Bioimpedance: From Body Composition Assessment to Emerging Technologies in Clinical Nutrition Authors: Ruth Nekane Pita, Isabel María Fernández, Oibar Martínez, Jose Miguel Miranda Published: 2026 ABSTRACT This review examines the physical foundations, predictive models, emerging technologies and recent research trends in bioimpedance for clinical nutrition. A bibliometric analysis of cumulative scientific production from 1980 to 2025 shows that weighted logistic models provide a good descriptive fit to the evolution of the literature (R2 ≥ 0.999). Bioimpedance research remains in an expanding phase, with its projected inflection point lying beyond 2026. These fits are descriptive and model-dependent: high coefficients of determination on cumulative counts are not in themselves evidence of predictive validity, and inflection points beyond the observation window are extrapolations. We introduce the Measurement-Independent Factor (MIF), which isolates the anthropometric contribution from the bioimpedance measurement in predictive equations. Phase Angle (PhA) has emerged as the most extensively studied bioimpedance descriptor, since it avoids the population-specific prediction equations required for estimates such as FFM or FM, although its absolute value remains dependent on measurement conditions, instrumentation and biological characteristics. Body-composition estimates obtained with devices from different manufacturers should not be interpreted interchangeably. Electrode technologies, bioimpedance wearables, and bioimpedance imaging are reviewed as major emerging research directions. A second bibliometric analysis covering the most-cited literature from 2020 to 2026 and 1655 original articles published during 2025–2026 show that recent research focuses primarily on obesity and adiposity, together with sarcopenia and frailty, whereas critical and acute care remain underrepresented. 1. Bioimpedance in Nutrition 1.1. Scope, Terminology, and Organization Bioelectrical impedance analysis (BIA) is currently one of the most widely used techniques for assessing body composition and indirectly estimating basal metabolic rate (BMR). Its widespread adoption is attributable to a unique combination of four major characteristics rarely found together in physiological measurements: non-invasive nature, portability, ease of operation, and low cost. BIA provides the estimates of several physiologically relevant body compartments, including total body water (TBW), fat-free mass (FFM), fat mass (FM), extracellular water (ECW), and intracellular water (ICW). Beyond conventional BIA, two major branches of bioimpedance emerged: bioimpedance spectroscopy (BIS), for characterizing tissue dielectric properties and body fluid distribution, and electrical impedance tomography (EIT), for functional conductivity imaging. This work is a narrative methodological review. All clinical evidence discussed in the following sections is drawn from the published literature. Three analyses were generated specifically for this manuscript: the bibliometric assessment of publication activity described in Section 1.3, the algebraic decomposition of published prediction equations introduced as the measurement-independent factor (MIF), and the Cole model simulation of Section 2.4, fitted to public bioimpedance records. All three are exploratory and descriptive in nature, and are presented as illustrative analyses rather than as validation studies. Three terms are used throughout this review. A reference method is a technique used to calibrate or benchmark another. A criterion method is the technique taken as comparator in a given validation study. A validated method is one whose agreement with such a comparator has been quantified in the population of interest. BIA is treated here as a validated method for specific populations and applications, and not as a reference method. The review is addressed to readers from both clinical nutrition and biomedical engineering, and is organized as follows. The historical development of the technique is outlined first, followed by the growth of the corresponding literature. Section 2 then presents the physical basis of the measurement. Two points receive particular attention there: the complex phasor representation of impedance and the capacitive behavior of cell membranes. Neither is always stated explicitly in the recent clinical literature, although both help to interpret what is measured. The conversion of electrical quantities into body composition estimates and the limits of that conversion are examined in Section 3. Emerging technologies related to those limits follow in Section 4. The recent literature is considered the last. Among the most-cited work published between 2020 and 2026, sustained attention is apparent in a limited number of clinical areas. Citation counts, however, need time to accumulate. Non-review articles from 2025 to 2026 are therefore examined separately at the end of Section 5. That comparison offers a tentative indication of which of the gaps noted above are being addressed. 1.2. First Steps The origins of bioimpedance date back to the nineteenth century, when investigators first recognized that biological tissues exhibit frequency-dependent electrical properties [1]. The theoretical description of dielectric dispersion that followed provided the physical framework underpinning impedance spectroscopy [2]. Early studies demonstrated that tissues containing high concentrations of water and electrolytes, such as blood and skeletal muscle, conduct electrical current far more readily than adipose tissue or bone. Thomasset reported electrical impedance measurements with needle electrodes to determine TBW [3] and was the first to employ two measurement frequencies (1 and 100 kHz) to estimate ECW [4]. Nyboer then introduced the tetrapolar surface-electrode configuration for bioimpedance measurements [5], while Hoffer et al. demonstrated a strong correlation between whole-body bioimpedance and TBW, validating the method against tritium dilution techniques [6]. During the 1980s, Lukaski et al. showed that whole-body electrical impedance could be used to accurately estimate FFM [7], subsequently validating the tetrapolar impedance method against these reference techniques [8]. Concurrently, Kushner and Schoeller established that the underlying physical basis of this estimation relies on the strong correlation between TBW and the impedance index, mathematically expressed as height squared divided by resistance [9]. Further studies combined raw electrical measurements with anthropometric variables in multivariate regression models, improving predictive accuracy across diverse populations [10]. However, Deurenberg et al. documented the errors that arise when equations developed in one population are applied to subjects differing in age, sex or body composition [11]. The National Institutes of Health Technology Assessment Conference reinforced these observations, emphasizing that the validity of BIA predictions depends on the characteristics of the calibration cohort and on key physiological assumptions of the method. The chief among these is the assumption that the hydration fraction of FFM remains constant at approximately 73.2% in healthy, normohydrated subjects [12]. Sun et al. later improved the basic BIA 2-component model (fat and fat-free) with a multicomponent reference method accounting for variations in mineral bone, water, and fat [13]. To overcome limitations in fluid distribution assessment and regional body composition, Ward et al. further developed multifrequency BIA (MF-BIA), in which impedance is measured at several discrete frequencies to improve the characterization of body fluid compartments [14]. This development was soon complemented by segmental BIA, which extended impedance measurements to obtain direct information from limbs and trunk [15]. Bioimpedance research expanded beyond whole-body composition analysis toward imaging techniques capable of reconstructing the spatial distribution of tissue electrical properties. EIT was originally inspired by geophysical electrical prospecting methods. It was introduced for medical applications by Barber, Brown, and co-workers, who developed one of the first systems capable of generating cross-sectional conductivity images from surface-electrode measurements [16]. Their initial work identified the thorax as the most promising clinical target because of the large conductivity contrast between air-filled lungs, blood, and surrounding tissues. Cheney, Isaacson, and Newell presented a comprehensive description of the physical and mathematical foundations of EIT together with the Adaptive Current Tomography system [17]. These theoretical developments were further consolidated by Borcea, who reviewed the inverse conductivity problem [18]. EIT evolution from an experimental imaging modality to a clinically useful bedside monitoring technique was reviewed by Adler and Holder [19], and Bodenstein et al., who focused on mechanically ventilated patients [20]. BIS emerged progressively from early bioimpedance developments [21]. Methodological analyses have shown that BIS provides greater biological insight than conventional regression-based BIA [22]. This advantage arises because BIS relies on electrical models of tissue behavior rather than solely on statistical prediction equations. By extrapolating the measured spectrum to extreme frequencies (f → 0, f → ∞), BIS estimates ECW and TBW from electrically meaningful parameters. This approach reduces dependence on arbitrary measurement frequencies and yields a more physiologically grounded description of body fluid distribution. Bioelectrical Impedance Vector Analysis (BIVA) technique is attributed to Piccoli et al., who first described the method in 1994 [23]. In a second study, the same group defined the bivariate reference tolerance ellipses that transformed BIVA into a practical tool for the assessment of hydration status and body cell mass (BCM) in both healthy subjects and patients [24]. While traditional BIA relies on empirical regression equations, BIVA directly analyzes the raw impedance vector plotted on a resistance–reactance graph. BIVA has been proven highly valuable in clinical conditions associated with altered hydration and body composition, including severe obesity, where conventional regression equations often provide inaccurate estimates [25]. The 2004 guidelines of the European Society for Clinical Nutrition and Metabolism (ESPEN) played a major role in standardizing and consolidating the clinical application of BIA. Before these consensus documents, the clinical implementation of BIA remained heterogeneous, with considerable variability in measurement protocols and interpretation across institutions. The first ESPEN guideline synthesized the theoretical foundations, methodological requirements, and technical limitations of BIA, providing a standardized scientific framework for its clinical application [26]. The companion paper translated these principles into evidence-based clinical recommendations [27]. Reference methods evolved from hydrodensitometry and direct anthropometry [28], toward more accurate techniques, including isotope dilution, neutron activation analysis, magnetic resonance imaging and computed tomography [29]. Dual-energy X-ray absorptiometry (DXA) has become one of the most popular reference methods during the last 25 years [30]. Three distinct uses of BIA are separated throughout this review. The first is the estimation of absolute compartment values, which depends entirely on prediction equations and on the population in which they were calibrated. The second is the intraindividual monitoring of change over time, in which a stable equation bias largely cancels, and the relevant quantity is the difference rather than the absolute value. The third is the direct interpretation of raw bioelectrical parameters, such as PhA and BIVA, which requires no regression step. Most of the discrepancies discussed in the following sections concern the first use only, and a conclusion established for one of the three does not transfer to the others. 