Greenness profile assessment of a novel pioneer approach for voltametric sensing of cloperastine fendizoate in a pharmaceutical dosage form Authors: Maha Mahmoud Abou El‑Alamin, Ola Abd Elkhalek, Marwa M. Azab Published: 2026 ABSTRACT A novel sustainable analytical approach based on an electrochemical sensor has been fabricated for the quantification of Cloperastine Fendizoate (CLP), which has recently been employed in the management of COVID-19. A carbon paste electrode modified with 2% multiwalled carbon nanotubes and 1% chitosan (2%MWCNT/1%Chitosan/CPE) was fabricated to implement the proposed procedure. The obtained results demonstrated irreversible oxidation behavior and diffusion-controlled oxidation process of CLP, characterized by a well-defined anodic peak (E/V = 0.8) in 0.2 M phosphate buffer at pH 8 utilising differential pulse voltammetry (DPV). A linear correlation between current signal and CLP concentration was obtained across the range of 7.500 × 10 7- 6.00 × 10− 5 mol L⁻¹ (0.5–40 µg mL− 1), The practically calculated limits of detection and quantification were 1.11 × 10⁻⁷ mol L⁻¹ and 3.70 × 10⁻⁷ mol L⁻¹, respectively. The procedure achieved excellent accuracy and precision, with a mean recovery ±%RSD of 100.95 ± 1.75. The suggested procedure was successfully extended for the quantification of CLP in its dosage formulation. Validation was performed in accordance with ICH guidelines, and the findings demonstrated good agreement with those taken utilizing a previously validated approach. Furthermore, the eco-friendliness of the developed approach was investigated utilising both the Green Analytical Procedure Index (GAPI), the Multi-color Assessment (MA) technique, and the Analytical Greenness metric (AGREE), confirming that the proposed approach exhibits superior greenness compared to reported techniques. Introduction Cloperastin fendizoate (CLP) is chemically known as 1-[2-[(4-chlorophenyl)-phenylmethoxy]ethyl]piperidine; 2-(4-hydroxy-3-phenylbenzoyl)benzoic acid, as shown in Fig. 1. It is a centrally acting antitussive reagent, primarily prescribed to manage the non-productive (dry) cough associated with conditions such as bronchitis, common cold, or certain allergic reactions. More recently, it has also been utilized to alleviate cough-related symptoms in patients with COVID-19 [1, 2]. Several analytical techniques have been employed for CLP determination, including UV spectroscopy [3], spectrofluorimetry [4], RP-HPLC [5, 6], and HPLC associated with tandem mass spectrometry [7]. However, chromatographic methods present certain limitations; they typically require multiple steps of sample preparation, separation, and detection, which can be labor-intensive and time-consuming, especially when dealing with complex mixtures. Moreover, these techniques often consume large volumes of samples and organic solvents, which are expensive, potentially hazardous, and environmentally unfriendly. Additionally, their implementation depends on sophisticated and costly detection equipment. To the best of our knowledge, no voltammetric method has been previously reported for the determination of CLP. Voltammetric techniques have attracted considerable attention in pharmaceutical analysis due to their high sensitivity, simplicity, low cost, rapid response, and minimal sample preparation requirements, making them well-suited for the determination of electroactive compounds. Accordingly, the present study introduces the first voltammetric approach for the sensitive and selective analysis of CLP. The proposed method is based on a modified carbon paste electrode with nanomaterials, which offers several advantages, including enhanced electron transfer kinetics, increased effective surface area, and improved sensitivity and selectivity toward the target analyte, thereby improving the overall analytical performance of the developed method. Additionally, the application of green assessment is crucial, as it allows the evaluation of the environmental friendliness, sustainability, and safety of the analytical method, guiding researchers to develop methods that minimize chemical waste, energy consumption, and exposure to hazardous reagents. Therefore, the proposed method was evaluated using established green assessment techniques, such as Analytical Greenness Metric (AGREE), Complex Green Analytical Procedure Index (Complex-GAPI), and the Analytical Eco-Scale, to ensure minimal environmental impact. Fig. 1Chemical structure of Cloperastin fendizoate Chemical structure of Cloperastin fendizoate Materials and methods Regents and materials CLP drug substance with purity (99.6%) and the pharmaceutical dosage form, Notussil® 35.4 mg/5 mL suspension, were supplied kindly by Chemipharm pharmaceutical company.Graphite powder, multi-walled carbon nanotubes (MWCNTs, > 95% carbon, outer diameter × length: 6–9 nm × 5 mm), silicon dioxide nanoparticles (SiO₂ NPs, < 50 nm particle size, 99.5% trace metals basis), zeolite (< 20 μm), graphene nanoplatelets, and chitosan were sourced from Sigma-Aldrich (Germany).Hydrochloric acid (HCl) and Sodium hydroxide (NaOH) were also obtained from Sigma-Aldrich (Germany).Paraffin oil and diethyl ether (Sigma-Aldrich, Germany) were utilized for carbon paste preparation.Supporting electrolytes included Britton–Robinson (BR) buffer (0.04 M, pH 2–11), phosphate buffer (0.2 M), and borate buffer (0.2 M, pH 8). All buffer systems were prepared following the procedures outlined in the United States Pharmacopeia [8].Double-distilled water was utilized throughout the approach. All trials were