Salvage of undeflatable intra-stent angioplasty-balloon-catheter with direct percutaneous needle puncture Authors: Leonardo Pasquetti, Edoardo Pasqui, Giuseppe Galzerano, Elisa Lazzeri, Bruno Gargiulo, Gianmarco de Donato Published: 2025 Main text Background To the Editor, Balloon angioplasty and stenting are standard components of endovascular treatment for peripheral arterial disease. Although balloon inflation and deflation are routine steps, failure of a balloon to deflate is an exceptionally rare but potentially dangerous complication [1]. Evidence from coronary interventions exists, but applicability to peripheral vessels is limited due to differences in vessel calibre, calcium burden, and device design [2]. Here, we describe a case of balloon deflation failure inside a stented superficial femoral artery (SFA), successfully managed with a direct percutaneous puncture. Case presentation A 65-year-old man with diabetes, dyslipidaemia, heavy smoking history, and chronic limb-threating ischemia (Rutherford V) presented with recurrent ulceration of the left forefoot stump. Angiography showed chronic total occlusion (CTO) of a previously stented SFA. The lesion was crossed and prepared with plain-old balloons (Advance Serenity 5 × 150 mm and 6 × 150 mm, Cook Medical, Bloomington, IN, USA) and a non-compliant balloon (Dorado 7 × 150 mm, BD, Franklin Lakes, NJ, USA) and then relined with two 6 × 170 mm Pulsar-18 T3 stents (Biotronik SE&Co. KG, Berlin, Germany). Post-dilation was performed with the previously used 6-mm hydrophilic-coated over-the-wire balloon. At the end of inflation, the balloon failed to deflate despite repeated attempts with standard measures, including repeated flushing and inflation above the rated burst pressure. The fully expanded balloon obstructed blood flow and could not be withdrawn without risking stent deformation or vascular injury (Fig. 1A). A percutaneous bailout technique was adopted. Under fluoroscopic guidance in a single-plane projection, a 21-gauge, 7-cm microset puncture needle with a transitionless-tip design was advanced through the mid-thigh skin and vessel wall to directly puncture the balloon through the stent struts (Fig. 1B). Subsequently, back “bleeding” (the mix of saline and contrast medium) was observed (Fig. 1C). Passive deflation was insufficient, so aspiration with a 10-cc syringe enabled complete balloon collapse. The catheter was removed intact. Reinflation outside the body confirmed the puncture site without other structural defects (Fig. 2). Final angiography demonstrated no arterial injury or recoil, with restored SFA patency. During stents post-dilation, upon completion of the inflation phase, the balloon failed to deflate despite repeated attempts using standard techniques. A The balloon catheter results in being undeflatable. B Direct puncture of the balloon catheter through the stent’s struts is performed under X-ray guidance. C: back “bleeding” (saline and contrast medium) is observed. Aspiration through the needle successfully facilitated the deflation of the balloon catheter At the end of the procedure, by inflating the balloon catheter, the site of the percutaneous puncture is observed. The rest of the device appeared intact Discussion Entrapment or inability to deflate a balloon is exceedingly uncommon, especially in peripheral interventions. In the coronary field, multiple bailout techniques have been described: intentional rupture via overinflation, high tip-load guidewire puncture, cutting the hypotube to restore luminal continuity [3], guide extension-mediated puncture [4], buddy balloon displacement [5], and excimer-laser perforation [6]. In our case, conventional measures were ineffective. Therefore, we opted for the direct puncture technique. The most plausible mechanism of deflation failure was hypotube damage, which prevented pressure transmission to the balloon. Such damage may occur during retrieval and reinsertion or when advancing through stent struts, heavily calcified CTO [7], or tortuous vessels. The presence of stents likely increased the risk of kinking or microperforation of the shaft. This case highlights the importance of careful balloon manipulation, particularly in previously stented vessels, and the need to replace any device showing resistance or suspected structural damage. When balloon deflation failure does occur, direct puncture represents, in percutaneously accessible vessels, a simple, rapid, and reproducible bailout technique. Acknowledgements Not applicable. Informed consent Informed consent was obtained from all individual participants included in the study. Abbreviations Chronic total occlusion Superficial femoral artery Authors’ contributions LP and EP contributed to the conception and design of the work and drafted the manuscript. GG, BG, and EL collected clinical data and contributed to data interpretation. GdD critically revised the manuscript for important intellectual content and supervised the overall project. All authors read and approved the final manuscript. Funding This study was not supported by any funding. Data availability Data sharing is not applicable to this article as no datasets were generated or analyzed during the current study. Declarations Ethics approval and consent to participate All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards. Consent for publication Consent for publication was obtained for every individual person’s data included in the study. Competing interests The authors declare that they have no financial and non-financial competing interests. Footnotes Publisher’s Note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations. References Associated Data Data Availability Statement Data sharing is not applicable to this article as no datasets were generated or analyzed during the current study.