1.3. Bibliometric Search Strategy The evolution of research activity in body composition assessment within the field of nutrition was studied through a systematic literature search. This search was performed on 22 July 2026 using the Web of Science (WoS) Core Collection database, with all ten editions available under our institutional subscription selected: Science Citation Index Expanded, Social Sciences Citation Index, Arts & Humanities Citation Index, Conference Proceedings Citation Index—Science, Conference Proceedings Citation Index—Social Sciences & Humanities, Book Citation Index—Science, Book Citation Index—Social Sciences & Humanities, Emerging Sources Citation Index, Current Chemical Reactions, and Index Chemicus. No restriction on publication year was applied at the search stage, and the analysis reported here covers the period 1980–2025. The analysis included five of the most widely used instrumental techniques for body composition assessment: BIA, DXA, computed tomography (CT), magnetic resonance imaging (MRI), and ultrasound (US). A second, separate corpus is analyzed in Section 5: it applies the same search equations to the 2020–2026 window, without the restriction to the Article document type, and retains only publications above a citation-rate threshold. Its counts are therefore neither a subset of nor comparable with those reported here. Scientific terminology is not fully standardized; therefore, a simple keyword search may miss a substantial number of relevant publications. To overcome this limitation, dedicated search equations were designed for each modality. Each equation combines the principal terms and accepted synonyms of the corresponding technique. These were then combined with a common set of nutrition-related terms using the Topic Search (TS) field of the WoS database. Table 1 shows the complete search equations. The inclusion of alternative terminology, including acronyms, full technique names and common descriptors, was intended to improve sensitivity without sacrificing specificity. This strategy reduces the risk of excluding relevant studies due to nomenclature differences across journals and research communities. Three main criteria governed the design of the equations. First, the five technique blocks were built to a comparable level of specificity. Generic expressions such as “magnetic resonance” also retrieve spectroscopy and non-imaging applications, so they were replaced by explicit imaging terms. Second, phrases were truncated inside quotation marks rather than left unquoted. WoS applies automatic lemmatization to unquoted Topic terms but disables it within exact phrases, so internal truncation preserves morphological variants (impedance/impedances, absorptiometry/absorptiometric, tomography/tomographic) while retaining the precision of the phrase. Third, acronyms were admitted only where unambiguous in this context. DXA, DEXA, MRI, BIA and BIVA were retained because the mandatory conjunction with the nutrition block confines them to their intended meaning. The isolated acronym CT was excluded, since it also denotes the cycle threshold of quantitative PCR and other unrelated concepts; bounded forms were used instead. Two domain-specific safeguards were applied. In the BIA equation, “phase angle” is retrieved only in conjunction with a bioimpedance term, which excludes the extensive electrical-engineering and gait-analysis literature that uses the same expression with an unrelated meaning. In the US equation, an exclusion clause removes ultrasound-assisted extraction and sonication, which is a large body of food-technology work that satisfies the nutrition block without measuring body composition; this contaminant is specific to US, and retaining it would have inflated that series alone. The nutrition block deliberately omits “body composition” as a search term, because it is a methodological descriptor rather than a marker of the clinical nutrition domain. Its inclusion would disproportionately expand the CT and MRI series through the radiological and oncological literature, again at the expense of comparability. All five queries were executed with the aim of reflecting publication activity rather than retrieval settings. Searches were therefore restricted to the Article document type, which excludes reviews. Reviews are commissioned and published at rates set by editorial dynamics that differ between modalities. Including them would therefore inflate the more established techniques. The Article document type is also the one most consistently assigned across the full time window. The retrieval procedure therefore involved no screening stage beyond the search itself. For each of the five techniques, the corresponding equation was run in the Topic Search field, restricted to the Article document type, and the complete record set was exported directly from the Web of Science interface without further selection. The annual and cumulative counts derived from these exports are those plotted in Figure 1. No record was excluded manually or automatically at any subsequent stage. 1.4. Evolution of Research Activity, 1980–2025 Figure 1 shows the results of the bibliometric search from the WoS Core Collection. The bibliometric data show that research volume in body composition techniques has increased steadily since 1980. Their trajectories, however, reflect different lifecycles. CT and MRI grew exponentially during the 1990s and established themselves as laboratory reference methods. In contrast, BIA has evolved from a basic predictive tool into a widely used methodology for nutritional assessment. BIA publications experienced a sharp acceleration during the 1990s. This growth was driven by the transition from single-frequency empirical equations to multifrequency models, alongside the validation of raw electrical parameters, such as PhA. DXA entered the literature later (1991) and quickly reached a plateau of 700–900 annual articles as the standard for the three-compartment model. BIA, in contrast, maintained a continuous upward trajectory and reached a peak of 1039 articles in 2025. Notably, the publication volume and growth rate of BIA are fully comparable to, and in some years exceed, those of DXA, CT and MRI. The inset of Figure 1 presents the cumulative scientific production alongside weighted logistic growth models fitted to the 1980–2025 period but was displayed for 2000–2025 for visual clarity. Model parameters were estimated independently for each technique by minimizing a weighted least-squares objective function, with cumulative counts normalized prior to optimization. To characterize the current maturity of each research field rather than its historical emergence, residuals were weighted using an exponential function that progressively increased the contribution of more recent years. This weighting reduced the influence of the sparse publication counts observed during the early stages of each field while preserving the information contained in the complete historical record. The weight assigned to year t was w(t) = exp[λ(t − 1980)], with a weighting rate λ = 0.03 yr−1. This rate is distinct from the intrinsic growth rate α of the logistic model. Weights were normalized to unity at the first year of the series. The relative contribution of a residual therefore increased monotonically by a factor of approximately 3.9 between 1980 and 2025. This value of λ is not a standard constant, and it was fixed a priori rather than estimated from the data. It corresponds to a weight doubling time of ln2/λ ≈ 23 years, about half of the 46-year window, so that the mature phase of each field carries roughly four times the weight of its emergence while no year is discarded. Sensitivity to the weighting and to the functional form was assessed by refitting all five series. Varying the weighting rate λ between zero and 0.06 yr−1 displaced the inflection year of BIA by 1.6 years in total, from 2032.5 to 2034.1, and changed its asymptote by less than 8%. Variations were smaller for the remaining techniques. The choice of λ therefore does not drive the conclusions. Replacing the logistic by a Gompertz model changed the estimated asymptotes far more substantially. For BIA and CT, the two series whose inflection lies beyond the observation window, the asymptote became effectively non-identifiable. K reached 1.9 × 107 and 1.3 × 108. The estimated inflection years moved into the twenty-second century. The confidence intervals exceeded the estimates themselves, and the Jacobian was numerically singular. The ordering of the five techniques by inflection year was nevertheless identical under both functional forms, from DXA and MRI through US to BIA and CT. The interpretation offered here therefore rests on that ordering and on the relative maturity it implies, and not on the absolute values of K or α. For the pre-inflection series, those values are governed by the functional form assumed rather than by the observed record. The logistic model provided an excellent description of the cumulative publication trends for all five techniques, with coefficients of determination equal to or above 0.999. Because cumulative counts are monotonic and highly structured variables, such values indicate a good descriptive fit rather than predictive validity. The estimated intrinsic growth rates α indicate that DXA expanded most rapidly over the analyzed period, whereas MRI and ultrasound showed intermediate rates and BIA and CT showed more gradual but sustained growth. MRI and US yielded inflection points close to the present, at y0 = 2025.4 ± 0.5 and 2026.4 ± 0.7, respectively. In contrast, BIA and CT produced substantially later inflection years, 2033.6 ± 2.8 and 2034.4 ± 1.7, whereas y0 for DXA was the earliest, at 2020.2 ± 0.7. Because the inflection years estimated for BIA and CT lie beyond the observation window, the corresponding K values should be interpreted as extrapolations rather than as observed saturation levels. All intervals reported here are nominal 95% weighted least-squares confidence intervals obtained from the Jacobian at the optimum. 2. Physical Principles 2.1. Why Is Bioimpedance a Complex Number? Bioimpedance (Z) is a complex quantity defined as the ratio of the voltage phasor (V) to the electric current phasor (I): (1)Z=V/I Phasors are complex-number representations of time-harmonic sinusoidal variables. They allow us to retrieve the time-dependent function they represent by multiplying the phasor by the complex exponential factor exp (jωt) and extracting the real part: (2)v(t)=ReV⋅ejωt=ReVmejθv⋅ejωt=Vmcos(ωt+θv) (3)i(t)=ReI⋅ejωt=ReImejθi⋅ejωt=Imcos(ωt+θi) where Vm and Im are the peak amplitudes of the voltage and current, respectively, ω is the angular frequency (ω=2πf), θv and θi are the phases of voltage and current, respectively, and j is the imaginary unit (j=−1). Complex impedance can be separated into its real and imaginary parts as (4)Z=R+jX The physical interpretation of these components determines how energy interacts with the medium:○Resistance. Quantifies the dissipation of electrical energy as heat. In biological tissues this dissipation is mostly due to collisions of ions moving through intra- and extracellular fluids. It is always a positive quantity.○Reactance. Quantifies the storage and release of electromagnetic energy within the system and can be either positive or negative. The sign depends entirely on the stored energy: a positive reactance indicates energy storage in a magnetic field, whereas a negative reactance indicates storage in an electric field. In circuital terms, it is customary to represent a positive reactance as an inductor and a negative one as a capacitor. From the circuit theory, it follows that the reactance of a pure inductor with inductance L is XL=Lω, whereas the reactance of a pure capacitor with capacitance C is XC=−1/(ωC). ○Resistance. Quantifies the dissipation of electrical energy as heat. In biological tissues this dissipation is mostly due to collisions of ions moving through intra- and extracellular fluids. It is always a positive quantity. Resistance. Quantifies the dissipation of electrical energy as heat. In biological tissues this dissipation is mostly due to collisions of ions moving through intra- and extracellular fluids. It is always a positive quantity. ○Reactance. Quantifies the storage and release of electromagnetic energy within the system and can be either positive or negative. The sign depends entirely on the stored energy: a positive reactance indicates energy storage in a magnetic field, whereas a negative reactance indicates storage in an electric field. In circuital terms, it is customary to represent a positive reactance as an inductor and a negative one as a capacitor. From the circuit theory, it follows that the reactance of a pure inductor with inductance L is XL=Lω, whereas the reactance of a pure capacitor with capacitance C is XC=−1/(ωC). Reactance. Quantifies the storage and release of electromagnetic energy within the system and can be either positive or negative. The sign depends entirely on the stored energy: a positive reactance indicates energy storage in a magnetic field, whereas a negative reactance indicates storage in an electric field. In circuital terms, it is customary to represent a positive reactance as an inductor and a negative one as a capacitor. From the circuit theory, it follows that the reactance of a pure inductor with inductance L is XL=Lω, whereas the reactance of a pure capacitor with capacitance C is XC=−1/(ωC). In some applications, bioimpedance can also be expressed in terms of the modulus |Z| and phase angle φA, (5)Z=|Z|ejφA where Z = R2+X2; and φA=θv−θi. Biological tissues store energy almost exclusively within an electric field. Because this capacitive storage of electrical energy is heavily dominant over any negligible magnetic inductive properties, bioimpedance characteristically exhibits a negative reactance. However, the imaginary part is typically small over the frequency range employed in BIA. For this reason, a number of analytical models treat each segment as a purely resistive element and neglect the phase contribution. Accordingly, the argument of Z in Equation (4) is negative for a capacitive biological load. The clinical phase angle reported throughout this review follows the positive BIA convention, PhA = arctan(|X|/R), and is thus the negative of that argument. 