performed at ambient laboratory temperature (25 °C). CLP drug substance with purity (99.6%) and the pharmaceutical dosage form, Notussil® 35.4 mg/5 mL suspension, were supplied kindly by Chemipharm pharmaceutical company. CLP drug substance with purity (99.6%) and the pharmaceutical dosage form, Notussil® 35.4 mg/5 mL suspension, were supplied kindly by Chemipharm pharmaceutical company. Graphite powder, multi-walled carbon nanotubes (MWCNTs, > 95% carbon, outer diameter × length: 6–9 nm × 5 mm), silicon dioxide nanoparticles (SiO₂ NPs, < 50 nm particle size, 99.5% trace metals basis), zeolite (< 20 μm), graphene nanoplatelets, and chitosan were sourced from Sigma-Aldrich (Germany). Graphite powder, multi-walled carbon nanotubes (MWCNTs, > 95% carbon, outer diameter × length: 6–9 nm × 5 mm), silicon dioxide nanoparticles (SiO₂ NPs, < 50 nm particle size, 99.5% trace metals basis), zeolite (< 20 μm), graphene nanoplatelets, and chitosan were sourced from Sigma-Aldrich (Germany). Hydrochloric acid (HCl) and Sodium hydroxide (NaOH) were also obtained from Sigma-Aldrich (Germany). Hydrochloric acid (HCl) and Sodium hydroxide (NaOH) were also obtained from Sigma-Aldrich (Germany). Paraffin oil and diethyl ether (Sigma-Aldrich, Germany) were utilized for carbon paste preparation. Paraffin oil and diethyl ether (Sigma-Aldrich, Germany) were utilized for carbon paste preparation. Supporting electrolytes included Britton–Robinson (BR) buffer (0.04 M, pH 2–11), phosphate buffer (0.2 M), and borate buffer (0.2 M, pH 8). All buffer systems were prepared following the procedures outlined in the United States Pharmacopeia [8]. Supporting electrolytes included Britton–Robinson (BR) buffer (0.04 M, pH 2–11), phosphate buffer (0.2 M), and borate buffer (0.2 M, pH 8). All buffer systems were prepared following the procedures outlined in the United States Pharmacopeia [8]. Double-distilled water was utilized throughout the approach. All trials were performed at ambient laboratory temperature (25 °C). Double-distilled water was utilized throughout the approach. All trials were performed at ambient laboratory temperature (25 °C). Instrument All voltammetric trials were performed utilizing a Metrohm 884 Professional VA voltameter, and the obtained data were processed with VIVA software (version 2.0). The electrochemical setup employed a 3 sensors configuration, composed of a bare or modified CPE as the working electrode, an Ag/AgCl reference electrode (3 mol L⁻¹ KCl), to which all cell potentials were referenced, and a platinum wire serving as the auxiliary electrode. Electrochemical measurements were performed in a 50 mL glass cell.pH values were determined using a Hanna pH meter (model HI 2211 PH/ORP).Bench-top ultrasonic bath (POWERSONIC410, USA). All voltammetric trials were performed utilizing a Metrohm 884 Professional VA voltameter, and the obtained data were processed with VIVA software (version 2.0). The electrochemical setup employed a 3 sensors configuration, composed of a bare or modified CPE as the working electrode, an Ag/AgCl reference electrode (3 mol L⁻¹ KCl), to which all cell potentials were referenced, and a platinum wire serving as the auxiliary electrode. Electrochemical measurements were performed in a 50 mL glass cell. All voltammetric trials were performed utilizing a Metrohm 884 Professional VA voltameter, and the obtained data were processed with VIVA software (version 2.0). The electrochemical setup employed a 3 sensors configuration, composed of a bare or modified CPE as the working electrode, an Ag/AgCl reference electrode (3 mol L⁻¹ KCl), to which all cell potentials were referenced, and a platinum wire serving as the auxiliary electrode. Electrochemical measurements were performed in a 50 mL glass cell. pH values were determined using a Hanna pH meter (model HI 2211 PH/ORP). pH values were determined using a Hanna pH meter (model HI 2211 PH/ORP). Bench-top ultrasonic bath (POWERSONIC410, USA). Bench-top ultrasonic bath (POWERSONIC410, USA). Working electrode Carbon Paste Electrode (CPE) The CPE was fabricated by blending 0.5 g of graphite powder with approximately 0.3 mL of paraffin oil in a porcelain dish. The obtained paste was inserted into the electrode body, and its surface was subsequently polished to obtain a smooth finish with filter paper until a smooth, shiny finish was achieved. MWCNT-Modified CPE (2% MWCNT/CPE) To prepare the MWCNT-modified electrodes, the required proportions of MWCNTs and graphite powder were mixed with diethyl ether to ensure homogeneity. The mixture was air-dried, after which paraffin oil was incorporated and thoroughly blended until a consistent and well-wetted paste was formed, indicating uniform mixing of all components. Nanomaterial-Modified CPEs Modified sensors containing 2% MWCNT/CPE with additional modifiers such as 3% SiO₂ NPs, 3% ZrO₂ NPs, 3% chitosan, and 3% graphene were fabricated using the same procedure described above. Standard solutions A standard stock solution of CLP (1.5 × 10⁻³ mol L⁻¹) was formed by solubilized 10.0 mg of the compound in a 10 mL volumetric flask and completing the volume with double-distilled water. Construction of calibration curves Differential pulse (DP) voltammograms were obtained in the potential range of 0.0–1.60 V using a scan rate of 150 mV s⁻¹, with a pulse amplitude of 50 mV and a pulse width of 10 ms. Aliquots from the CLP stock solution (1.5 × 10⁻³ mol L⁻¹) were introduced into a 10 cm³ electrochemical cell containing 10 mL of 0.2 M phosphate buffer at pH 8, yielding final concentrations in the range