Protocol for quantifying coccidian parasite mechanics and rupture force using advanced micromanipulation-based piercing Authors: Jana El Husseiny, Pierre-Henri Puech, Aurélien Dumètre, Julien Husson Published: 2025 ABSTRACT SummaryCoccidian parasites can rely on their mechanical properties to infect their hosts. Here, we detail a microindentation-based technique for quantifying the wall mechanics of coccidian oocysts. We describe steps for micropipette fabrication, calibration of flexible microindenters, system assembly, and quantifying indentation and rupture forces. We then present procedures for automated data analysis using open-source software and programming languages. The protocol enables measurement of wall rupture forces and is applicable to study the integrity of parasites under various physiological conditions.Subject areas: Biophysics, Microbiology, microscopy Before you begin This protocol outlines an advanced microindentation technique for quantifying the mechanical properties of the oocyst wall of coccidian parasites, specifically its bending stiffness (N/m) and rupture forces. The method is adapted from the cell indentation system designed by Husson.1 This approach operates across a wide force range (typically in the nN to μN range) and is particularly suitable for weakly or non-adhesive, asymmetric or inhomogeneous biological objects with cellular dimensions (∼5–30 μm). Figure 1 shows the main components of the microindentation setup. This configuration is essentially identical to the one described in Husson1 with the primary modification being the use of a needle-shaped indenter. Components not required for this application, such as the manual microinjector, are removed. Schematic of the main components of the profile microindentation setup 1 Glass-bottom Petri dish containing buffer and parasites; 2 micropipette holder (for the holding micropipette, IM-H1); 3 micropipette holder (for the microindenter, IM-H1); 4 right-angle bracket (MT402) and rod-holding clamp (accessory supplied with the MP-285 micromanipulator); 5 piezoelectric single-axis translation stage (P-611.1S Piezo Nanopositioner and E-709.SRG Digital Piezo Controller); 6 right-angle bracket (MT402) and metal plates (custom made) linking the 3-axis motorized micromanipulator and piezoelectric stage; 7 motorized micromanipulator (MP-285); 8 rigid stand with platform supporting the motorized micromanipulator (MP250); 9 pillar and mounting plates (XT66 and XT66P3); 10 vertical manual stage supporting the water reservoir (Unislide A1512DM-S1.5), with coarse and fine translation; 11 50 mL plastic syringe serving as water reservoir body; 12 auxiliary 30 mL plastic syringe for water level adjustment and pressure control, connected via a four-way stopcock; 13 microscope x-y stage translation knobs; 14 antivibration optical table; 15 camera; 16 rigid stand with platform (MP250); 17 three-axis manual translation stage (PT3/M); 18 rod-holding joint (UT-2, Universal Joint with Swing/Tilt Mechanism) with right angle joint (X-4 Rotating X-Block) and adjustment plate (P-1A, height adjustment plate) for attachment to the three-axis manual translation stage; 19 microscope (Eclipse Ti-U); 20 microscope illumination. The left inset shows a schematic view of the micropipette tips (not to scale); 21 holding micropipette; 22 microindenter. The right inset shows a schematic view of an aspirated parasite in contact with the microindenter tip; 23 parasite aspirated by the holding micropipette. Part of the microscope body is not shown, including oculars, fluorescent light source, and objective. We used a 40X air objective (Plan Fluor, 0.75 numerical aperture). Scale bar: 10 μm. This figure adapted from Husson1, Methods in Molecular Biology (https://doi.org/10.1007/978-1-0716-2851-5_1). The microindentation and piercing technique enables controlled piercing of the multilayered polymeric oocyst walls of coccidian parasites and accurate measurement of rupture forces while allowing precise control of parasite orientation. This capability supports a comprehensive understanding of the structure’s mechanical characteristics and integrity when exposed to biochemical or biophysical stresses that may influence parasite infectivity. The ability of this technique to explore different orientations is particularly valuable for studying asymmetric biological structures, as it allows for a complete assessment of wall strength variations that may be critical to the organism’s survival. Furthermore, the ability to perform these measurements on non-adherent biological objects significantly expands the range of microorganisms that can be studied, including coccidian parasites. By combining force measurements with real-time visual analysis, this protocol enables researchers to correlate structural deformations with specific force thresholds, providing insights into the wall mechanical properties, such as its elasticity and breaking points. This level of detail is crucial for deciphering how microorganisms withstand environmental pressures and how their walls might respond to potential therapeutic interventions. The consistency and reliability of the measurements, together with its relative ease of use for trained experimentalist, make it an invaluable tool for comparative studies across different species or strains of microorganisms. It also enables assessment of the effect of treatments or genetic modifications on wall integrity. Innovation The microindentation technique provides key advantages over other biophysical indentation techniques, such as atomic force microscopy (AFM), which generally requires adhesive cells and can restrict object orientation. In contrast, the approach described here enables mechanical characterization of non-adherent and asymmetrically shaped organisms and supports controlled variation of sample orientation. This protocol integrates