.00114.65< 0.001t-value-−0.351.860.96---P-value-0.730.060.34---P-values were derived from linear mixed-effects models with Bonferroni-Holm correction for multiple comparisons. Comparison of VAS scores between the two groups. P-values were derived from linear mixed-effects models with Bonferroni-Holm correction for multiple comparisons. ASIA scores: Both ASIA sensory and motor scores improved markedly after surgery in both cohorts. There were no significant differences between the guide plate and control groups at the preoperative, 1-week postoperative, or final follow-up assessments (Tables 6 and 7). Table 6Comparison of ASIA sensory scores between the two groups.GroupCasesPre-op1 WeekFinal Follow-upt (Pre-op vs. 1 week)P (Pre-op vs. 1 week)t (Pre-op vs. final)P (Pre-op vs. final)Control15198.67 ± 12.76210.53 ± 6.92221.07 ± 4.20−6.02< 0.001−6.15< 0.001Guide24201.16 ± 13.48213.50 ± 5.37222.21 ± 4.03−6.18< 0.001−9.35< 0.001t-value-−0.58−1.43−0.81---P-value-0.570.160.42---P-values were derived from linear mixed-effects models with Bonferroni-Holm correction for multiple comparisons. Comparison of ASIA sensory scores between the two groups. P-values were derived from linear mixed-effects models with Bonferroni-Holm correction for multiple comparisons. Table 7Comparison of ASIA motor scores between the two groups.GroupCasesPre-op1 weekFinal follow-upt (Pre-op vs. 1 week)P (Pre-op vs. 1 week)t (Pre-op vs. final)P (Pre-op vs. final)Control1587.27 ± 8.4292.27 ± 5.6498.87 ± 1.77−3.600.003−5.82< 0.001Guide2489.33 ± 10.6395.00 ± 5.6798.92 ± 1.82−4.60< 0.001−4.98< 0.001t-value-−0.64−1.40−0.08---P-value-0.530.160.94---P-values were derived from linear mixed-effects models with Bonferroni-Holm correction for multiple comparisons. Comparison of ASIA motor scores between the two groups. P-values were derived from linear mixed-effects models with Bonferroni-Holm correction for multiple comparisons. Fusion rates Fusion status was assessed at the final follow-up using CT and dynamic radiographs. In the guide plate group, 23 of 24 patients (95.8%) achieved solid fusion at the final follow-up, compared to 13 of 15 patients (86.7%) in the control group. The difference between groups was not statistically significant (χ2 = 0.342, P = 0.559). The single case of non-union in the guide group and two cases in the control group were managed with continued collar immobilization; no patient required revision surgery for non-union during the follow-up period. Postoperative complications No severe neurovascular complications related to screw malposition occurred in either group. Specific complications are summarized below. Guide Plate Group: One patient experienced postoperative dyspnea requiring transient intensive care unit admission and intubation, with subsequent full recovery. One case of occipital pressure ulcer was noted, which healed with enhanced wound care. Control Group: Two patients developed cerebrospinal fluid leakage associated with delayed wound healing; both were managed successfully with conservative measures including head elevation and localized wound care. During follow-up, some patients in both groups reported transient neck or shoulder stiffness or discomfort, which improved with guided rehabilitation. Discussion The atlantoaxial region presents a unique surgical challenge due to its intricate and highly variable anatomy (19, 20). Achieving accurate pedicle screw placement in this area is technically demanding, as even minor deviations risk injury to critical neurovascular structures such as the vertebral artery, spinal cord, and nerve roots (21). Beyond immediate injury, malpositioned screws can compromise biomechanical stability, potentially leading to construct failure and spinal instability. Therefore, the pursuit of a reliable and precise method for upper cervical pedicle screw insertion is of paramount clinical importance. Posterior atlantoaxial fusion is a well-established treatment for instability. While various techniques exist (e.g., Gallie, Brooks, and transarticular screws), cervical pedicle screw fixation, popularized since Abumi's report, has become a mainstay due to its superior three-column fixation, biomechanical strength, and high fusion rates (22). However, the procedure's inherent technical difficulty and proximity to vital structures necessitate exceptional precision. The conventional free-hand technique, reliant on anatomical landmarks and intraoperative fluoroscopy, is limited by its dependence on two-dimensional imaging to navigate complex three-dimensional anatomy. This often results in the need for repeated fluoroscopic verification and