of 7.5 × 10− 7 − 6.0 × 10− 5 mol L⁻¹ (0.5–40 µg mL⁻¹). A calibration curve was generated by plotting the current signal values against the corresponding drug concentrations, and the regression equation was subsequently determined. Pharmaceutical preparations A 14.1 mL portion of Notussil® suspension (labeled to contain 100 mg of CLP) was sonicated in an ultrasonic bath at 25 °C for 15 min, then filtered by a 0.45 μm disposable syringe filter. The filtrate was diluted to 100 mL with distilled water to obtain a 1 mg mL⁻¹ (≈ 1.5 × 10⁻³ mol L⁻¹) stock solution. Aliquots of this stock were transferred to a 10 cm³ electrolytic cell, and the electrochemical method described above was applied. CLP concentration was calculated using the linear regression equation obtained from the calibration curve. Validation and green assessment The analytical method was validated according to ICH Q2 (R1) guidelines. Validation parameters, including linearity, accuracy, precision, LOD, LOQ, and selectivity, were evaluated. In addition, the method was assessed using green analytical metrics (The Multi-color Assessment (MA) Tool, Analytical GREEnness Metric (AGREE), Complex Green analytical procedure index (Complex-GAPI), and the analytical Eco-scale to ensure environmentally friendly practices. Result and discussion Electrochemical behavior Cyclic voltammetry (CV) was employed to study the electrochemical behavior and redox characteristics of CLP at the modified carbon paste electrode. In contrast, differential pulse voltammetry (DPV), known for its enhanced sensitivity and lower background current, was utilized for quantitative determination, resulting in higher signal intensity compared to CV [9]. CV was performed over the potential range of 0.0–1.3 V at a scan rate of 150 mV s⁻¹. for CLP (1.5 × 10⁻5 mol L⁻¹, equivalent to 10.0 µg mL⁻¹) using the MWCNT-Chitosan/CPE electrode. A single anodic peak was detected at a potential of 0.8 V (Fig. 2a). No cathodic peaks appeared during the reverse scan, revealing that the electrochemical oxidation of the CLP at the 2%MWCNT-1%Chitosan/CPE surface is an irreversible process. DPV was also applied using a pulse amplitude of 50 mV, step potential of 5 mV, pulse width of 10 ms, and scan rate of 150 mV s⁻¹. All experiments were carried out at room temperature in 0.2 M phosphate buffer (pH 8). (Fig. 2b), showing a more defined peak with higher current intensity compared to CV. Therefore, DPV was chosen for quantitative analysis based on its balance between sensitivity, peak resolution, and reproducibility, rather than peak current intensity alone. Control trials using blank buffer solutions under both CV and DPV conditions showed no observable peaks. Fig. 2a Cyclic voltammogram and b differential pulse (DPV) signal of 1.5 × 10⁻⁵ mol L⁻¹ CLP solution in 0.2 M phosphate buffer (pH 8) recorded at a 2% MWCNT/1%Chitosan/CPE electrode with a scan rate of 150 mV s⁻¹ a Cyclic voltammogram and b differential pulse (DPV) signal of 1.5 × 10⁻⁵ mol L⁻¹ CLP solution in 0.2 M phosphate buffer (pH 8) recorded at a 2% MWCNT/1%Chitosan/CPE electrode with a scan rate of 150 mV s⁻¹ Factor optimization Influence of pH and buffer types The influence of pH on the electrochemical response of CLP (1.5 × 10⁻5 mol L⁻¹) was examined utilizing 0.04 M BR buffer solutions over the pH range 2–11 at a CPE with a scan rate of 150 mV s⁻¹, employing DP voltammetry (Fig. 3a). At pH 2, a single clearly distinguishable anodic signal was detected. Increasing the buffer pH resulted in a progressive rise in the current signal, achieving the highest value at pH 8. Beyond this value, a sharp decline in signal intensity was noted at pH 9, while at pH 11 the anodic peak completely disappeared. Based on these results, pH 8 was chosen for further investigations into the influence of buffer composition. The pKa of CLP is reported in the literature to be around 8–9, indicating that the drug exists in a partially protonated form near neutral to slightly alkaline media. At pH 8, the equilibrium between protonated and neutral species facilitates electron transfer at the electrode surface, resulting in enhanced oxidation current and improved analytical sensitivity. This explains why pH 8 was selected as the optimal condition for voltammetric determination. When the potential signal was plotted against the solution pH, a shift to more negative values was observed, supporting the assumption of involvement of protons in the oxidation procedure of the investigated drug at the surface of the sensor. As illustrated in Fig. 3b, a linear correlation was obtained between the potential signal (E/V) and pH, which was expressed by the following regression Equation [10]. \documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$ {\text{E }} = {\text{ K}} - \left( {0.0{\text{59 y}}/{\mathrm{n}}} \right){\text{ pH}} $$\end{document} where y represents the number of protons included in the electrode reaction, and n denotes the number of electrons transferred, as expressed by the following relation: \documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$ {\mathrm{E}}_{{\mathrm{p}}} = {\mathrm{1}}.0{\mathrm{9}}0{\mathrm{6}} - {\text{ }}0.0{\text{311pH, R}}^{{\mathrm{2}}} = 0.