custom-fabricated, needle-shaped glass microindenters with automated image-guided control to improve precision over existing micromanipulation and indentation techniques. Each microindenter is calibrated for bending stiffness, enabling reliable force measurements in the nanonewton to micronewton range. The system combines optical tracking with real-time feedback regulation of a piezoelectric stage to achieve spatial and temporal resolution during indentation and allowing detection of discrete rupture events associated with different wall layers. In parallel, an open-source Python-based computational pipeline performs automated signal filtering, peak detection, and force quantification from time-resolved datasets. By unifying microtool fabrication, calibration, mechanical testing, and data analysis, this method offers a reproducible, accessible, and versatile alternative to atomic force microscopy for investigating microorganism wall biomechanics and their responses to environmental stressors or therapeutics. Eimeria parasite sample preparation Timing: 20 min This section outlines how to prepare Eimeria oocysts for experimental use by washing them to remove storage solution and contaminants and resuspending them at a defined concentration for micromanipulation assays. 1.Remove Eimeria oocysts from potassium dichromate storage solution. Remove Eimeria oocysts from potassium dichromate storage solution. Note: Oocysts are collected and purified following established protocols.2,3,4 After purification, Eimeria oocysts are provided in potassium dichromate solution for long-term preservation. This solution also effectively inhibits microbial growth, maintaining oocyst viability and integrity. 2.Centrifuge the parasites at 900 × g for 5 min, then carefully remove the supernatant without disturbing the pellet containing the oocysts. Centrifuge the parasites at 900 × g for 5 min, then carefully remove the supernatant without disturbing the pellet containing the oocysts. 3.Resuspend the pellet in 1 mL of 1X DPBS (Dulbecco’s Phosphate-Buffered Saline) to wash the parasites. Resuspend the pellet in 1 mL of 1X DPBS (Dulbecco’s Phosphate-Buffered Saline) to wash the parasites. 4.Repeat steps 2–3 for a total of three washes. Repeat steps 2–3 for a total of three washes. 5.After the final wash, resuspend the oocysts in 1X DPBS to achieve a concentration of 1 million oocysts/mL. After the final wash, resuspend the oocysts in 1X DPBS to achieve a concentration of 1 million oocysts/mL. Note: If the solution remains orange due to residual potassium dichromate, perform additional wash steps until the solution becomes colorless.2 CRITICAL: While washing helps reduce contaminants, it is important to be aware that some debris may remain with the oocysts. DPBS/BSA 1% medium preparation Timing: 10 min This section describes how to prepare a 1% Bovine Serum Albumin (BSA) solution in DPBS for use in micromanipulation experiments. 6.Preparation of DPBS/BSA 1% solution.a.Weigh 0.5 g of BSA using an analytical balance.b.Dissolve the BSA in 50 mL of 1X DPBS in a sterile container. Mix gently to ensure complete dissolution while avoiding excessive foaming.c.Pass the prepared solution through a 0.22 μm sterile filter to remove any particulate matter and ensure sterility. Preparation of DPBS/BSA 1% solution.a.Weigh 0.5 g of BSA using an analytical balance.b.Dissolve the BSA in 50 mL of 1X DPBS in a sterile container. Mix gently to ensure complete dissolution while avoiding excessive foaming.c.Pass the prepared solution through a 0.22 μm sterile filter to remove any particulate matter and ensure sterility. a.Weigh 0.5 g of BSA using an analytical balance. Weigh 0.5 g of BSA using an analytical balance. b.Dissolve the BSA in 50 mL of 1X DPBS in a sterile container. Mix gently to ensure complete dissolution while avoiding excessive foaming. Dissolve the BSA in 50 mL of 1X DPBS in a sterile container. Mix gently to ensure complete dissolution while avoiding excessive foaming. c.Pass the prepared solution through a 0.22 μm sterile filter to remove any particulate matter and ensure sterility. Pass the prepared solution through a 0.22 μm sterile filter to remove any particulate matter and ensure sterility. CRITICAL: Filtration is essential to avoid clogging micropipettes or other equipment during downstream applications. Pause point: After this step, the prepared medium can be used immediately or stored at 4°C for up to one week in a tightly sealed container to maintain sterility to be used later. For longer storage, aliquot and freeze to prevent contamination and repeated freeze-thaw cycles. Key resources table REAGENT or RESOURCESOURCEIDENTIFIERChemicals, peptides, and recombinant proteinsBovine serum albumin (BSA)Sigma-AldrichA7030Dulbecco’s phosphate-buffered saline (DPBS)Thermo Fisher Scientific14190144Experimental models: Organisms/strainsEimeria acervulina oocysts, P3A strainRépérant et al.3NCBI Taxonomy: 5801Eimeria tenella oocysts, PAPt36 strainTomal et al.4NCBI Taxonomy: 5802Software and algorithmsAnaconda PythonAnaconda, Inc.https://www.anaconda.com/products/individualConda 24.7.1conda.orghttps://conda.org/blog/2024-07-23-july-releases/Fiji/ImageJ 2.16.0 (Java 1.8.0_322, 64-bit)Schneider et al.5https://imagej.net/software/fiji/MATLABMathWorkshttps://www.mathworks.com/products/matlab.htmlmatplotlib 3.5.3Matplotlib development teamhttps://matplotlib.org/3.5.3/api/_as_gen/matplotlib.pyplot.htmlMicro-manager 1.4.22Edelstein et al.6https://micro-manager.org/NumPy 1.21.6NumPy developershttps://numpy.org/devdocs/release/1.21.6-notes.htmlpandas 1.3.5pandas development teamhttps://pandas.pydata.org/pandas-docs/version/1.3/reference/api/pandas.read_csv.htmlPython 3.12.1Python Software Foundationhttps://www.python.org/downloads/release/python-3121/SciPy 1.7.3SciPy development teamhttps://docs.scipy.org/doc/scipy/release/1.7.3-notes.htmlseaborn 