2.2. Segmental Bioimpedance Segmental bioimpedance analysis (SBIA) is based on the measurement of individual body segment impedances. Compared to conventional whole-body bioimpedance analysis, SBIA provides more localized information about tissue composition and physiological changes, allowing independent assessment of each limb and trunk. Figure 2 illustrates the electrode arrangement and the equivalent electrical circuits for representative measurements. The measurements are performed using the tetrapolar method, which minimizes the influence of the electrode–skin interface impedance. Although the measurement instrument incorporates eight surface electrodes (E1–E8), only four are active during each acquisition. An electronic switching network automatically selects the appropriate electrodes depending on the body segment to be measured. As summarized in the auxiliary table, two electrodes are used to inject a sinusoidal current (I), while two different electrodes are employed to measure the resulting voltage drop (V). Since the voltage-sensing circuit has a very high input impedance, virtually no current flows through the voltage electrodes, making the measured voltage independent of the contact impedance at these electrodes. Consequently, the calculated impedance is determined almost exclusively by the electrical properties of the selected body segment. The electrode selection depends on the anatomical region under investigation. For example, when the left arm is measured, current is injected through electrodes E2 and E4, whereas the voltage is sensed between E3 and E7. Similarly, different electrode combinations are automatically selected for the trunk, each leg, or palm-to-foot measurements for full body assessment. This switching strategy enables all segmental measurements to be performed without repositioning the electrodes, ensuring repeatability and fast acquisition. The usefulness of the tetrapolar method becomes evident when analyzing the equivalent circuits. As an example, the equivalent circuit shown on the left of Figure 2 represents the measurement of the right arm. The excitation current is injected through the upper electrodes and flows through the arm segment under investigation. Because the voltmeter draws negligible current, no significant voltage drop is produced across the voltage electrode contact impedances. Furthermore, the current flowing through the trunk and the remaining limbs produces only a negligible contribution to the measured voltage because the voltage electrodes are positioned immediately across the right arm. Under these conditions, the measured impedance is ZAR. A similar principle applies to trunk measurements, illustrated in the right-hand equivalent circuit of Figure 2. Z1 and Z2 represent the local tissue impedances between the current-injecting electrode and the voltage-sensing electrode located on the same hand or foot. Since these impedances lie exclusively within the current injection path, they are connected in series with the body segment under investigation and carry the same excitation current. However, they are located outside the voltage measurement terminals and therefore do not contribute to the measured voltage. Consequently, neither Z1 nor Z2 affects the measured impedance. These impedances must remain finite rather than being assumed equal to zero. Otherwise, the current-injecting and voltage-sensing electrodes would become electrically equivalent, effectively reducing the measurement to a two-terminal (bipolar) configuration. If a conventional two-electrode configuration was used, where the same pair of electrodes injects the current and measures the resulting voltage drop, the high impedance of the skin–electrode interface would dominate the measurement, masking the impedance contributions of the internal tissues. 2.3. Body Composition Compartments Bioimpedance does not measure body composition directly. Instead, it determines the electrical properties of biological tissues. From these properties, several physiologically meaningful compartments can be estimated through validated prediction models. Figure 3 summarizes the hierarchical organization of human body composition. It also illustrates the principal compartments that can be estimated using SBIA. At the highest level, total body weight can be divided into FM and FFM. FM comprises all lipid-containing tissues, whereas FFM includes all metabolically active tissues, including skeletal muscle, internal organs, bone, connective tissues, and TBW. This compartment can be further subdivided into mineral mass and soft lean tissues. A total of 95% of minerals are contained in bones, and, in fact, the mineral compartment is usually referred to as Bone Mineral Content (BMC). The combination of FM, BMC, proteins and water is usually referred to as the four-compartment (4C) model. Cortical bone exhibits high electrical resistivity because of its low water content and highly mineralized structure. However, the contribution of bone to the measured impedance is limited, as is that of the tissues enclosed by the cortical bone. This is because bone represents a relatively small fraction of the conductive volume. Conversely, skeletal muscles and visceral organs contain large amounts of intracellular and extracellular fluids and therefore dominate the electrical conduction pathways. As a result, bioimpedance measurements are particularly sensitive to changes in lean soft tissue. The lean soft-tissue compartment can be further separated into proteins and TBW. Structural proteins constitute the solid component of muscle and organ tissues, whereas TBW forms the conductive medium through which electrical current flows. Both MF-BIA and BIS enable the independent estimation of ECW, ICW, and TBW by exploiting the frequency dependence of tissue impedance. Figure 4 depicts a simplified equivalent circuit of a cell together with the corresponding current pathways at low and high excitation frequencies. The selective sensitivity of low frequencies to ECW and of high frequencies to TBW provides the physical foundation for estimating ICW and for assessing changes in hydration, nutritional status, and cellular integrity. Each biological cell can be represented by an equivalent electrical circuit consisting of an intracellular resistance (RI) surrounded by a cell membrane. Electrically, the membrane behaves predominantly as a capacitor (CM) because its phospholipid bilayer acts as an excellent dielectric separating two conductive electrolyte solutions. This makes the membrane impedance strongly frequency-dependent, decreasing as the frequency increases. At low excitation frequencies, the membrane capacitive reactance is very large. This effectively prevents alternating current from entering the cells. The current follows the path of least impedance and flows almost exclusively through ECW, bypassing the intracellular compartment. As illustrated in the upper part of Figure 4, current streamlines deviate around the cells because membranes behave as insulating barriers. Consequently, the measured impedance primarily reflects extracellular electrical conductivity. It is therefore highly correlated with ECW volume. As the excitation frequency increases, the membrane reactance progressively decreases. At high frequencies, the capacitive impedance becomes sufficiently small for the current to cross the cell membranes and enter the intracellular space. The membranes no longer constitute a significant obstacle to current flow, allowing both the extracellular and intracellular electrolyte solutions to participate in electrical conduction. As illustrated in the lower part of Figure 4, the current traverses the cells rather than circumventing them, increasing the effective conductive volume. ICW can be obtained simply as (6)ICW=TBW−ECW 2.4. Frequency Dependence and Tissue Electrical Properties The frequency-dependent dielectrical behavior of biological tissues dictates how different body constituents can be identified and quantified across the spectrum. The resistance of a biological tissue is fundamentally determined by its electrical conductivity σ, an intrinsic physical property that quantifies how easily a material allows free ions to move under an applied electric field. Resistance is inversely proportional to σ, and mediated by the geometry of the tissue. For the simple geometry of a cylinder of length L and cross-sectional area A, (7)R=L/(σA) The electrolytic solutions making up the intracellular and extracellular fluids provide a high density of mobile ions that drive this conductivity, allowing the medium to dissipate electrical energy as heat. Tissue reactance is directly connected to dielectric permittivity εr′, a physical property that measures a material’s capacity to polarize and store energy within an electric field. Cell membranes—formed by an insulating lipid bilayer sandwiched between conductive media—act as biological capacitors whose capacitance is proportional to their permittivity. When an alternating current is applied, this structural arrangement temporarily stores charge and induces a phase shift in the voltage signal, which macroscopically translates into capacitive reactance. Within the frequency range typically employed in bioimpedance measurements, the impedance spectrum is well-described by the classical Cole–Cole model: (8)Zω=R∞+R0−R∞1+(jωτ)α where R0 is the resistance at zero frequency, corresponding to current flowing exclusively through the extracellular fluid, R∞ is the high frequency resistance, at which current is assumed to flow through both extracellular and intracellular compartments, τ is the characteristic relaxation time, and α (0<α≤1) describes the broadening of the dielectric dispersion. In whole-body bioimpedance measurements, the impedance locus follows a depressed semicircular arc rather than the ideal semicircle predicted by a single-time-constant RC circuit. This indicates that the frequency-dependent electrical behavior of biological tissues is governed by a distributed relaxation process. As the excitation frequency increases, the measured impedance moves along the Cole arc from right to left. Figure 5 illustrates this behavior for three individuals from the multifrequency bioimpedance records of the NHANES 1999–2000 cycle, a public and de-identified dataset [32]. No measurement was performed for the present work. The three loci differ mainly in their position along the real axis. Both R0 and R∞ increase as body size decreases, which is consistent with the dependence of whole-body resistance on the geometry of the conductive path. The diameter of the arc, proportional to R0 − R∞, reflects the progressive recruitment of ICW as the frequency increases. Its height is related to the reactance and therefore to the capacitive behavior of cell membranes. Two parameters vary far less than the resistances. The exponent α stays close to 0.32 in the three records, and the characteristic time τ is indistinguishable at the resolution reported here, near 1.2 μs, which corresponds to a characteristic frequency of about 130 kHz. This is not unexpected. R0 and R∞ scale with the length and cross-sectional area of the conductive volume, whereas τ and α depend on intensive properties of the tissue, namely the membrane capacitance per unit area and the resistivity of the intracellular and extracellular media. A similar relaxation time across individuals is therefore consistent with the model. The observation should not be over-interpreted. With a finite number of measured frequencies, τ and α are the least well-determined parameters of the fit, and three records say nothing about the range of variation expected in a population. Both electrical conductivity and