{\mathrm{9965}} $$\end{document} The slope of the potential signal –pH plot was determined to be 0.0311 V per pH unit, which is lower than the theoretical Nernstian value of 0.059 V per pH unit. This suggests that the number of protons participating in the oxidation process is lower than the number of electrons transferred. Such behavior implies that the oxidation of CLP at the surface of the sensor followed a proton–electron coupled mechanism, in which electron transfer is the dominant step, while proton involvement is limited. Since pH 8 was identified as the optimum condition, 0.2 M phosphate buffer and 0.2 M borate buffer (both at pH 8) were employed to assess their influence on the electrochemical response of CLP, in comparison with 0.04 M BR buffer. Among the tested media, 0.2 M phosphate buffer (pH 8) produced the highest current signal (Fig. 4). Fig. 3a DPV signal of 1.5 × 10⁻⁵ mol L⁻¹ CLP solution in 0.04 M BR buffer with pH ranging from 2 to 11, recorded at a CPE electrode and a scan rate of 150 mV s⁻¹. b Plot of potential signal (E) as a function of pH (2–11) for DP voltammograms of 1.5 × 10⁻⁵ mol L⁻¹ CLP solution in 0.04 M BR buffer, recorded at a CPE electrode with a scan rate of 150 mV s⁻¹ a DPV signal of 1.5 × 10⁻⁵ mol L⁻¹ CLP solution in 0.04 M BR buffer with pH ranging from 2 to 11, recorded at a CPE electrode and a scan rate of 150 mV s⁻¹. b Plot of potential signal (E) as a function of pH (2–11) for DP voltammograms of 1.5 × 10⁻⁵ mol L⁻¹ CLP solution in 0.04 M BR buffer, recorded at a CPE electrode with a scan rate of 150 mV s⁻¹ Fig. 4DPV signal of 1.5 × 10⁻⁵ mol L⁻¹ CLP solution in 0.04 M BR buffer, 0.2 M phosphate buffer, and 0.2 M borate buffer (pH 8), recorded at a CPE electrode with a scan rate of 150 mV s⁻¹ DPV signal of 1.5 × 10⁻⁵ mol L⁻¹ CLP solution in 0.04 M BR buffer, 0.2 M phosphate buffer, and 0.2 M borate buffer (pH 8), recorded at a CPE electrode with a scan rate of 150 mV s⁻¹ Different electrode types CPE was selected for studying the electrochemical performance of the analyte owing to its simple preparation, easy surface renewal, wide potential window, and the possibility of modification with different materials to improve selectivity and sensitivity [11]. Since metal oxide nanoparticles are often reported to enhance electrochemical signals due to their catalytic properties, different types were tested. However, adding 2% of zirconia (ZrO₂), silica (SiO₂), or titanium dioxide (TiO₂) nanoparticles to the CPE did not improve the response. Instead, the current decreased noticeably, as shown in Fig. 5. Other modifiers, such as 2% graphene nanoparticles, were also examined; however, graphene resulted in lower current and higher background noise compared to the unmodified electrode. These results highlight the well-known advantage of MWCNTs. Adding 2% MWCNTs to the paste improved the electrode’s uniformity, increased its surface area, and enhanced its conductivity[12–14]. Therefore, the 2% MWCNT/CPE was selected for further electrode modification because it provided the best balance between improved signal and stable background current. Next, several modifiers were tested in combination with the MWCNT/CPE. Adding 1% ZrO₂ or 1% SiO₂ nanoparticles did not enhance the response; instead, the signal decreased compared to the MWCNT/CPE alone (Fig. 6). On the other hand, incorporating chitosan produced a significant increase in the electrochemical signal. Thus, the electrode containing 2% MWCNT and 1% chitosan in the CPE was chosen as the optimal sensor for this study. The improved performance is likely due to chitosan’s ability to increase the effective surface area, enhance electron transfer, and form a uniform, biocompatible film that helps disperse the MWCNTs evenly. These properties make chitosan a valuable modifier for improving the sensitivity and stability of carbon-based electrochemical sensors[15–17]. Fig. 5DPV signal of 1.5 × 10⁻⁵ mol L⁻¹ CLP solution in 0.2 M phosphate buffer (pH 8), recorded at a bare CPE and at CPEs modified with 2% graphene, MWCNT, silica, Zr, and TiO₂, utilizing a scan rate of 150 mV s⁻¹ DPV signal of 1.5 × 10⁻⁵ mol L⁻¹ CLP solution in 0.2 M phosphate buffer (pH 8), recorded at a bare CPE and at CPEs modified with 2% graphene, MWCNT, silica, Zr, and TiO₂, utilizing a scan rate of 150 mV s⁻¹ Fig. 6DPV signal of 1.5 × 10⁻⁵ mol L⁻¹ CLP solution in 0.2 M phosphate buffer (pH 8), recorded at electrodes modified with 2% MWCNT–1% Zr, 2% MWCNT–1% silica, and 2% MWCNT–1% chitosan, using a scan rate of 150 mV s⁻¹ DPV signal of 1.5 × 10⁻⁵ mol L⁻¹ CLP solution in 0.2 M phosphate buffer (pH 8), recorded at electrodes modified with 2% MWCNT–1% Zr, 2% MWCNT–1% silica, and 2% MWCNT–1% chitosan, using a scan rate of 150 mV s⁻¹ Influence of scan rate The influence of scan rate on the current signal and potential was examined using a 1.5 × 10⁻5 mol L⁻¹ CLP solution in 0.2 M phosphate buffer (pH 8) with the 2% MWCNT/1% Chitosan/CPE sensor over a range of 20–200 mV s⁻¹. A linear relationship was obtained when plotting the logarithm of current signal (log I) versus the logarithm of scan rate (log ʋ) as illustrated in the Figure. 7. The corresponding regression equations were determined to be: \documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$${\text{log I }} = {\text{ }}0.{\text{4674 log }}\upsilon {\text{ }} - {\text{ }}0.{\mathrm{3652}},{\text{ R}}^{{\mathrm{2}}} = 0.