0.12.2seaborn development teamhttps://seaborn.pydata.org/whatsnew/v0.12.0.htmlOther4-way Luer stopcockWorld Precision Instrumentshttps://www.wpiinc.com/14036-15-4-way-luer-stopcock.html40X air objective (Plan Fluor, 0.75 numerical aperture)Nikon Instrumentshttps://www.microscope.healthcare.nikon.com/en_EU/products/optics/selector/comparison/-196150 and 30 mL plastic syringeTerumohttps://www.terumo-europe.com/en-emea/products/terumo%C2%AE-syringe-3-part-syringeAntivibration optical table M-VIS3636-PG4-325ANewport, Irvine, CA, USAhttps://www.newport.com/p/M-VIS3636-PG4-325ABorosilicate glass capillaries (1 mm OD, 0.78 mm ID)Sutter InstrumentBF100-78-10Eclipse Ti-U inverted microscopeNikon Instrumentshttps://www.microscope.healthcare.nikon.com/en_EU/products/inverted-microscopes/eclipse-ti-seriesForce transducer, model 406A (0–500 nN)Aurora Scientifichttps://aurorascientific.com/support/discontinued-products/400a-force-transducers/Glass-bottom Petri dish, FluoroDish, clear wall, 35 mm, 23 mm wellWorld Precision InstrumentsFD35-100MicroFil flexible needleWorld Precision InstrumentsMF34G-5Microforge MF2World Precision Instrumentshttps://www.wpiinc.com/var-3114-analog-microforge.html?srsltid=AfmBOoojwrpj4-76t4tzuGXAMBP9KU5KpM-UnkPMXbimWcoPd2z3-YORMicroforge MF-900Narishigehttps://www.tritechresearch.com/MF-900.htmlMicropipette holder IM-H1Narishigehttps://products.narishige-group.com/group1/IM-H1/injection/english.htmlMotorized micromanipulator MP-285Sutter Instrumenthttps://www.sutter.com/micromanipulation/mp-285ORCA-Flash4.0 V3 digital CMOS cameraHamamatsu Photonics, Hamamatsu City, Japanhttps://www.hamamatsu.com/us/en/product/cameras/cmos-cameras/C13440-20CU.htmlP-1A height adjustment plateNarishigehttps://products.narishige-group.com/group1/P-1A/injection/english.htmlP-97 flaming/brown micropipette pullerSutter Instrumenthttps://www.sutter.com/MICROPIPETTE/p-97.html?gad_source=1&gclid=Cj0KCQjw7Py4BhCbARIsAMMx-_Kzm31Dy1RNdefdki3RZd7gJpyc5KmGDeq14gYOvmQWk1omTd4D4qkaAuDaEALw_wcBPiezoelectric single-axis translation stage P-611.1S Piezo nanopositioner and E-709.SRG digital Piezo controllerPhysik Instrumente, Karlsruhe, Germanyhttps://www.physikinstrumente.com/en/Pillar and mounting plates XT66 and XT66P3Thorlabshttps://www.thorlabs.com/thorproduct.cfm?partnumber=XT66P3Right-angle bracket MT402Thorlabshttps://www.thorlabs.com/thorproduct.cfm?partnumber=MT402#ad-image-0Rigid stand with platform MP250Thorlabshttps://www.thorlabs.com/thorproduct.cfm?partnumber=MP250Stericup Quick Release Millipore Express Plus 0.22 μmSigma-AldrichS2GPU02REThree-axis manual translation stage PT3/MThorlabshttps://www.thorlabs.com/thorproduct.cfm?partnumber=PT3/M-Z8Universal joint with swing/tilt mechanism UT-2Narishigehttps://www.tritechresearch.com/UT-2.htmlVertical manual stage supporting the water reservoir UniSlide A1512DM-S1.5Velmex, Inc.https://velmex.com/products/unislide/unislide_a15_seriesX-4 rotating X-blockNarishigehttps://products.narishige-group.com/group1/X-4/electro/english.html Step-by-step method details Micropipette fabrication Timing: 5 min This section details how to fabricate a micropipette to hold and manipulate coccidian oocysts. The process includes pulling glass capillaries, trimming the tip, and bending the pipette for optimal access within a FluoroDish (see Figure 2). 1.Pull microneedles from glass capillaries.a.Mount a borosilicate glass capillary in the P-97 Flaming/Brown Micropipette Puller.b.Configure the puller parameters: Pressure: 500, Heat: 475, Pull: 40, Velocity: 140 and Time: 250. These are dimensionless parameters specific to the Sutter Instruments puller.c.Activate the puller to form the micropipettes. Pull microneedles from glass capillaries.a.Mount a borosilicate glass capillary in the P-97 Flaming/Brown Micropipette Puller.b.Configure the puller parameters: Pressure: 500, Heat: 475, Pull: 40, Velocity: 140 and Time: 250. These are dimensionless parameters specific to the Sutter Instruments puller.c.Activate the puller to form the micropipettes. a.Mount a borosilicate glass capillary in the P-97 Flaming/Brown Micropipette Puller. Mount a borosilicate glass capillary in the P-97 Flaming/Brown Micropipette Puller. b.Configure the puller parameters: Pressure: 500, Heat: 475, Pull: 40, Velocity: 140 and Time: 250. These are dimensionless parameters specific to the Sutter Instruments puller. Configure the puller parameters: Pressure: 500, Heat: 475, Pull: 40, Velocity: 140 and Time: 250. These are dimensionless parameters specific to the Sutter Instruments puller. c.Activate the puller to form the micropipettes. Activate the puller to form the micropipettes. Note: Thin-walled capillaries are preferred for flexible micropipettes, while thick-walled capillaries require adjusted puller settings. CRITICAL: Puller settings may vary depending on the desired micropipette characteristics, such as length and overall shape. Optimize parameters for your specific experimental requirements by referring to the puller provider documentation. 