dielectric permittivity exhibit strong frequency-dependent variations known as dielectric dispersions. In the spectrum where commercial devices operate, this phenomenon is primarily driven by the structural polarization of cell membranes and the relaxation of proteins, causing permittivity to drop dramatically and conductivity to increase as the frequency of the applied electric field rises. Figure 6 depicts the relative permittivity and electrical conductivity of five primary body constituents as a function of the frequency, according to IT’IS database [33]. A pronounced contrast appears between highly hydrated tissues with rich extracellular fluid networks, such as muscle and blood, and poorly hydrated tissues such as adipose tissue and cortical bone. This occurs because current is restricted to extracellular pathways at low frequencies. Highly hydrated tissues manifest both elevated permittivity and conductivity, whereas fat and bone maintain significantly lower values. This wide gap in dielectric properties at low frequencies provides the necessary sensitivity to differentiate between FFM and FM, making low-frequency measurements very useful for characterizing basic tissue composition. As the measurement frequency increases into the megahertz and gigahertz ranges, the dielectric properties undergo significant transitions known as beta and gamma dispersions. During beta dispersion, which occurs primarily in the radiofrequency range, the capacitive reactance of the cell membranes decreases to the point where the lipid bilayers become electrically transparent. While this enables the measurement of TBW, it simultaneously reduces the structural contrast between different tissue types. At the gigahertz range gamma dispersion takes over, dominated by the dielectric relaxation of free water molecules. In this domain, the conductivity of all tissues rises sharply and converges toward similar orders of magnitude, which drastically diminishes the sensitivity required to isolate or differentiate specific body constituents based purely on their electrical properties. The electrical behavior of skin introduces a challenge in SBIA instrumentation. At low frequencies, skin exhibits very low values of σ, being several orders of magnitude lower than that of blood or muscle. This low conductivity is primarily a characteristic of the stratum corneum, the outermost layer of dead skin cells that acts as a robust insulating barrier. In addition, the electrical conductivity of the cortical bone remains notably low at frequencies below 1 GHz, implying that the dense cortical shell restricts current penetration into deeper tissues. Consequently, physiological changes within the bone marrow are unlikely to be detected using BIA. 3. From Bioimpedance to Body Composition 3.1. BIA Prediction Equations BIA Prediction equations (BPEs) remain largely empirical. Most models combine impedance variables with anthropometric parameters calibrated for specific devices and populations. A total of 106 BPEs derived from 64 studies involving more than 40,000 participants have recently been compiled and analyzed [34]. Likewise, Kampo et al. reviewed more specifically multifrequency BPEs published between 2000 and 2025 [35]. Population-specific equations continue to be developed, such as an FFM equation derived and validated in Greek Caucasian adults [36]. For most manufacturers, the BPEs implemented in commercial bioimpedance analyzers are not publicly disclosed. These equations are typically developed from reference body composition techniques applied to extensive population studies, conducted by the manufacturer itself. Consequently, body composition estimates obtained with different commercial instruments are not interchangeable. Moreover, estimates from a commercially available bioimpedance analyzer can still be globally refined using DXA-derived correction factors, even when cohorts comprise more than 900 individuals [37]. Table 2 presents three representative BPEs proposed by Lukaski et al. [38], Kyle et al. [39], and Hughes et al. [40], all of which showed an excellent agreement with established reference methods. We use the term ALSM (appendicular lean soft mass) to refer to the variable estimated by the equation of Kyle et al., although the original publication uses ASMM (appendicular skeletal muscle mass). Both terms refer to closely related physiological compartments and are frequently used interchangeably in the literature. ALSM, or the equivalent of appendicular lean soft tissue (ALST), is now preferred in DXA-based research because DXA measures this component directly. Each equation can be decomposed into two additive components: an electrical term, which depends on the bioimpedance, and a measurement-independent factor (MIF). MIF is introduced in this work to represent the contribution of all regression terms that are independent of the measurement. The decomposition is an exact algebraic rearrangement of the published equation rather than a statistical estimate, and it applies only where the impedance and the anthropometric terms enter additively. The coefficient of determination quantifies the proportion of the variance in the reference body composition variable explained by the equation. Values closer to unity indicate a better fit. The standard error of estimation (SEE) represents the typical prediction error of the regression model with respect to the reference method. As illustrated in Figure 7, MIF is primarily determined by the subject’s anthropometric characteristics, namely body weight, age, sex, and regression constants. Among these variables, body weight exerts the strongest influence, whereas age introduces a comparatively smaller negative correction. The effect of sex is reflected as an approximately constant offset between the male and female prediction surfaces, without modifying their overall shape or gradient. MIF varies over a wide range within the physiological intervals considered. For all three prediction equations, the largest values are obtained in young individuals with high body weight, whereas the lowest values correspond to elderly subjects with low body weight. The nearly parallel contour lines indicate that the anthropometric contribution changes almost linearly throughout the entire range of age and body weight, reflecting the linear structure of the underlying regression models. This example illustrates that the MIF can contribute as much as, or even more than, the electrical measurement itself. Consequently, a substantial proportion of the estimated body composition may already be determined before any bioimpedance data are incorporated into the prediction model. However, the opposite may also occur: for certain weight–age combinations, the ALSM equation of Kyle et al. predicts that the MIF cancels out, so the resulting ALSM depends solely on the measurement. MIF is not a measure of statistical importance, and it is not a share of explained variance. A term may dominate the predicted value while accounting for little of the variance explained across a cohort. The magnitude of the MIF is furthermore specific to each equation and to the anthropometric range considered, so the values illustrated in Figure 7 should not be extrapolated to other equations or to populations outside that range. 3.2. DXA-Based Calibration and Validation of BPEs DXA plays a major role in the validation of BPEs. Although originally introduced for the assessment of bone mineral density [41], advances in detector technology and image processing have transformed DXA into one of the most widely accepted methods for whole-body composition assessment [42,43]. The physical principles underlying DXA have been comprehensively described by Pietrobelli et al. [44]. The technique is based on irradiating the body with two X-ray photon energies. Because bone mineral, lean soft tissue, and adipose tissue exhibit different X-ray attenuation coefficients, the simultaneous acquisition of low- and high-energy images allows each pixel to be decomposed into its constituent tissue components. By integrating information from all image pixels, DXA provides estimates of whole-body and regional body composition with high spatial resolution and excellent reproducibility. BPEs can be calibrated against the 4C model. The 4C model is generally regarded as the criterion method for in vivo body composition assessment because it minimizes the simplifying assumptions required to estimate body composition. FM, TBW, BMC and proteins are quantified using independent reference techniques, typically isotope dilution for TBW, air-displacement plethysmography or hydrodensitometry for body volume, and DXA for BMC. FM and proteins are subsequently calculated with multicompartment equations [45]. Despite its superior accuracy, the 4C model is impractical for the development of BPEs because it requires multiple independent measurements and time-consuming protocols. In contrast, DXA provides simultaneous estimates of all body compartments in a single examination with high precision and low cost. For this reason, commercial BPEs can be calibrated directly against DXA. DXA depends on expensive instrumentation and entails regular calibration, radiation safety procedures, and trained operators. Although the radiation dose is relatively low—typically between 1 and 10 μSv for a whole-body scan—it nevertheless limits the frequency of repeated measurements, particularly in children, pregnant women, and longitudinal studies requiring frequent follow-up. A complete examination generally requires 5–15 min, substantially longer than a conventional BIA measurement. Furthermore, DXA systems are neither portable nor suitable for bedside or home monitoring. Whole-body DXA measurements typically exhibit coefficients of variation below 1% for total body mass and approximately 1–2% for lean and fat soft tissues [46]. Nevertheless, Park et al. showed that body composition estimates obtained with different DXA systems are not always directly interchangeable, particularly for FM estimation [47]. More recently, several studies have demonstrated the feasibility of cross-calibrating measurements obtained with different DXA platforms [48,49]. Caution must be taken when comparing DXA results from different manufacturers, since terms like lean mass or ALST are used inconsistently [50]. Another limitation concerns the estimation of visceral adipose tissue (VAT). Although modern DXA systems incorporate dedicated algorithms for VAT quantification, these estimates are derived indirectly from two-dimensional projection images rather than true tomographic data. Kaul et al. [51] reported good agreement between DXA-derived VAT estimates and CT, although systematic differences persisted, particularly in individuals with severe obesity or atypical fat distribution. Likewise, Borga et al. identified CT and MRI as the preferred techniques for accurately quantifying VAT, diffuse fat infiltration, and individual muscle volume [52]. More recently, Mancilha et al. found that DXA systematically overestimates VAT compared to MRI [53]. DXA is recommended for assessing FM across many clinical conditions; however, its validity for evaluating FFM in clinical populations remains uncertain [54]. In addition, Basty et al. recently reported that DXA is less accurate than MRI for FFM assessment, particularly in longitudinal studies [55]. Several studies have evaluated the agreement between BIA and DXA. Achamrah et al. compared FM and FFM estimated by both techniques and reported good agreement, particularly in individuals with a body mass index (BMI) between 16 and 18.5 kg/m2. However, for BMI values above 18.5 and below 40 kg/m2, BIA overestimated FFM by 3.38–8.28 kg and underestimated FM by 2.51–5.67 kg relative to DXA [56]. Similarly, Farias et al. demonstrated that simple correction equations can improve the agreement between DXA measurements and commercially available single-frequency leg-to-leg BIA devices [57]. More recently, Zaplatosch et al. assessed the validity of body composition estimates obtained with a multifrequency BIA device in women [58]. FM is generally not predicted directly by BIA but obtained as the difference between body weight and FFM, so the entire error of the FFM estimate is transferred to FM. FFM, in turn, is reached either from TBW, under the assumption that the hydration fraction of FFM remains constant at approximately 73.2%, or by direct regression against DXA in a particular cohort. Both routes are vulnerable in the same direction, because the constant-hydration assumption and the cohorts used for calibration are least representative at the extremes of adiposity. The magnitude of the resulting uncertainty is not negligible: the SEE of the FFM equation was previously shown in Table 2: it is close to 7% of the FFM of an adult of average build, and it corresponds precisely to the equation derived in the cohort with the highest median BMI, 31 kg/m2. Accuracy therefore degrades as adiposity increases, that is, in precisely the population in which body composition is most frequently assessed. This is also the regime in which descriptors that require no regression step offer the clearest advantage: BIVA has been reported as particularly valuable in severe obesity, where conventional equations often provide inaccurate estimates [25]. Commercial analyzers also report parameters whose relationship to the electrical measurement is considerably more remote. VAT and BMC are provided routinely, and BMC obtained with a commercial analyzer has been used as a primary endpoint in recent clinical work [59], yet neither is a quantity that impedance can resolve. Cortical bone is highly resistive and occupies a small fraction of the conductive volume, so a BMC output is almost entirely a product of the regression model rather than of the measurement. VAT is not a distinct electrical compartment either. Any reference used to calibrate such an estimate also carries its own systematic error, as illustrated by the discrepancies between DXA, CT and MRI described above. The practice of the recent literature is consistent with this. In the only primary study of the present corpus in which VAT enters the analysis, bioimpedance supplied body fat percentage alone and VAT was obtained by other means [60]. Likewise, methodological reviews assign visceral fat and bone mass to imaging and cite bioimpedance as a low-cost but imprecise alternative [61]. The standing of these two outputs is therefore best described as uncharacterized rather than quantified: the assessments available are qualitative. 