{\mathrm{9949}}$$\end{document} The slope of the obtained equation was 0.4674, close to the theoretical value expected for diffusion-controlled processes, suggesting that the electrochemical reaction of the drug at the sensor surface is mainly diffusion-controlled[18]. Moreover, a linear dependence between the peak current and the square root of the scan rate was observed, providing additional confirmation that the electrochemical process is governed by diffusion control, as shown in Fig. 8. As the scan rate increased, a positive shift in potential signal was obtained, demonstrating the irreversible nature of the drug’s oxidation procedure at the electrode[19]. A well-defined peak with excellent repeatability was achieved at a scan rate of 150 mV s⁻¹, which was therefore chosen for subsequent method development and validation. Fig. 7Plot of log current signal (log I) versus log scan rate (log ʋ) in the range of 20–200 mV s⁻¹ for a 1.5 × 10⁻⁵ mol L⁻¹ CLP solution in 0.2 M phosphate buffer (pH 8) at the 2% MWCNT/1% Chitosan/CPE electrode Plot of log current signal (log I) versus log scan rate (log ʋ) in the range of 20–200 mV s⁻¹ for a 1.5 × 10⁻⁵ mol L⁻¹ CLP solution in 0.2 M phosphate buffer (pH 8) at the 2% MWCNT/1% Chitosan/CPE electrode Fig. 8Plot of peak current (I) versus square root of scan rate (\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$\:\surd\:$$\end{document} ʋ) for a 1.5 × 10⁻⁵ mol L⁻¹ CLP solution in 0.2 M phosphate buffer (pH 8) at the 2% MWCNT/1% chitosan/CPE electrode Plot of peak current (I) versus square root of scan rate (\documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$\:\surd\:$$\end{document} ʋ) for a 1.5 × 10⁻⁵ mol L⁻¹ CLP solution in 0.2 M phosphate buffer (pH 8) at the 2% MWCNT/1% chitosan/CPE electrode Fig. 9Plot of potential signal (E) versus log scan rate (log ʋ) for a 1.5 × 10⁻⁵ mol L⁻¹ CLP solution in 0.2 M phosphate buffer (pH 8) at the 2% MWCNT/1% chitosan/CPE electrode Plot of potential signal (E) versus log scan rate (log ʋ) for a 1.5 × 10⁻⁵ mol L⁻¹ CLP solution in 0.2 M phosphate buffer (pH 8) at the 2% MWCNT/1% chitosan/CPE electrode Proposed mechanism of CLP oxidation Based on the results from previous investigations, the electrochemical oxidation of the target analyte at the sensor surface was confirmed to be irreversible. Laviron’s equation was therefore applied to determine the number of electrons (n) involved in the observed peak, \documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$ {\mathrm{E}} = {\text{ E}}^{0} + {\text{ 2}}.{\mathrm{3}}0{\text{3 R T}}/\alpha {\text{ n F log RTK}}^{ \circ } /\alpha {\text{nF }} + {\text{ log }}\upsilon , $$\end{document} An α value of 0.5 was assumed. By plotting the potential signal (E) against log ʋ, linear relationships were observed (Fig. 9). From the slope value, the number of electrons was estimated to be two. Based on the data obtained, the oxidation process was found to involve one proton. These findings support the proposed oxidation mechanism of the studied drug, presented in Fig. 10. Fig. 10Schematic representation of the suggested electrochemical oxidation mechanism of CLP at the 2%MWCNT/1%chitosan/CPE electrode Schematic representation of the suggested electrochemical oxidation mechanism of CLP at the 2%MWCNT/1%chitosan/CPE electrode Electrode characterization As described above, functionalizing the bare CPE with MWCNTs and chitosan nanoparticles enlarges the electrode’s effective surface area and improves the electrical conductivity of the sensor. The larger surface provides more active sites, which facilitate greater adsorption and accumulation of the analyte on the modified sensor. To confirm this hypothesis, the electroactive surface areas of three sensors CPE, 2% MWCNT/CPE, and 2%MWCNT/1%Chitosan/CPE, were determined utilizing the Randles–Sevcik equation: \documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$ {\mathrm{Ip}} = {\text{ }}\left( {{\mathrm{2}}.{\mathrm{69}} \times {\mathrm{1}}0^{{\mathrm{5}}} } \right)\,{\mathrm{n}}^{{{\mathrm{3}}/{\mathrm{2}}}} \,{\text{x A}}\,{\text{x C x D}}^{{{\mathrm{1}}/{\mathrm{2}}}} \,\upsilon \,^{{{\mathrm{1}}/{\mathrm{2}}}} $$\end{document} The calculated electroactive surface areas were 0.03 cm² for the bare CPE, 0.05 cm² for the 2% MWCNT/CPE, and 0.14 cm² for the 2%MWCNT/1%Chitosan/CPE. These results show that introducing MWCNTs increased the surface area by approximately 1.67 fold, and the subsequent incorporation of chitosan nanoparticles further enhanced it to about 4.67 fold compared with the unmodified electrode. This marked enlargement of the effective surface area accounts for the high sensitivity observed for the designed electrode, as previously discussed. Furthermore, scanning electron microscopy (SEM) was performed on the CPE and 2% MWCNT/1%Chitosan/CPE electrodes, as illustrated in Fig. 11. Figure 11a shows the surface morphology of the bare CPE, while Fig. 11b demonstrates that the incorporation of MWCNTs significantly improved the surface uniformity and homogeneity. Furthermore, the three-dimensional network formed by chitosan nanoparticles introduced numerous cavities that serve as active sites for drug adsorption, thereby enhancing the electrode’s effective surface area. Fig. 11SEM micrographs of (a) bare CPE and (b) 2% MWCNT/1% chitosan/CPE electrode surface, showing improved homogeneity and uniformity after modification SEM micrographs of (a) bare CPE and (b) 2% MWCNT/1% chitosan/CPE electrode surface, showing improved homogeneity and uniformity after modification Influence of accumulation time The influence of accumulation time was investigated utilizing 1.5 × 10⁻⁵ M CLP solution in 0.2 M phosphate buffer of pH 8 at 2% MWCNT/1% Chitosan/CPE. The current signal was recorded against time for 25 s with 5 s intervals, utilizing DP at open circuit, as illustrated in Fig. 12. A sharp decline in the current signal was obtained as the accumulation time elevated. This behavior indicates that prolonged accumulation leads to surface saturation and possible adsorption of oxidation products on the active sites of the electrode, which hinders further electron transfer