2.Trim the capillary end.a.Transfer the pulled capillary to the MF2 microforge and carefully trim the capillary end by using capillary wetting of the glass tube by melted glass from a bead.b.Measure the inner diameter of the capillary end using the calibrated graduations in the microforge ocular.c.Fine-tune trimming to achieve the target inner diameter of 7–9 μm for our experiments. Trim the capillary end.a.Transfer the pulled capillary to the MF2 microforge and carefully trim the capillary end by using capillary wetting of the glass tube by melted glass from a bead.b.Measure the inner diameter of the capillary end using the calibrated graduations in the microforge ocular.c.Fine-tune trimming to achieve the target inner diameter of 7–9 μm for our experiments. a.Transfer the pulled capillary to the MF2 microforge and carefully trim the capillary end by using capillary wetting of the glass tube by melted glass from a bead. Transfer the pulled capillary to the MF2 microforge and carefully trim the capillary end by using capillary wetting of the glass tube by melted glass from a bead. b.Measure the inner diameter of the capillary end using the calibrated graduations in the microforge ocular. Measure the inner diameter of the capillary end using the calibrated graduations in the microforge ocular. c.Fine-tune trimming to achieve the target inner diameter of 7–9 μm for our experiments. Fine-tune trimming to achieve the target inner diameter of 7–9 μm for our experiments. Note: The optimal inner diameter depends on your specific experimental needs. For Eimeria acervulina oocysts, we used a pipette with a 7–8 μm inner diameter, as their apparent diameter is around 18 μm. For E. tenella, we used an 8–9 μm pipette, corresponding to an apparent oocyst diameter of around 21 μm. Adjust the sizes to the objects of interest accordingly during the procedure. In general, the pipette inner diameter should be approximately 40% of the object’s apparent diameter and can be adjusted according to other species. 3.Angle the micropipette tip for 45° angle FluoroDish entry.a.Position the micropipette 20–100 μm from the microforge filament of the MF-900 microforge.b.With the filament in a horizontal position, approach and align the pipette from above to a position where the pipette is approximately 50 μm large and close to the heating point.c.Set the microfilament-only microforge to 80% power and apply heat in 1-second intervals.i.Repeat the heating cycles until the pipette bends to a 45° angle. In this configuration, the bending goes downward, i.e., toward the filament.ii.Readjust the pipette position if it moves too close to the heating element.iii.Verify the bend angle using the angular grid in the microforge ocular. Angle the micropipette tip for 45° angle FluoroDish entry.a.Position the micropipette 20–100 μm from the microforge filament of the MF-900 microforge.b.With the filament in a horizontal position, approach and align the pipette from above to a position where the pipette is approximately 50 μm large and close to the heating point.c.Set the microfilament-only microforge to 80% power and apply heat in 1-second intervals.i.Repeat the heating cycles until the pipette bends to a 45° angle. In this configuration, the bending goes downward, i.e., toward the filament.ii.Readjust the pipette position if it moves too close to the heating element.iii.Verify the bend angle using the angular grid in the microforge ocular. a.Position the micropipette 20–100 μm from the microforge filament of the MF-900 microforge. Position the micropipette 20–100 μm from the microforge filament of the MF-900 microforge. b.With the filament in a horizontal position, approach and align the pipette from above to a position where the pipette is approximately 50 μm large and close to the heating point. With the filament in a horizontal position, approach and align the pipette from above to a position where the pipette is approximately 50 μm large and close to the heating point. c.Set the microfilament-only microforge to 80% power and apply heat in 1-second intervals.i.Repeat the heating cycles until the pipette bends to a 45° angle. In this configuration, the bending goes downward, i.e., toward the filament.ii.Readjust the pipette position if it moves too close to the heating element.iii.Verify the bend angle using the angular grid in the microforge ocular. Set the microfilament-only microforge to 80% power and apply heat in 1-second intervals.i.Repeat the heating cycles until the pipette bends to a 45° angle. In this configuration, the bending goes downward, i.e., toward the filament.ii.Readjust the pipette position if it moves too close to the heating element.iii.Verify the bend angle using the angular grid in the microforge ocular. i.Repeat the heating cycles until the pipette bends to a 45° angle. In this configuration, the bending goes downward, i.e., toward the filament. Repeat the heating cycles until the pipette bends to a 45° angle. In this configuration, the bending goes downward, i.e., toward the filament. ii.Readjust the pipette position if it moves too close to the heating element. Readjust the pipette position if it moves too close to the heating element. iii.Verify the bend angle using the angular grid in the microforge ocular. Verify the bend angle using the angular grid in the microforge ocular. Note: The 45° angle ensures that when the micropipette holder is held at this position, its tip remains almost parallel to the bottom of the dish. This orientation facilitates the capture of objects at the bottom of the FluoroDish. CRITICAL: When heating sections thinner than 20 μm in diameter, the bending will curve the pipette upward, i.e. away from the filament. Note: The protocol indeed uses two different microforges: the MF2 for capillary trimming and the MF-900 for pipette bending at a 45° angle. While it would be in principle possible to adapt the protocol to use only the MF-900 microforge, this would present some technical challenges. The MF2 employs a glass bead on the heating filament for precise capillary trimming, whereas the MF-900 uses direct filament heating for non-contact bending of the pipettes. Adapting the MF-900 for both functions would require adding a glass bead to the filament and switching between these two heating modes within the same protocol, which might require upgrading the MF-900 with a high magnification objective (35X) to achieve the resolution typically provided by the MF2 system. For laboratories with budget constraints, we acknowledge that modifications to use a single MF-900 unit could be implemented, performing each step for a set of pipettes in a sequence. Pause point: After this step, the micropipette can be used immediately or stored in a large