3.3. Boundary Limitations Across the Frequency Spectrum The frequency dependence of tissue impedance provides valuable information on the physiological parameters estimated by bioimpedance analysis. In practice, however, the usable frequency range is constrained by physical phenomena and instrumentation-related parasitics that progressively degrade measurement accuracy. Table 3 summarizes the operating frequencies of representative commercial bioimpedance analyzers. Most of them perform measurements between approximately 1–5 kHz and 1 MHz, with only a few extending into the low-megahertz range. The availability of bioimpedance measurements at very low frequencies would improve the characterization of current flow through the extracellular compartment and, consequently, provide a more reliable estimation of ECW. Why, then, do commercial bioimpedance analyzers seldom operate below approximately 1 kHz? The answer lies in the increasingly dominant contribution of electrode polarization (EP). EP is most commonly associated with electrodes immersed in liquid electrolytes [62,63], but the same electrochemical processes also occur at the interface between a metallic electrode and biological tissue [64]. In bioimpedance measurements, EP originates at the electrode–skin interface. Whereas electrical conduction in metals is mediated by electrons, biological tissues conduct current through the movement of ions. Because these charge carriers cannot be exchanged directly across the interface, ions accumulate near the electrode’s surface, forming an electrical double layer that behaves as a capacitive element. As the excitation frequency decreases, the impedance of this interfacial capacitance increases rapidly and may eventually become several times larger than the impedance of the body segment being measured. Under these conditions, the measured impedance increasingly reflects the electrode–skin interface rather than the tissue itself. EP therefore introduces a systematic error into the measurement. Why not simply remove this contribution through calibration or mathematical compensation? Several equivalent circuit models have been proposed using a constant phase element aiming at compensating for EP effects [65,66]. These approaches can improve measurement accuracy under controlled laboratory conditions. However, their effectiveness is limited in routine clinical practice. EP depends on numerous factors, including electrode material, skin preparation, contact pressure, hydration, and subject-to-subject variability. Consequently, minimizing EP at its source remains the preferred strategy. The tetrapolar electrode configuration adopted in most commercial bioimpedance analyzers substantially reduces the influence of EP by separating the current-injecting and voltage-sensing electrodes, although it cannot eliminate the effect completely. A complementary approach is the use of Ag/AgCl electrodes, which remain the reference standard for bioelectrical measurements owing to their highly reversible electrochemical behavior based on the Ag/AgCl/Cl− equilibrium [67]. This reversible reaction provides a stable half-cell potential and a considerably lower polarization impedance than conventional metallic electrodes. However, their performance relies on a chloride-containing hydrogel to maintain a stable electrode–skin interface, making them susceptible to gel dehydration, skin irritation, and reduced long-term stability. At the opposite end of the spectrum, extending bioimpedance measurements beyond the conventional frequency range could provide additional physiological information by probing tissue dispersion phenomena. Moreover, because the impedance of the electrode–skin interface decreases markedly with increasing frequency, EP becomes progressively less important. However, these benefits are accompanied by new challenges. As the operating frequency approaches the megahertz range, parasitic effects associated with the measurement hardware become increasingly significant [68]. Stray capacitances in cables, connectors, electrode leads and switching circuits introduce amplitude and phase errors, while mutual inductive coupling and electromagnetic interference become more difficult to control. In addition, when the physical dimensions of the measurement system are no longer negligible compared to the effective signal wavelength, the lumped-element approximation ceases to be valid and distributed transmission-line effects must be considered. These phenomena ultimately limit the practical upper frequency that can be used for accurate in vivo bioimpedance measurements. 4. Latest Advances of Interest for Clinical Nutrition 4.1. Electrode Technologies and Bioimpedance Wearables (BWs) In clinical nutrition, bioimpedance analysis is increasingly used not only to estimate body composition but also to monitor changes in hydration, muscle mass and nutritional status during disease progression and nutritional interventions. As nutritional care shifts toward longitudinal patient follow-up, there is an increasing demand for reliable bioimpedance devices. Advances in electrode technologies, flexible electronics and wearable devices have become key drivers. Electrode technology is a critical component of BWs because the electrode–skin interface largely determines measurement reproducibility and long-term stability. It has become an active research field not only within the bioimpedance community but also in related disciplines such as biopotential monitoring [69] and human–machine interaction [70], which face similar challenges regarding electrode–skin impedance and contact reliability. The transition from conventional Ag/AgCl gel electrodes toward dry, flexible and printable electrodes has become one of the main research directions in biopotential wearables and BWs. Materials under investigation include noble metals, conductive polymers, graphene and conductive textiles [71]. Kusche et al. [72] evaluated dry, miniaturized electrodes (177 mm2 surface area) fabricated from various materials. While conventional Ag/AgCl electrodes exhibited the lowest interface impedance, metal (gold-plated and stainless steel) and carbon rubber electrodes achieved comparable performance, demonstrating their suitability for bioimpedance measurements. Nevertheless, maintaining a stable, low-impedance interface during prolonged use and under dynamic conditions remains the main challenge for wearable bioimpedance devices. Recent advances in printed electronics have further expanded the capabilities of BWs. Levit et al. [73] developed flexible skin electrodes fabricated by combining screen-printed carbon with inkjet-printed PEDOT:PSS. Soft, conformable electrode arrays were achieved with low contact impedance and long-term stability. The electrodes enabled reliable four-electrode bioimpedance measurements over several hours, including the continuous monitoring of arterial pulsations and during simultaneous muscle activity. These advances support future BWs for repeated assessment of body composition and hydration in free-living individuals, older adults and patients receiving nutritional support. Despite these advances, continuous monitoring under daily living conditions imposes stringent requirements on electrode design. Sweat accumulation, changes in contact pressure and body motion influence the electrical properties of the electrode–skin interface, leading to motion artifacts and measurement variability. Machine-learning approaches are increasingly being employed to suppress motion artifacts, extract clinically relevant physiological features and integrate bioimpedance measurements with complementary sensors. This improves measurement robustness and enables reliable physiological monitoring during daily activities [74]. In addition, efficient low-power signal-processing algorithms have become an essential component of BWs [75]. Groenendaal et al. [76] highlighted the potential of BWs for chronic disease management, emphasizing continuous home monitoring with low-cost, low-power devices. In clinical nutrition, BWs could facilitate the longitudinal assessment of hydration, lean tissue preservation and nutritional recovery in patients with chronic diseases. Wearable EIT (WEIT) devices provide spatially resolved information on tissue composition. Chen et al. [77] identified WEIT as an emerging technology for continuous physiological monitoring, highlighting the integration of flexible electronics, textile-based sensors and intelligent signal processing. These systems enable real-time monitoring of muscle activity and soft-tissue composition, supporting the longitudinal evaluation of muscle health outside conventional clinical settings. In clinical nutrition, the continuous monitoring of muscle quality, body composition and functional decline could facilitate the early detection of malnutrition and sarcopenia, while improving the follow-up of patients receiving nutritional interventions. WEIT has also shown promise for the non-invasive evaluation of hepatic steatosis. Luo et al. [78] demonstrated that a portable WEIT system can estimate liver fat content, with measurements correlating with MRI-derived liver fat and discriminating between different degrees of hepatic fat accumulation. Although further clinical validation is required, these findings suggest a potential role for BWs in the longitudinal monitoring of metabolic dysfunction-associated steatotic liver disease during nutritional interventions. 