and favors a diffusion-controlled mass-transfer process of the analyte toward the sensor surface. It should be noted that the pre-accumulation step was employed solely as an auxiliary parameter to optimize the analytical signal and was not intended to imply an adsorption-controlled mass transport mechanism. The diffusion-controlled behavior was independently confirmed by the scan-rate study, as discussed in the previous section. Consequently, an accumulation time of 5 s was chosen as the optimum value for current signal measurements. Fig. 12Plot of current signal (I) versus accumulation time for a 1.5 × 10⁻⁵ mol L⁻¹ CLP solution in 0.2 M phosphate buffer (pH 8) at the 2% MWCNT/1% chitosan/CPE electrode, recorded at a scan rate of 150 mV s⁻¹ Plot of current signal (I) versus accumulation time for a 1.5 × 10⁻⁵ mol L⁻¹ CLP solution in 0.2 M phosphate buffer (pH 8) at the 2% MWCNT/1% chitosan/CPE electrode, recorded at a scan rate of 150 mV s⁻¹ Validation As highlighted earlier, differential pulse voltammetry (DPV) provided a stronger electrochemical response with sharper and more defined peaks. DP voltammograms were obtained, and the approach was validated in accordance with ICH guidelines for the direct quantification of CLP in both pure substance and marketed formulation [20]. Validation was carried out by assessing linearity, range, limit of detection (LOD), limit of quantity (LOQ), accuracy, precision, and selectivity (Table 1). Table 1Validation data of pure CLPParameterValueLinear range (M)7.50 × 10− 7- 6.00 × 10− 5Limit of detection (LOD) (M)1.11 × 10− 7Limit of quantification (LOQ) (M)3.70 × 10− 7Determination coefficient (R 2)0.9995Slope) µA·M⁻¹(0.7737Intercept (µA)1.0421Standard deviation of the slope) µA·M⁻¹(1.76Error %0.58RSD %1.75 Table 1Validation data of pure CLPParameterValueLinear range (M)7.50 × 10− 7- 6.00 × 10− 5Limit of detection (LOD) (M)1.11 × 10− 7Limit of quantification (LOQ) (M)3.70 × 10− 7Determination coefficient (R 2)0.9995Slope) µA·M⁻¹(0.7737Intercept (µA)1.0421Standard deviation of the slope) µA·M⁻¹(1.76Error %0.58RSD %1.75 Validation data of pure CLP Fig. 13Calibration curve showing the relationship between current signal (I) and CLP concentration (µg mL⁻¹) in 0.2 M phosphate buffer (pH 8), recorded at the 2% MWCNT/1% Chitosan/CPE electrode with a scan rate of 150 mV s⁻¹ for the detected peak Calibration curve showing the relationship between current signal (I) and CLP concentration (µg mL⁻¹) in 0.2 M phosphate buffer (pH 8), recorded at the 2% MWCNT/1% Chitosan/CPE electrode with a scan rate of 150 mV s⁻¹ for the detected peak Linearity and range After optimization of experimental parameters, the linearity of the proposed method was evaluated using nine different concentration levels over the range 7.5 × 10⁻⁷ – 6.0 × 10⁻5 mol L⁻¹ (equivalent to 0.5–40.0 µg mL⁻¹). For each concentration, three replicate voltammetric measurements were recorded, and the mean peak current values were used to construct a calibration plot between CLP concentration and current signal. The regression equation was found to be. \documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$${\mathrm{I}} = 0.{\mathrm{7737C}} + {\mathrm{1}}.0{\mathrm{421}},{\text{ R}}^{{\mathrm{2}}} = 0.{\mathrm{9995}}$$\end{document} where I is the current signal and C is the CLP concentration in µg mL⁻¹. Based on the regression data in Table I, the method showed a low residual standard deviation and a small percentage relative error (%E), verifying the precision of the calibration curve. Limit of detection (LOD) and Limit of quantitation (LOQ) According to ICH guidelines, the LOD and LOQ were determined experimentally, based on signal-to-noise considerations presented in Fig. 13. The LOD and LOQ were found to be 1.11 × 10− 7- 3.70 × 10− 7 mol L⁻¹, respectively. Accuracy and precision The accuracy of the suggested approach was investigated by assaying CLP in its pure form at 3 different concentrations, each analyzed in triplicate. The mean % recovery was 99.75%, with a relative standard deviation (%RSD) of 1.81%. Multiple replicate measurements of CLP evaluated the precision of the proposed approach were performed at a fixed concentration (n = 3), and the %RSD was calculated using the standard formula (%RSD = SD/Mean ×100), where SD is the standard deviation of the measured values. Repeatability (intra-day precision) was evaluated within a single day, while intermediate precision (inter-day) was determined across three separate days. The approach demonstrated high accuracy and precision, reflected by excellent % recovery and low % RSD, as summarized in Table 2. Table 2Accuracy and precision data for CLP were obtained utilizing the suggested electrochemical approachParameterIntra-day precisionInter-day precisionConcentration (mol L⁻¹)7.5 × 10− 6 1.5 × 10− 5 3.0 × 10− 57.5 × 10− 5 1.5 × 10− 5 3.0 × 10− 5%Recovery*100.59 101.99 99.0398.86 98.78 98.82±SD1.49 0.75 0.751.49 1.29 0.98%RSD1.48 0.73 0.751.51 1.31 0.99 Table 2Accuracy and precision data for CLP were obtained utilizing the suggested electrochemical approachParameterIntra-day precisionInter-day precisionConcentration (mol L⁻¹)7.5 × 10− 6 1.5 × 10− 5 3.0 × 10− 57.5 × 10− 5 1.5 × 10− 5 3.0 × 10− 5%Recovery*100.59 101.99 99.0398.86 98.78 98.82±SD1.49 0.75 0.751.49 1.29 0.98%RSD1.48 0.73 0.751.51 1.31 0.99 Accuracy and precision data for CLP were obtained utilizing the suggested electrochemical approach Selectivity The selectivity of the proposed method was investigated in the presence of compounds commonly found in pharmaceutical dosage forms, including