closed Petri dish, protected from dust, to prevent tip damage. Schematic and experimental setup for the microindentation technique (A) Schematic illustration of the fabrication and use of the holding pipette (moved using a piezoelectric device) and indenter (the motion of which is optically detected at high frame rates). The preparation process includes trimming, smoothing, and bending the holding pipette at a 45° angle. The indenter is then prepared by (a) sealing and shaping its tip to a conical-like form, followed by two bending steps: (b) bending the body at 90° angle for deflection measurement and (c) bending the tip at 45° angle for entering the FluoroDish. (B) Experimental setup showing the holding pipette and indenter under a microscope. (C) Microscopic view highlighting the interaction between the indenter and the target, with the target motion (red arrow) and deflection (blue arrow) indicated. Scale bar: 10 μm. Microindenter fabrication Timing: 5 min This section details how to create a custom microindenter from a glass capillary, essential for precise force application in mechanical property measurements of coccidian oocysts (see Figure 2). 4.Pull the glass capillary into a microneedle.a.Mount a borosilicate glass capillary, 1 mm OD, 0.78 mm ID, in the micropipette puller.b.Configure the puller parameters: Pressure: 500, Heat: 475, Pull: 45, Velocity: 140 and Time: 250. These are dimensionless parameters specific to the Sutter Instruments pullerc.Activate the puller to shape the capillary. Pull the glass capillary into a microneedle.a.Mount a borosilicate glass capillary, 1 mm OD, 0.78 mm ID, in the micropipette puller.b.Configure the puller parameters: Pressure: 500, Heat: 475, Pull: 45, Velocity: 140 and Time: 250. These are dimensionless parameters specific to the Sutter Instruments pullerc.Activate the puller to shape the capillary. a.Mount a borosilicate glass capillary, 1 mm OD, 0.78 mm ID, in the micropipette puller. Mount a borosilicate glass capillary, 1 mm OD, 0.78 mm ID, in the micropipette puller. b.Configure the puller parameters: Pressure: 500, Heat: 475, Pull: 45, Velocity: 140 and Time: 250. These are dimensionless parameters specific to the Sutter Instruments puller Configure the puller parameters: Pressure: 500, Heat: 475, Pull: 45, Velocity: 140 and Time: 250. These are dimensionless parameters specific to the Sutter Instruments puller c.Activate the puller to shape the capillary. Activate the puller to shape the capillary. 5.Seal the microneedle tip.a.Place the microneedle tip near the MF-900 filament microforge.b.Carefully melt the tip to seal it without deforming its shape. Seal the microneedle tip.a.Place the microneedle tip near the MF-900 filament microforge.b.Carefully melt the tip to seal it without deforming its shape. a.Place the microneedle tip near the MF-900 filament microforge. Place the microneedle tip near the MF-900 filament microforge. b.Carefully melt the tip to seal it without deforming its shape. Carefully melt the tip to seal it without deforming its shape. CRITICAL: Ensure complete tip sealing to prevent aspiration in subsequent steps. 6.Bend the microneedle tip to a 90° angle.a.Heat the filament on the MF-900 microforge to 45°C for 5s.b.Before the filament cools down, approach the microneedle tip to the heated filament. Then apply gentle pressure for 3 seconds to create a 90° angle at the tip, repeat this step until obtaining the exact angle. Bend the microneedle tip to a 90° angle.a.Heat the filament on the MF-900 microforge to 45°C for 5s.b.Before the filament cools down, approach the microneedle tip to the heated filament. Then apply gentle pressure for 3 seconds to create a 90° angle at the tip, repeat this step until obtaining the exact angle. a.Heat the filament on the MF-900 microforge to 45°C for 5s. Heat the filament on the MF-900 microforge to 45°C for 5s. b.Before the filament cools down, approach the microneedle tip to the heated filament. Then apply gentle pressure for 3 seconds to create a 90° angle at the tip, repeat this step until obtaining the exact angle. Before the filament cools down, approach the microneedle tip to the heated filament. Then apply gentle pressure for 3 seconds to create a 90° angle at the tip, repeat this step until obtaining the exact angle. Note: This 90° angle ensures that one can then perform optical deflection measurement under the microscope with the described setup configuration (see Figure 2). 7.Bend the microneedle body to a 45° angle in the plane perpendicular to the 90° angle bending.a.Identify a point on the microneedle with 50 μm thickness.b.Position this point near the MF-900 filament microforge.c.Apply heat in short bursts until the microneedle bends to a 45° angle, while verifying the angle using the angular grid in the microforge ocular. Bend the microneedle body to a 45° angle in the plane perpendicular to the 90° angle bending.a.Identify a point on the microneedle with 50 μm thickness.b.Position this point near the MF-900 filament microforge.c.Apply heat in short bursts until the microneedle bends to a 45° angle, while verifying the angle using the angular grid in the microforge ocular. a.Identify a point on the microneedle with 50 μm thickness. Identify a point on the microneedle with 50 μm thickness. b.Position this point near the MF-900 filament microforge. Position this point near the MF-900 filament microforge. c.Apply heat in short bursts until the microneedle bends to a 45° angle, while verifying the angle using the angular grid in the microforge ocular. Apply heat in short bursts until the microneedle bends to a 45° angle, while verifying the angle using the angular grid in the microforge ocular. Note: The two bends have to be the most precisely possible in perpendicular planes since the two micromanipulators, for the holding pipette and the indenter, are set perpendicular on our setup (see Figures 1 and 2). The overall geometry can be adapted to any different specific configuration. Note: Fine-tune heating parameters based on your specific microforge model. Pause point: The microindenter can be used immediately or stored in a dust-free, large closed Petri dish to protect the tip from damage. Microindenter calibration Timing: 20 min/indenter This protocol describes how to measure the bending stiffness of a newly pulled microindenter or flexible micropipette using a reference microindenter of known stiffness. 