4.2. Bioimpedance Imaging Whole-body BIA condenses the electrical behavior of the entire organism into a few global descriptors, which is insensitive to the regional nature of most nutrition-related alterations. Muscle wasting in sarcopenia and cachexia progresses unevenly across limbs and trunk. Fluid overload accumulates preferentially in dependent regions, and refeeding-associated fluid shifts are inherently local. Imaging-based bioimpedance resolves where conductivity changes occur rather than only how large they are, while remaining non-invasive, radiation-free, inexpensive, and repeatable at the bedside. These properties suit the longitudinal monitoring of nutritional interventions, where the relevant signal is a change in regional muscle quality or fluid distribution rather than a single cross-sectional estimate. EIT and Electrical Impedance Myography (EIM) extend conventional bioimpedance analysis by providing regional information about tissue function rather than whole-body electrical properties. EIT reconstructs the internal conductivity distribution of biological tissues from voltage measurements acquired using multiple surface electrodes. Unlike conventional anatomical imaging modalities, EIT provides continuous bedside monitoring with high temporal resolution, making it particularly suitable for functional assessment rather than structural diagnosis. EIT does not provide detailed anatomical information but continuously visualizes functional changes in tissue conductivity. The growing clinical adoption of EIT was further supported by the international consensus statement of the TREND group [79], which established standardized examination protocols, unified terminology, data analysis methods, and recommendations for the clinical interpretation of thoracic EIT. Youssef Baby et al. [80] summarized the evolution of both EIT and EIM technologies, highlighting the growing role of digital signal processing, machine learning, and integrated electronic systems in improving image quality and diagnostic performance. Recent progress in chest EIT has been reviewed by Frerichs et al. [81], who describe its increasing use for the bedside monitoring of pulmonary perfusion and patient-specific optimization of mechanical ventilation. Their review illustrates how technological advances and improved reconstruction algorithms have accelerated the clinical adoption of EIT in intensive care medicine, particularly for patients requiring mechanical ventilation. Beyond respiratory monitoring, EIT is progressively expanding toward cardiovascular assessment, neurocritical care and gastrointestinal monitoring [80]. Although limitations in spatial resolution remain, ongoing improvements in reconstruction algorithms and hardware design continue to broaden its diagnostic potential. EIM focuses on the localized assessment of skeletal muscle, providing quantitative information on muscle composition, membrane integrity and neuromuscular function. Sonbas–Cobb et al. [82] recently demonstrated that combining multifrequency EIM with machine-learning algorithms improves the screening of neuromuscular disorders compared to conventional impedance metrics alone. Their results illustrate the growing integration of artificial intelligence into impedance-based diagnostic systems and support the further evaluation of EIM as a candidate objective biomarker for neuromuscular disease detection. Current research is also extending impedance-based muscle assessment beyond clinical diagnosis toward rehabilitation engineering and human–machine interaction. Park et al. [70] describe how EIM can be combined with biopotential recordings to improve prosthetic control, rehabilitation monitoring and wearable biomechanical interfaces. Hybrid impedance–EMG sensing decodes movement intention more robustly than EMG alone, particularly during dynamic contractions and when conventional EMG signals become unstable. This illustrates the increasing convergence between bioimpedance imaging, intelligent sensing, and digital healthcare. Bioimpedance-based imaging technologies grapple with intrinsic physical limitations. Spatial resolution remains notably low compared to structural modalities such as CT or MRI. Furthermore, image fidelity strictly depends on the exact placement of electrodes and is heavily affected by the anatomical shape variability of each patient’s chest or limbs. 4.3. Toward Standardized Bioelectrical Biomarkers Despite its widespread use for body composition assessment in healthy populations, BIA has yet to achieve broad clinical acceptance. Reference values for BIA-derived parameters have been proposed to aid clinical interpretation, such as those of Barbosa et al. for PhA [83]. However, in its 2020 evidence-based clinical guideline, the American Society for Parenteral and Enteral Nutrition (ASPEN) reached a different conclusion. In its view, the available evidence was insufficient to recommend BIA as a reference method for body composition assessment in hospitalized patients or in individuals with acute or chronic diseases [54]. This conclusion was based on four main limitations: (i) limited validation of BIA methods in clinical populations, (ii) substantial variability among commercial devices, (iii) the influence of disease-related factors like inflammation, and (iv) insufficient high-quality evidence. This last point refers to the lack of adequately powered studies using appropriate reference methods and clinically relevant outcomes. ASPEN also set priorities for future research: validating bioimpedance methods in clinical populations; standardizing terminology and measurement protocols. Further priorities were investigating raw bioelectrical variables as clinically meaningful biomarkers. It also called for establishing disease-specific reference values and diagnostic cutoff points, and adopting more rigorous study designs with appropriate reference methods. Interest in raw bioelectrical variables has a long clinical history, including early applications of PhA in nutritional and oncological assessment during the 1990s. Since 2020, the field has received renewed attention, centered on standardization, reference values and outcome-based validation. PhA and BIVA emerged as the most promising candidates for clinical implementation. Oncology illustrates this trajectory. Early studies used BIA to assess malnutrition and body cell mass in patients with cancer, and to monitor body composition during nutritional support [84,85]. The introduction of BIVA in 1994 allowed the impedance vector to be interpreted directly [23], and in 2000 patients with lung cancer were shown to occupy a distinct vector position relative to matched healthy controls [86]. By 2004, PhA had been reported as a prognostic indicator in advanced pancreatic cancer [87]. Outcome-based validation against survival developed mainly from 2020 onward. Unlike conventional estimates of FM, FFM or TBW, PhA is a direct electrical parameter obtained from the relationship between resistance and reactance, without relying on regression equations or body composition models. Consequently, it is less affected by the methodological variability associated with proprietary algorithms and provides information that more directly reflects cellular integrity, hydration status and BCM. Its absolute value nevertheless remains dependent on instrumentation, electrode configuration and measurement conditions. Likewise, BIVA enables the interpretation of the measured impedance vector itself, avoiding the uncertainties associated with predictive equations while providing an integrated assessment of hydration and soft-tissue characteristics. Large-scale systematic reviews have since compiled age- and sex-specific reference values for PhA, representing an important step toward clinical standardization [88]. At the same time, comprehensive reviews highlighted the complementary roles of PhA and BIVA for assessing muscle quality, hydration status and functional performance, extending their applications well beyond conventional body composition analysis [89]. Growing evidence also demonstrated the prognostic value of PhA in different clinical settings. Prospective studies showed that low PhA predicts future functional disability in older adults [90], while systematic reviews confirmed consistently reduced PhA values in patients with cardiovascular disease, supporting its role as a prognostic biomarker for cardiovascular risk stratification [91]. In parallel, considerable efforts were devoted to establishing the physiological basis underlying these observations. Comprehensive reviews consolidated the interpretation of PhA as an indicator of cellular hydration, membrane integrity and BCM [92], while also emphasizing the need for standardized measurement protocols and clinical reference values to facilitate routine implementation [93]. Experimental studies further demonstrated that segmental PhA correlates with local muscle contractile properties in healthy individuals [94] and longitudinal investigations showed that increases in segmental PhA accompany improvements in muscle quality during post-stroke rehabilitation [95]. More recently, long-term prospective studies have shown that segmental PhA combined with the ECW/ICW ratio predicts future functional disability in older adults, introducing regional bioelectrical markers for the early detection of frailty [96]. Similarly, BIVA has proven useful for monitoring body composition changes during stroke rehabilitation, providing clinically relevant information that complements conventional anthropometric and body composition measurements [97]. The two descriptors are not, however, equally consolidated. Age- and sex-specific reference values, prognostic evidence and segmental applications have accumulated for PhA, whereas for BIVA the corresponding reference framework remains largely limited to the tolerance ellipses defined at its introduction [24]. Extending to BIVA, the reference material already available for PhA is therefore part of the standardization effort that ASPEN identified as a research priority. A methodological refinement of particular relevance in this context is specific BIVA (sBIVA), in which resistance and reactance are normalized by an estimate of the geometry of the conductive volume instead of by height alone [98]. Because the classical vector depends strongly on body size, differences between individuals or groups may reflect body geometry rather than tissue composition. The specific normalization attenuates this dependence and improves the separation between fluid distribution and soft-tissue quality, an advantage that is most evident in obesity, where the classical vector is most affected [99]. sBIVA is conceptually aligned with the argument developed in this review: like PhA, it is obtained from the measured impedance and therefore avoids prediction-equation calibration bias, while additionally addressing the size dependence that limits the classical vector. Its interpretation nevertheless remains anchored to the tolerance ellipses of the population in which they were derived, so vector positions referred to different reference samples are not directly comparable [24]. Recent studies have also expanded the biological interpretation of PhA beyond its purely electrical definition. Mechanistic reviews have linked PhA to growth, body composition and cellular physiology across the lifespan [100]. Prospective cohort studies have demonstrated that raw bioelectrical variables, particularly PhA, can monitor body composition changes during growth without relying on prediction equations, supporting their use for longitudinal follow-up [101]. More broadly, recent reviews have summarized the growing number of clinical applications of BIA in nutrition, fluid management, oncology, cardiology and intensive care. These reviews highlight the ongoing transition from model-dependent body composition estimates toward physiologically meaningful bioelectrical biomarkers [102]. This evidence is, however, largely observational. The associations reported between raw bioelectrical parameters and clinical or nutritional outcomes are consistent across settings. However, they do not establish a causal relationship in themselves, nor do they demonstrate that management guided by these parameters improves patient outcomes. 5. Bibliometric Analysis of the Literature (2020–2026) 5.1. Article Selection and Citation Impact The methodology and nutrition equations described in Section 1 were reused for this analysis, but without the restriction to the Article document type. Consensus documents, clinical practice guidelines and reviews are themselves objects of study in this section rather than duplicated evidence. This search was run on 28 July 2026. It was restricted to the 2020–2026 window and retrieved 5625 records, exported directly from the Web of Science interface and provided as Supplementary Materials. To identify influential studies while maintaining a manageable dataset for qualitative analysis, we retained only publications averaging at least 20 citations per year, [54,59,60,61,88,89,92,93,103,104,105,106,107,108,109,110,111,112,113,114,115,116,117,118,119,120,121,122,123,124,125,126,127,128,129,130,131,132,133,134,135,136,137,138,139,140,141,142,143,144,145,146,147,148,149,150,151,152,153,154,155,156,157,158]. The resulting dataset was reviewed to classify publication type, study design, clinical application, and the specific role played by bioimpedance within each study. Large language models were used to cross-check the summarized contents tabulated in Appendix A against the corresponding full texts. Articles were first classified according to the “Document Type” field assigned by WoS, yielding two categories: review articles and non-review articles. Following a manual analysis of the article, it became apparent that several articles labeled as “non-review” by WoS were, in substance, consensus guidance documents, clinical practice guidelines, or systematic reviews with meta-analysis rather than studies reporting primary data. These were therefore reclassified and grouped together with the review articles for the purposes of this analysis. This process yielded a final classification of 37 review-type articles and 27 non-review articles. Although the number of highly cited publications decreases in the most recent publication years, this pattern should not be interpreted as evidence of declining research activity. The total volume of publications continues to increase across all body composition modalities, whereas the apparent reduction in highly cited papers is consistent with citation maturation effects and the shorter time available for recently published articles to accumulate citations. Table 4 summarizes citations per year (Cites/year) by group and by the role assigned to BIA within each document. Review-type articles accumulated more citations per year than non-review articles (Mann–Whitney U = 698.5, p = 0.007), and articles assigning BIA a complementary role received more citations than those treating it as the primary measurement technique (U = 322.5, p = 0.030). Both distributions were strongly right-skewed, driven by a small number of very highly cited consensus documents and guidelines, so the median is the more representative summary statistic. Part of this difference is a generic feature of citation behavior rather than a property of the bioimpedance literature, since review articles are cited more frequently than original research across virtually all disciplines. A selection threshold based on citations per year therefore favors synthesis over primary research by construction, and the resulting ratio of 37 review-type to 27 non-review articles should be read as reflecting both this structural effect and a genuine predominance of synthesis within the field. 