ascorbic acid, lactose, mannitol, and sodium chloride. These potential interfering species were tested at concentrations approximately 100-fold higher than that of CLP. The observed variation in peak current was less than 5.0% for all examined compounds, indicating negligible interference. The experimental interference results are summarized in Table 3. All measurements were carried out, and demonstrating high recovery values and confirming that the tested interferents did not significantly affect the peak current of CLP. Table 3Effect of some excipients and interferents on the voltammetric response of 1.5 × 10⁻⁵ M of CLPInterferants% signal changeLactose1.9Sodium chloride3.1Mannitol2.4Ascorbic acid4.6 Table 3Effect of some excipients and interferents on the voltammetric response of 1.5 × 10⁻⁵ M of CLPInterferants% signal changeLactose1.9Sodium chloride3.1Mannitol2.4Ascorbic acid4.6 Effect of some excipients and interferents on the voltammetric response of 1.5 × 10⁻⁵ M of CLP Application in a pharmaceutical dosage form The proposed approach was successfully carried out to quantify CLP in its market dosage form, as presented in Table 4. A statistical comparison with the reported spectrofluorimetric approach [4] utilizing Student’s t-test and variance ratio F-test at the 95 confidence level showed no marked difference between the two approaches [21]. The standard addition technique was utilized to assess the influence of excipients and their potential influence on the assay results. Results, summarized in Table 5, indicate that the approach demonstrated high accuracy and excellent precision, with satisfactory % recoveries and low % RSD values. No marked matrix influence was observed. Table 4Determination of CLP in Notussil ® suspension dosage formClaimed Concentaration (mol L⁻¹)Suggested approachReported spectrofluorimetric approach 4Found concentration (mol L⁻¹)Average recovery (%)% error% RSDAverage recovery (%)% RSD1.50 × 10− 51.49 × 10− 599.33%0.670.0299.800.252.25 × 10− 52.23 × 10− 599.11%0.890.4799.503.75 × 10− 53.72 × 10− 599.20%0.800.69100.0Student’s t-test2.05 (2.78)*F-test1.03 (19)**The tabulated F- and t-values at a confidence level of P = 0.05Each value presented in the table represents the mean of 3 independent measurements. Table 4Determination of CLP in Notussil ® suspension dosage formClaimed Concentaration (mol L⁻¹)Suggested approachReported spectrofluorimetric approach 4Found concentration (mol L⁻¹)Average recovery (%)% error% RSDAverage recovery (%)% RSD1.50 × 10− 51.49 × 10− 599.33%0.670.0299.800.252.25 × 10− 52.23 × 10− 599.11%0.890.4799.503.75 × 10− 53.72 × 10− 599.20%0.800.69100.0Student’s t-test2.05 (2.78)*F-test1.03 (19)**The tabulated F- and t-values at a confidence level of P = 0.05Each value presented in the table represents the mean of 3 independent measurements. Determination of CLP in Notussil ® suspension dosage form *The tabulated F- and t-values at a confidence level of P = 0.05 Each value presented in the table represents the mean of 3 independent measurements. Table 5Results of CLP determination in its dosage form by standard addition techniqueAmount taken (mol L⁻¹)Amount added (mol L⁻¹)Amount found (mol L⁻¹)% Recovery7.50 × 10− 67.50 × 10− 61.48 × 10− 598.631.50 × 10− 52.23 × 10− 599.233.00 × 10− 53.72 × 10− 599.23Mean ± % RSD99.03 ± 1.26Each result represents the mean of 3 independent measurements Table 5Results of CLP determination in its dosage form by standard addition techniqueAmount taken (mol L⁻¹)Amount added (mol L⁻¹)Amount found (mol L⁻¹)% Recovery7.50 × 10− 67.50 × 10− 61.48 × 10− 598.631.50 × 10− 52.23 × 10− 599.233.00 × 10− 53.72 × 10− 599.23Mean ± % RSD99.03 ± 1.26Each result represents the mean of 3 independent measurements Results of CLP determination in its dosage form by standard addition technique Each result represents the mean of 3 independent measurements Comparative assessment A comparative evaluation of the proposed voltametric method with previously reported techniques is presented in Table 6. The results demonstrate that the voltammetric method offers notable advantages, including simplicity, fast response, good selectivity, low operational cost, and no requirement for sample pretreatment. Table 6Comparison of the proposed approach with the reported approachesMethodLinear rangeSlope(R2)LODReferenceSpectrophotometric2.0–10.0 µg/mL0.0390.9991.012 µg/mL[3]Spectrofluorimetric0.2–5.0 µg/mL39.100.99930.04 µg/mL[4]HPLC/UV20.0–80.0 µg/mL30.1010.99920.146 µg/mL[6]Our proposed method0.5–40.0 µg/mL0.77370.99950.05 µg/mL Table 6Comparison of the proposed approach with the reported approachesMethodLinear rangeSlope(R2)LODReferenceSpectrophotometric2.0–10.0 µg/mL0.0390.9991.012 µg/mL[3]Spectrofluorimetric0.2–5.0 µg/mL39.100.99930.04 µg/mL[4]HPLC/UV20.0–80.0 µg/mL30.1010.99920.146 µg/mL[6]Our proposed method0.5–40.0 µg/mL0.77370.99950.05 µg/mL Comparison of the proposed approach with the reported approaches Green assessment The environmental impact of the analytical methods was investigated according to four primary factors: elevated energy demand, excessive waste generation, intensive chemical consumption with related hazards, and overall chemical usage [22, 23]. In addition, the cost efficiency and eco-friendliness of the developed approach were examined utilizing four different assessment tools, as outlined below. The Multi-color Assessment (MA) Tool The Multi-color Assessment (MA) Tool [24] was utilized to investigate the environmental sustainability of the developed voltametric approach. This online platform combines four major evaluation