8.Set up.a.Mount a reference microindenter of known rigidity (with glass bead tip) on the calibration setup.b.Mount the newly pulled microindenter on the piezoelectric stage. Set up.a.Mount a reference microindenter of known rigidity (with glass bead tip) on the calibration setup.b.Mount the newly pulled microindenter on the piezoelectric stage. a.Mount a reference microindenter of known rigidity (with glass bead tip) on the calibration setup. Mount a reference microindenter of known rigidity (with glass bead tip) on the calibration setup. b.Mount the newly pulled microindenter on the piezoelectric stage. Mount the newly pulled microindenter on the piezoelectric stage. CRITICAL: Ensure the reference microindenter has an intact glass bead tip to prevent inaccurate measurements due to breakage. Note: This reference microindenter serves as a caliper of intermediate bending stiffness. The latter is measured using a calibrated force transducer.1 9.Alignment.a.Using the 40x air objective (Plan Fluor, 0.75 numerical aperture, Nikon Instruments), align the tips of both microindenters so they are facing each other with a good relative orientation.b.Ensure the microindenters are in the same focal plane. Alignment.a.Using the 40x air objective (Plan Fluor, 0.75 numerical aperture, Nikon Instruments), align the tips of both microindenters so they are facing each other with a good relative orientation.b.Ensure the microindenters are in the same focal plane. a.Using the 40x air objective (Plan Fluor, 0.75 numerical aperture, Nikon Instruments), align the tips of both microindenters so they are facing each other with a good relative orientation. Using the 40x air objective (Plan Fluor, 0.75 numerical aperture, Nikon Instruments), align the tips of both microindenters so they are facing each other with a good relative orientation. b.Ensure the microindenters are in the same focal plane. Ensure the microindenters are in the same focal plane. 10.Force application.a.Translate the base of the new microindenter using the piezoelectric stage.b.Press the tip of the new microindenter against the tip of the reference microindenter. Force application.a.Translate the base of the new microindenter using the piezoelectric stage.b.Press the tip of the new microindenter against the tip of the reference microindenter. a.Translate the base of the new microindenter using the piezoelectric stage. Translate the base of the new microindenter using the piezoelectric stage. b.Press the tip of the new microindenter against the tip of the reference microindenter. Press the tip of the new microindenter against the tip of the reference microindenter. 11.Deflection measurement.a.Record the deflection (dref) over time of the reference microindenter.b.Simultaneously record the deflection (dmi) of the new microindenter over time. Deflection measurement.a.Record the deflection (dref) over time of the reference microindenter.b.Simultaneously record the deflection (dmi) of the new microindenter over time. a.Record the deflection (dref) over time of the reference microindenter. Record the deflection (dref) over time of the reference microindenter. b.Simultaneously record the deflection (dmi) of the new microindenter over time. Simultaneously record the deflection (dmi) of the new microindenter over time. 12.Analysis.a.Plot dmi vs. dref to obtain a linear curve.b.Calculate the slope s = dmi/dref of this curve. Due to the equilibration of forces between both microindenters, Kmidmi=Krefdref, so that this slope is equal to Kref/Kmi, where Kref is the bending stiffness of the reference bead microindenter and Kmi is the bending stiffness of the microindenter to be calculated.c.Determine Kmi using the equation: Kmi = Kref/s, where s is the slope determined at step b. Analysis.a.Plot dmi vs. dref to obtain a linear curve.b.Calculate the slope s = dmi/dref of this curve. Due to the equilibration of forces between both microindenters, Kmidmi=Krefdref, so that this slope is equal to Kref/Kmi, where Kref is the bending stiffness of the reference bead microindenter and Kmi is the bending stiffness of the microindenter to be calculated.c.Determine Kmi using the equation: Kmi = Kref/s, where s is the slope determined at step b. a.Plot dmi vs. dref to obtain a linear curve. Plot dmi vs. dref to obtain a linear curve. b.Calculate the slope s = dmi/dref of this curve. Due to the equilibration of forces between both microindenters, Kmidmi=Krefdref, so that this slope is equal to Kref/Kmi, where Kref is the bending stiffness of the reference bead microindenter and Kmi is the bending stiffness of the microindenter to be calculated. Calculate the slope s = dmi/dref of this curve. Due to the equilibration of forces between both microindenters, Kmidmi=Krefdref, so that this slope is equal to Kref/Kmi, where Kref is the bending stiffness of the reference bead microindenter and Kmi is the bending stiffness of the microindenter to be calculated. c.Determine Kmi using the equation: Kmi = Kref/s, where s is the slope determined at step b. Determine Kmi using the equation: Kmi = Kref/s, where s is the slope determined at step b. Pause point: The calibrated microindenter can be used immediately or stored in a large closed Petri dish away from dust and shocks to protect the tip. It can be reused for multiple experiments as long as the tip remains intact. Microscope setup and sample preparation Timing: 20 min This section outlines how to prepare the observation chamber, microindentation system setup, and sample loading for coccidian oocyst wall analysis. 