5.2. Reviews, Guidelines, and Consensus Documents Table A1 (Appendix A) summarizes the reviewed articles. Narrative, overview and critical reviews comprise the largest share of the literature, 59%. Articles employing formal systematic evidence synthesis, such as scoping reviews or systematic reviews with or without meta-analysis, account for 22%. Consensus statements and clinical practice guidelines constitute the remaining 19%. This distribution reveals that the most-cited BIA literature in clinical nutrition rests predominantly on expert interpretation rather than pooled quantitative synthesis. The chronological distribution is comparatively even, indicating sustained review interest across the period. Analysis of the articles showed that sarcopenia was the most frequently addressed clinical area within the review group. These covered diagnostic criteria and screening algorithms [103,118,123,126,127], as well as prevalence estimates in the general population and in specific settings such as dialysis [115,117,132]. They also addressed underlying molecular and pathogenic mechanisms [126,134], the use of raw bioelectrical parameters, particularly PhA, in this population [109], and the interplay with cardiovascular aging [133]. A consensus document on sarcopenic obesity outlines research priorities for this overlapping phenotype [124]. The second-largest group is methodological rather than disease-oriented, focusing on body-composition assessment itself [54,61,88,92,93,104,107,110,119,128,131]. Four of these studies are specifically devoted to PhA [88,92,93,128]. This confirms the shift toward raw bioelectrical biomarkers described above. Reviews devoted to sports and athletic populations [108,112], together with a systematic review addressing methodological aspects of BIA in athletes [119], malnutrition assessment, muscle-health knowledge-translation resources [116,120,121], and oncology [113,114,122] followed. The remaining areas were represented by one or two articles each: obesity by two, one addressing metabolic phenotypes and biomarkers [105] and the other the adequacy of BMI-based diagnosis [125]. Chronic kidney disease [106], critical illness [111], cystic fibrosis [129], and menopause-related nutrition [130] were represented by a single article. Fluid overload and altered hydration are precisely the conditions in which prediction-equation-based estimates of body composition are least reliable. The scarce high-impact review coverage of critical illness and renal disease is therefore notable, and marks these settings as the clearest gaps in the current evidence. A further distinction emerges from the role assigned to BIA within each document. Only about one quarter of the reviewed documents treat BIA as their primary object of study [88,92,93,108,109,111,119,122,128,131]. In the rest, it appears as a complementary technique, typically listed as one acceptable muscle-mass or FM method alongside DXA, CT, or MRI. Raw bioelectrical descriptors are discussed in 21 of the 37 documents and BIVA in 11. Five documents identify device-, population- or equation-specific variability as a central limitation [93,107,110,116,120], so the reporting gap is flagged by the same literature that leaves it open. DXA acts as the comparator of reference in 18 entries, ahead of CT in ten and MRI in three, which frames BIA as a technique benchmarked against imaging rather than against a criterion measure of body composition. Where numerical PhA thresholds are offered, they do not converge. Reported values include 3.55–5.05° for sarcopenia screening [109], <5.58° for nutritional risk in critical illness [111], 5.95° for 5-year survival in advanced head-and-neck cancer [122], and <5.10° in men and <4.73° in women for low muscle radiodensity [128]. The only value repeated anywhere in Table A1 and Table A2, 5.95°, is derived for two unrelated endpoints, which illustrates that numerical agreement between studies can be coincidental rather than substantive. 5.3. Studies Reporting Primary Data Table A2 (Appendix B) presents 27 non-review articles. By study design, cross-sectional studies were the most frequent (n = 11, 41%), followed by prospective cohort studies (n = 8, 30%), randomized controlled trials (n = 4, 15%), other observational designs (n = 2, 7%), and methodological or development studies (n = 2, 7%). One of the four randomized trials used cluster randomization at the practice level [144]. One of the eight prospective cohorts is a post hoc analysis of a previous cohort [141], and one of the cross-sectional studies is a cross-sectional analysis nested within a prospective cohort [158]. Designs capable of supporting temporal or causal inference—prospective cohorts and randomized trials—thus accounted for 44% of this group, a stronger evidentiary profile than that of the review literature, in which formal evidence synthesis represented only 22% of the articles. Sarcopenia and frailty formed the dominant clinical focus. These spanned endocrine and biomarker correlates [59], mortality, frailty and physical-performance outcomes [137,140], cognition and incident dementia [139], heart failure [141], and validation of BIA against DXA [142]. Other topics included body composition in pre-frail older adults [143], PhA as a muscle-quality index [146], gut–microbiome associations [148], cancer-related sarcopenia and malnutrition [149,155], exercise and multidomain interventions [150,153], and sarcopenic obesity [154]. Obesity and adiposity constituted the second cluster, ranging from ketogenic dietary intervention [135] and school-based pediatric programs [144] to anthropometric and adiposity indices [138,152]. These indices were associated with cognition [60], cardiovascular disease [157], and heart failure [158]. The remaining clinical areas included athletic populations [89,136] and hepatic disease [147]. Two additional articles focused primarily on research infrastructure and methodology: automated CT segmentation [145], wearable impedance biosensors [156], and the profile of a large population cohort [151]. The role assigned to BIA as a body-composition measurement technique was more central here than in the review literature: it was the primary measurement technique in two thirds of the studies and of marginal relevance in only two. In those two, adiposity was assessed by DXA and BIA figures were quoted only for comparison with results from other authors [152], or the impedance measurement was confined to the electrode–skin interface of a novel wearable biosensor rather than to body composition [156]. The contrast is informative: whereas high-impact reviews and guidelines tend to list BIA as one option among several accepted reference methods, the original research reaching comparable citation levels is largely built on BIA as the measurement instrument of record. This asymmetry in the role assigned to BIA between the two groups was confirmed by contingency-table analysis. Table 5 shows the cross-tabulation together with the associated χ2 statistics. An article was therefore roughly five times more likely to treat BIA as the primary measurement technique if it reported primary data than if it was a review or guideline document. The same instrument-level reading of Table A2 reveals an asymmetry with the review literature. PhA is used in only five of the 27 original studies [89,143,146,150,156], and in one of them the impedance measurement concerns the electrode–skin interface of a wearable sensor rather than the body [156]; equation-derived muscle-mass indices underpin fourteen studies. The most-cited primary-data studies continue to rely predominantly on model-dependent body-composition estimates, despite the growing use of raw bioelectrical descriptors in a smaller subset of studies. The transition toward raw descriptors is therefore more evident in reviews and consensus documents than in the current body of highly cited primary research. Device provenance is also highly concentrated: the analyzers used come from only two manufacturers, eight of them InBody platforms [59,137,142,143,148,149,151,154] and four Tanita [140,147,155,157]. Because both rely on proprietary, device-specific equations, a substantial share of the high-impact evidence is anchored to two commercial calibrations. Scale and provenance are equally uneven. Sample sizes are traceable for 23 of the 27 studies and span four orders of magnitude, from 50 participants in an exercise intervention [150] to 468,333 in a UK Biobank analysis [157], with a median of 1417. Three population-based resources contribute about 96% of the pooled participants: two analyses of the same UK Biobank cohort [147,157] and one Norwegian population survey [151]. Most of the statistical weight of this group therefore comes from secondary analyses of pre-existing cohorts rather than from purpose-designed BIA studies. Where the study population can be located, seven were conducted in East Asian populations [59,137,138,142,146,148,154], against two in the United Kingdom [147,157], one in Norway [151] and one in Ireland [153]. The sarcopenia evidence therefore derives largely from populations in which the Asian Working Group for Sarcopenia (AWGS) criteria and cutoffs apply, whereas the large-cohort adiposity evidence is European. This constrains the transferability of population- and device-specific thresholds between the two clusters. Hard endpoints also remain scarce: mortality or survival is examined in four studies [137,141,149,155] and incident disease in three [139,147,157], with the remainder reporting cross-sectional or short-term associations. 5.4. Shared Findings and Evidence Gaps The chronological distribution peaked in 2021 (nine studies), followed by 2022 (six), 2023 (five), 2020 (four), 2024 (two), and 2025 (one), with the decline in the final years again reflecting the citations-per-year threshold. Table 6 regroups the 64 articles into seven broad clinical areas by group. The obesity/adiposity literature was disproportionately drawn from large-cohort primary-data studies rather than reviews (Fisher exact OR = 6.12, p = 0.029), consistent with its dependence on population-scale resources such as UK Biobank. The methodology and PhA literature, by contrast, was almost exclusively contained in the review literature. Taken together, these findings suggest that the recent high-impact BIA literature follows a characteristic translational trajectory. Methodological development is largely driven by review articles and consensus documents, whereas primary investigations focus on validating the clinical usefulness of already established body-composition metrics within a limited number of disease areas, particularly sarcopenia and obesity. The adoption of raw bioelectrical biomarkers in clinical research remains comparatively limited, despite their growing prominence in methodological reviews. At the same time, several settings in which reliable body-composition assessment is most needed, notably critical illness, advanced renal disease and acute care, remain substantially underrepresented. These patterns indicate that future research priorities should extend beyond incremental validation studies toward broader clinical implementation and the standardization of raw bioelectrical parameters across diverse patient populations. These barriers are not confined to instrumentation. The standardization of electrode placement and of pre-measurement conditions, population-specific reference values, transparent analytical pipelines and clinically interpretable thresholds are equally necessary before raw bioelectrical parameters can be used in routine practice. 