tools, GEMAM, BAGI, RAPI, and VIGI, which yielded respective scores of 74.5%, 90.0%, 97.5%, and 35.0% (Table 7), providing an integrated evaluation of greenness, applicability, performance, and innovation. Method-specific information, including solvent type and volume, waste produced per analysis, instrument energy demand, sample preparation procedures, and analytical characteristics such as LOD, linear range, and precision, was entered into the MA-Tool interface. The system generated color-coded ratings for each assessment domain and an overall MA-score of 74.3 (Table 7), representing a high level of environmental compatibility in line with the principles of white analytical chemistry. The Analytical Eco-Scale The Analytical Eco-Scale is a metric utilized to evaluate the environmental friendliness of an analytical approach [25]. The system begins with a total score of 100, from which penalty points are subtracted. A final score above 75 indicates excellent environmental friendliness, a score above 50 reflects acceptable greenness, and anything below 50 is considered environmentally insufficient. It offers a quantitative way to evaluate analytical methods because it accounts for every reagent involved, not only the most hazardous ones, as some other evaluation systems do. Table 7 shows the calculated scores for the proposed method. It received 88 points, showing a high level of environmental friendliness. Analytical GREEnness metric (AGREE) The AGREE tool evaluates analytical approaches in accordance with the 12 principles of Green Analytical Chemistry (GAC), which encompass factors such as sample preparation, number of steps, waste generation, energy requirements, and toxicity [26, 27]. Each criterion is expressed on a unified scale ranging from 0 to 1. Within the pictogram, methods approaching a score of 1 and highlighted in dark green are considered more environmentally sustainable. Table 7 presents the pictograms of the proposed. The suggested approach achieved an AGREE score of 0.77. Complex Green analytical procedure index (Complex-GAPI) The GAPI tool offers a comprehensive investigation of analytical methodologies through fifteen parameters, each represented by a pentagram that reflects different stages of the procedure including sampling, sample preparation, instrumentation, solvents, and reagents employed, as well as the overall purpose of the method. The color coding in GAPI indicates the level of environmental impact: green denotes high ecological compatibility, yellow reflects moderate impact, and red signifies considerable environmental risk [28] As illustrated in the Table 7. The proposed approach obtained ten green regions. These findings, as demonstrated by the Comple-GAPI evaluation, indicate that the developed approach is environmentally sustainable. To further emphasize the sustainability of the proposed voltammetric method, the obtained greenness metrics were compared with those of previously reported analytical methods for CLP determination, including spectrofluorimetric and HPLC techniques. As summarized in Table 7, the proposed approach demonstrated a superior or comparable environmental profile across all applied assessment tools. Although the spectrofluorimetric method exhibited a slightly higher Eco-Scale score (89) compared to the proposed method (88), both approaches fall within the category of excellent environmental friendliness. In contrast, the HPLC method showed a noticeably lower Eco-Scale score (74), reflecting increased solvent consumption, higher waste generation, and greater energy demand. Table 7Environmental evaluation of the proposed method and other reported methods using four tools: Eco-Scale, Complex-GAPI, AGREE, and the Multi-color Assessment tool Table 7Environmental evaluation of the proposed method and other reported methods using four tools: Eco-Scale, Complex-GAPI, AGREE, and the Multi-color Assessment tool Environmental evaluation of the proposed method and other reported methods using four tools: Eco-Scale, Complex-GAPI, AGREE, and the Multi-color Assessment tool Moreover, the AGREE assessment revealed that the proposed voltammetric method achieved a higher overall greenness score, indicating better compliance with the principles of green analytical chemistry, particularly in terms of minimal sample preparation, reduced chemical usage, and lower environmental burden. The Complex-GAPI and MA-tool evaluations further supported these findings, highlighting a greater proportion of green regions and a favorable multi-criteria sustainability profile for the proposed method compared to the reported techniques. Collectively, these results confirm that the developed voltammetric approach offers a greener and more sustainable alternative to conventional methods, while maintaining reliable analytical performance. Conclusion The proposed green method represents the first practical trial to characterize the electrochemical performance of CLP. A 2%MWCNT/1%Chitosan/CPE electrode was fabricated for the quantification of CLP in its pure form and pharmaceutical formulation. Differential pulse voltammetric measurements using the modified sensor demonstrated excellent selectivity, high accuracy, low detection and quantification limits, and a broad linear range. The method proved to be suitable for drug assay without the need for complex procedures or costly instrumentation. Moreover, the electrochemical results generated in this study provide a valuable reference for future investigations, given that no electrochemical approaches for CLP have been previously reported.