13.Prepare the Petri dish used for microindentation experiments.a.Place a FluoroDish on an inverted microscope equipped with a 40x air objective.b.Add 3 ml of DPBS/BSA 1% to the dish and keep it at 20°C–22°C. Prepare the Petri dish used for microindentation experiments.a.Place a FluoroDish on an inverted microscope equipped with a 40x air objective.b.Add 3 ml of DPBS/BSA 1% to the dish and keep it at 20°C–22°C. a.Place a FluoroDish on an inverted microscope equipped with a 40x air objective. Place a FluoroDish on an inverted microscope equipped with a 40x air objective. b.Add 3 ml of DPBS/BSA 1% to the dish and keep it at 20°C–22°C. Add 3 ml of DPBS/BSA 1% to the dish and keep it at 20°C–22°C. 14.Set up the micropipette system.a.Fill the water reservoir with ultra-pure water.b.Configure the 4-way stopcock to fill the auxiliary micropipette and tubing.c.Ensure no bubbles are present in the system. If any, purge the system from them. Set up the micropipette system.a.Fill the water reservoir with ultra-pure water.b.Configure the 4-way stopcock to fill the auxiliary micropipette and tubing.c.Ensure no bubbles are present in the system. If any, purge the system from them. a.Fill the water reservoir with ultra-pure water. Fill the water reservoir with ultra-pure water. b.Configure the 4-way stopcock to fill the auxiliary micropipette and tubing. Configure the 4-way stopcock to fill the auxiliary micropipette and tubing. c.Ensure no bubbles are present in the system. If any, purge the system from them. Ensure no bubbles are present in the system. If any, purge the system from them. CRITICAL: Bubbles can interfere with aspiration efficiency and pressure control. 15.Prepare the holding micropipette.a.Connect a syringe to a blunt fill needle with a piece of rubber joint at its tip.b.Connect this auxiliary air syringe to the back of a pre-forged stiff micropipette and dip its tip in a tube of DPBS 1X while making sure not to touch the tube walls and exert aspiration pressure with the syringe.c.Complete filling of the micropipette through its back using a water-filled syringe with a microfil flexible needle. Prepare the holding micropipette.a.Connect a syringe to a blunt fill needle with a piece of rubber joint at its tip.b.Connect this auxiliary air syringe to the back of a pre-forged stiff micropipette and dip its tip in a tube of DPBS 1X while making sure not to touch the tube walls and exert aspiration pressure with the syringe.c.Complete filling of the micropipette through its back using a water-filled syringe with a microfil flexible needle. a.Connect a syringe to a blunt fill needle with a piece of rubber joint at its tip. Connect a syringe to a blunt fill needle with a piece of rubber joint at its tip. b.Connect this auxiliary air syringe to the back of a pre-forged stiff micropipette and dip its tip in a tube of DPBS 1X while making sure not to touch the tube walls and exert aspiration pressure with the syringe. Connect this auxiliary air syringe to the back of a pre-forged stiff micropipette and dip its tip in a tube of DPBS 1X while making sure not to touch the tube walls and exert aspiration pressure with the syringe. c.Complete filling of the micropipette through its back using a water-filled syringe with a microfil flexible needle. Complete filling of the micropipette through its back using a water-filled syringe with a microfil flexible needle. CRITICAL: One should be able to avoid any bubbles, but if you trap one (which will be obvious by eye inspection), gently flick the micropipette with the nail, several times. You will see the bubble deform and eventually detach from the inner wall of the micropipette. Use the long capillary tip again to make sure the bubble doesn’t stay trapped at the meniscus at the back of the capillary. 16.Connect this liquid-filled pipette with the micropipette holder.a.Hold the micropipette holder below the water reservoir level.b.Allow drops to flow and coalesce with the meniscus at the back of the pipette.c.Ensure no bubbles are trapped. Connect this liquid-filled pipette with the micropipette holder.a.Hold the micropipette holder below the water reservoir level.b.Allow drops to flow and coalesce with the meniscus at the back of the pipette.c.Ensure no bubbles are trapped. a.Hold the micropipette holder below the water reservoir level. Hold the micropipette holder below the water reservoir level. b.Allow drops to flow and coalesce with the meniscus at the back of the pipette. Allow drops to flow and coalesce with the meniscus at the back of the pipette. c.Ensure no bubbles are trapped. Ensure no bubbles are trapped. 17.Place the micropipette tip on the microscope and center it in the field of view, typically 30 μm above the bottom glass slide. Place the micropipette tip on the microscope and center it in the field of view, typically 30 μm above the bottom glass slide. 