5.5. Are the Identified Gaps Being Addressed? The bibliometric analysis presented in Section 5.1, Section 5.2, Section 5.3 and Section 5.4 selects for citation impact and therefore cannot describe the most recent literature, in which citations have not yet had time to accumulate. To establish whether the gaps identified above are being addressed, a complementary search was performed on 30 July 2026 using the same technique and nutrition equations, restricted to the Article document type and to the 2025–2026 publication window, and without any citation threshold. The search retrieved 1672 records, of which 14 were discarded as non-human studies belonging to animal or food science and three because their print year fell outside the window, leaving 1655 articles. Each record was assigned to a single clinical area from its full text, using the categories of Table 6 with two refinements. First, the residual other clinical category was resolved into critical/acute care and renal disease, with the two settings identified as underrepresented in Section 5. Second, records spanning more than one area were assigned by a fixed priority order: critical/acute care, renal disease, oncology, malnutrition, sports, sarcopenia/frailty, obesity/adiposity, methodology, and other. The same rule was applied retrospectively to the 64 articles of Section 5, so that the two distributions are directly comparable. Table 7 reports the resulting distributions side by side. Both distributions are expressed as shares of their own corpus. They therefore describe how bioimpedance research is distributed across clinical areas rather than how intensively the technique is used within each area: an area that is large in the nutrition literature as a whole will tend to contribute a large share irrespective of its adoption of bioimpedance. Normalizing each area to its own total output would require field-level denominators outside the scope of the present search. The comparisons that follow are therefore drawn between the two corpora within a given area, and the shares are not read as a ranking of uptake between areas. Critical and acute care remains the setting with the smallest share of the corpus: 22 of the 1655 recent articles (1.3%), essentially unchanged from the single article (1.6%), were retained in the most-cited corpus. This is the clearest outcome of the comparison, and it is robust to the differences between the two datasets, since the proportion is equally marginal under either selection criterion. Renal disease follows the opposite trajectory, rising from two articles (3.1%) to 138 (8.3%), which is the largest proportional increase in any area. The two settings that Section 5.1, Section 5.2, Section 5.3 and Section 5.4 identified jointly as underrepresented have therefore diverged: the share of the corpus occupied by renal disease is expanding, whereas that of intensive and acute care is not. Within the areas that were already well-covered, obesity and adiposity has become the largest single category of the recent corpus (35.4%), ahead of sarcopenia and frailty (24.4%), reversing the order observed in the most-cited corpus. Because the two areas differ in the overall volume of nutrition literature they generate, this ordering locates where recent bioimpedance articles are concentrated and does not by itself indicate a higher uptake of the technique in obesity than in sarcopenia. The contraction of the methodological category, from 17.2% to 6.0%, should not be read as a decline in methodological activity. The corpus of Section 5.1, Section 5.2, Section 5.3 and Section 5.4 was selected by citations per year without restriction of document type and therefore concentrates reviews, guidelines and consensus documents, which is precisely where methodological work is published, whereas the present corpus contains original articles only. The comparison is accordingly informative for clinical areas, where both corpora indicate where research effort is directed, and uninformative for categories whose size depends on document type. Raw bioelectrical parameters are explicitly reported in 306 of the 1655 recent articles (18.5%). Their use concentrates precisely where fluid status is least stable: 45% of the critical and acute care articles and 34% of the malnutrition articles report PhA or vector analysis, against 4% of the obesity articles. Unlike the column shares of Table 7, these are proportions computed within each area and are therefore independent of the relative size of the areas. The direction is unambiguous: the adoption of raw bioelectrical descriptors, described as comparatively limited in the most-cited literature, is progressing, and it is progressing fastest in the settings where fluid redistribution most rapidly invalidates regression-based estimates. 6. Concluding Remarks The scope of BIA has widened considerably since its early use as an estimator of TBW. Its value nevertheless rests on an indirect measurement: what is recorded is the electrical response of biological tissues, which is then converted into physiological variables through predictive models. The accuracy of the resulting compartment estimates therefore depends as much on the reference method used for calibration and on the population from which the equation was derived as on the precision of the electrical measurement itself. The cumulative publication record places the maturity of BIA technologies in perspective. Weighted logistic models reproduce the growth of five of the most widely used body composition techniques with coefficients of determination at or above 0.999. The weighted logistic analyses should be interpreted as descriptive, model-dependent analyses rather than as established forecasts of future publication activity. Very high R2 values are not in themselves evidence of predictive validity when fitting smooth cumulative publication counts. However, a logistic curve accumulates half of its asymptote at the inflection point, which falls beyond the observation window for BIA. Under the fitted model, it is likely that fewer than half of the BIA articles it projects have so far been published. Interpreting a bioimpedance measurement requires attention to both tissue electrical properties and instrumental limits. EP, skin impedance, parasitic capacitances, distributed electromagnetic effects, and frequency-dependent dielectric phenomena all influence the measured impedance and bound the physiological information that can be extracted. This is why extending the measurement toward both lower and higher frequencies remains technically challenging. A second constraint lies in the equations rather than in the hardware. Anthropometric variables frequently contribute as much as, or more than, the electrical measurement, and MIF analysis makes that contribution explicit. This explains why instruments that record nearly identical impedances can return different body composition estimates. The consequence is not uniform across patients. FM is obtained by subtracting FFM from body weight, so the error of the FFM estimate passes to FM in full. Both the constant-hydration assumption and the cohorts used for calibration are least representative at high BMI, so accuracy is lowest at the upper end of the BMI range. We can conclude that greater transparency regarding proprietary prediction equations would facilitate independent validation and improve inter-device comparability. It would also clarify the standing of outputs such as VAT and BMC, which commercial devices report routinely and clinical studies already use as endpoints, but for which no quantitative validation was identified in this review. Considerable progress has also been achieved in the direct interpretation of raw bioelectrical variables, particularly PhA and BIVA. These parameters provide clinically relevant information without relying on regression equations and are therefore not subject to the calibration bias that grows with adiposity. Together with advances in multifrequency measurements and segmental analysis, these developments are progressively shifting the field from purely empirical body composition estimation toward a more mechanistic understanding of tissue physiology. Beyond body composition assessment, the electrode technologies, BWs and bioimpedance imaging mark the most likely direction of expansion. The regional distribution of change carries clinical information of its own: muscle wasting is uneven across limbs. Trunk and fluid overload accumulates in dependent regions. The weight that conventional imaging has acquired in nutrition research is evidence of that interest. Bioimpedance imaging brings regional resolution within reach of a radiation-free measurement that can be repeated at the bedside. It is therefore of direct use to sarcopenia and cachexia, to fluid management in dialysis and critical care, to refeeding in malnutrition, and to neuromuscular rehabilitation. Its present limit is spatial resolution. Closing that gap would be consequential. DXA needs fixed, high-cost instrumentation, radiation safety procedures and trained operators. Its dose is low, but it still limits how often a measurement can be repeated, particularly in children, in pregnancy and in longitudinal follow-up. A bioimpedance measurement takes a fraction of the time. It can be made at the bedside or at home, and it can be repeated daily. Bioimpedance may offer limited accuracy for some body-composition parameters, yet it provides major advantages for research. With BIA, sample size is no longer the primary constraint: large population cohorts already incorporate bioimpedance measurements, and individual studies can include hundreds of thousands of participants, a scale that is difficult to achieve with conventional imaging techniques. Prediction equations may introduce systematic errors. But these errors are largely canceled when they remain stable within individuals and the outcome of interest is change rather than absolute value. Segmental PhA and BIVA are already used this way in dialysis, critical care, nutrition, rehabilitation, and sport. The most-cited literature of the recent years converges on sarcopenia and, secondarily, on obesity and adiposity, and rests predominantly on narrative and consensus documents rather than on pooled quantitative synthesis. Critical illness and renal disease, where fluid redistribution is fastest and least predictable, occupy the smallest share of that corpus. The primary evidence is also concentrated on the analyzers of only two manufacturers, both relying on proprietary equations, so a large part of it rests on two commercial calibrations. The literature published in 2025 and 2026 shows that this picture is already changing unevenly. Renal disease has risen from 3% to 8% of the bioimpedance corpus, and obesity has displaced sarcopenia as the largest single area of application within it, so the technique is now applied most intensively where the equations are least accurate. Raw bioelectrical parameters now appear in nearly one fifth of recent articles, and most often in critical care and malnutrition, that is, precisely where fluid status is least stable. Critical and acute care, however, remains marginal. Two limitations qualify these conclusions. First, the statements concerning measurement principles, the dependence of compartment estimates on prediction equations, and the clinical use of raw bioelectrical parameters derive from the published evidence reviewed here. The statements concerning the maturity, distribution and expected trajectory of each technique derive instead from the bibliometric analysis. They describe research activity only and should not be read as evidence of clinical validity. Second, the bibliometric results depend on the database, the document type and the search terminology used. Web of Science indexes journals selectively and its coverage of early work is less complete, so publication counts reflect indexed visibility rather than historical priority. Applications described with heterogeneous terminology may also be underrepresented. Reproducing the analysis in other databases, or with different search equations, would be expected to shift the absolute counts. It is unlikely, however, to alter the relative ordering of the five techniques.