trajectory adjustments, which prolongs operative time and increases radiation exposure. Literature reports pedicle breach rates for this technique ranging from 6.7 to 12% (23, 24). In our study, the control group exhibited a breach rate of 37.74% (20/53 screws), with nearly half of these breaches (10 screws) classified as Grade 2 (>2 mm). This high rate of significant breaches underscores the limitations and potential risks associated with the free-hand method, particularly in anatomically challenging regions. Modern technologies, such as computer-assisted navigation and robotics, address these limitations by providing real-time, three-dimensional spatial guidance, significantly enhancing accuracy and safety (25). However, their clinical adoption is often hindered by substantial costs, procedural complexity, a steep learning curve, sensitivity to intraoperative movement requiring re-registration, and concerns regarding increased radiation exposure during setup (26, 27). Advances in digital orthopedics have positioned 3D printing technology as a compelling alternative for spine surgery, enabling the creation of patient-specific surgical guides based on meticulous preoperative virtual planning (28, 29). The present study demonstrates its significant benefits: the guide plate group achieved a markedly lower overall pedicle breach rate of 18.18% (16/88 screws) and a significantly higher proportion of perfectly placed (Grade 0) screws (81.82 vs. 62.26%) and clinically acceptable screws (97.73 vs. 81.13%) compared to the free-hand group. These findings are corroborated by existing literature, where Guo et al. (30) reported a >20% improvement in acceptable screw placement with a substantial reduction in fluoroscopy use, Niu et al. (31) documented superior accuracy (95.7 vs. 80.0%) alongside reductions in operative time, blood loss, and fluoroscopy, and Garg et al. (28) confirmed significantly higher accuracy with 3D-printed templates in complex deformities. Furthermore, the technology offers enhanced preoperative visualization and personalized planning, directly contributing to improved intraoperative accuracy (32). Importantly, while the guide plate technique demonstrably enhanced procedural safety and efficiency—reducing technical difficulty, operative time, and radiation exposure—both groups in our study ultimately achieved comparable and significant postoperative improvements in VAS, JOA, and ASIA scores. Therefore, 3D-printed guide-assisted pedicle screw placement in the upper cervical spine is a feasible and precise method that offers a valuable and cost-effective alternative to conventional free-hand techniques. Despite the promising outcomes, this study has several limitations. First, as a single-center, non-randomized controlled study with a relatively small sample size, potential selection bias cannot be excluded. Second, the application of 3D-printed guides requires additional preoperative time and costs for imaging processing and printing, rendering it unsuitable for emergency cases. Furthermore, the accuracy of the guide depends on its precise intraoperative fit to the bone surface, which may be compromised by inadequate soft-tissue dissection or anatomical variations. Third, the long-term clinical outcomes and cost-effectiveness of this technique need to be further validated through multicenter, randomized trials with larger sample sizes. Fourth, the chronological cohort design introduces potential confounding from technical evolution and surgeon experience over time. While the same senior surgical team performed all procedures to minimize operator-related variability, the inherent limitations of a non-randomized, single-center design remain. Additionally, the dissociation between improved screw accuracy and comparable clinical outcomes may be attributed to ceiling effects, the safety-oriented nature of accuracy (rather than functional recovery), limited statistical power, and insufficient follow-up to detect late sequelae. Conclusion This study demonstrates that patient-specific 3D-printed guide plates effectively translate preoperative digital planning into precise intraoperative execution for upper cervical pedicle screw placement. Compared to the conventional free-hand technique, this method significantly improves screw accuracy, reduces operative time, blood loss, and radiation exposure, while achieving comparable clinical outcomes. These findings support the conclusion that 3D-printed guide plate-assisted pedicle screw fixation is a safe, reliable, and clinically valuable alternative for upper cervical spine surgery, meriting broader clinical adoption.