18.Mount the calibrated microindenter on the holder. Place the tip of the indenter at about 30 μm above the bottom of the dish, which is in the same plane as the holding pipette. Mount the calibrated microindenter on the holder. Place the tip of the indenter at about 30 μm above the bottom of the dish, which is in the same plane as the holding pipette. Note: Working about 30 μm above the dish is a trade-off between good image quality (the better the closer from the bottom of the dish) and not having the image disturbed by oocysts present at the bottom of the dish. Make sure the microindenter shaft does not touch another oocyst outside of the field of view. 19.Establish aspiration pressure.a.Set an initial aspiration pressure equivalent to a 5 mm water column height difference.b.Allow BSA in the medium to coat the outer pipette walls, minimizing oocyst adhesion, for 5 min at 20°C–22°C. Establish aspiration pressure.a.Set an initial aspiration pressure equivalent to a 5 mm water column height difference.b.Allow BSA in the medium to coat the outer pipette walls, minimizing oocyst adhesion, for 5 min at 20°C–22°C. a.Set an initial aspiration pressure equivalent to a 5 mm water column height difference. Set an initial aspiration pressure equivalent to a 5 mm water column height difference. b.Allow BSA in the medium to coat the outer pipette walls, minimizing oocyst adhesion, for 5 min at 20°C–22°C. Allow BSA in the medium to coat the outer pipette walls, minimizing oocyst adhesion, for 5 min at 20°C–22°C. Note: For optimal results, extend the incubation time to 15–20 min if working with highly adhesive parasite or if initial trials show significant oocyst sticking. CRITICAL: The effectiveness of BSA coating can vary depending on the specific parasite type and experimental conditions. Monitor parasite behavior during initial experiments and adjust the incubation time accordingly to achieve optimal anti-adhesion properties. Note: The pressure difference (ΔP) is determined by the height difference (Δh) between the water level in the reservoir and the dish buffer level: where: 20.Load and position the sample.a.Gently inject 10 μL of parasite suspension (2×106 oocysts/mL) at the bottom of the dish.b.Allow oocysts to sediment for 2–3 min.c.Locate the area with the highest oocyst density. Load and position the sample.a.Gently inject 10 μL of parasite suspension (2×106 oocysts/mL) at the bottom of the dish.b.Allow oocysts to sediment for 2–3 min.c.Locate the area with the highest oocyst density. a.Gently inject 10 μL of parasite suspension (2×106 oocysts/mL) at the bottom of the dish. Gently inject 10 μL of parasite suspension (2×106 oocysts/mL) at the bottom of the dish. b.Allow oocysts to sediment for 2–3 min. Allow oocysts to sediment for 2–3 min. c.Locate the area with the highest oocyst density. Locate the area with the highest oocyst density. 21.Calibrate aspiration pressure.a.Adjust the reservoir height to achieve zero aspiration pressure, by first setting the zero level of the water tank to the height of the sample and then fine-tuning its position to have no aspiration or pushing away residual small particles in the suspension.b.Set the aspiration level for the experiment (typically 3–5 mm water column height difference). Calibrate aspiration pressure.a.Adjust the reservoir height to achieve zero aspiration pressure, by first setting the zero level of the water tank to the height of the sample and then fine-tuning its position to have no aspiration or pushing away residual small particles in the suspension.b.Set the aspiration level for the experiment (typically 3–5 mm water column height difference). a.Adjust the reservoir height to achieve zero aspiration pressure, by first setting the zero level of the water tank to the height of the sample and then fine-tuning its position to have no aspiration or pushing away residual small particles in the suspension. Adjust the reservoir height to achieve zero aspiration pressure, by first setting the zero level of the water tank to the height of the sample and then fine-tuning its position to have no aspiration or pushing away residual small particles in the suspension. b.Set the aspiration level for the experiment (typically 3–5 mm water column height difference). Set the aspiration level for the experiment (typically 3–5 mm water column height difference). 22.Prepare for indentation.a.Lower the holding pipette toward the bottom of the dish, close to a chosen parasite.b.Aspirate a single parasite, tuning its orientation by successive manipulations.c.Go up and position the parasite in front of the microindenter tip. Prepare for indentation.a.Lower the holding pipette toward the bottom of the dish, close to a chosen parasite.b.Aspirate a single parasite, tuning its orientation by successive manipulations.c.Go up and position the parasite in front of the microindenter tip. a.Lower the holding pipette toward the bottom of the dish, close to a chosen parasite. Lower the holding pipette toward the bottom of the dish, close to a chosen parasite. b.Aspirate a single parasite, tuning its orientation by successive manipulations. Aspirate a single parasite, tuning its orientation by successive manipulations. c.Go up and position the parasite in front of the microindenter tip. Go up and position the parasite in front of the microindenter tip. Note: By gently flicking the micropipette holder, the held parasite can be slightly reoriented. This procedure can be repeated as necessary until the desired parasite position is achieved. Image acquisition Timing: 3–5 min/sample This section describes how to set up and perform automated image acquisition during parasite indentation experiments. It details the configuration of the microscope and camera, initialization of control software, and acquisition of live imaging data used for feedback regulation of the piezoelectric stage. 23.Configure global imaging parameters.a.Set room temperature to 20°C–22°C to avoid thermal drift.b.Activate LED fluorescence light sources (CoolLED pE-300 ultra).c.Launch the computer, camera, and microscope and allow the system to thermally equilibrate before starting image acquisition. Configure global imaging parameters.a.Set room temperature to 20°C–22°C to avoid thermal drift.b.Activate LED fluorescence light sources (CoolLED pE-300 ultra).c.Launch the computer, camera, and microscope and allow the system to thermally equilibrate before starting image acquisition. a.Set room temperature to 20°C–22°C to avoid thermal drift. Set room temperature to 20°C–22°C to avoid thermal drift. b.Activate LED fluorescence light sources (CoolLED pE-300 ultra). Activate LED fluorescence light sources (CoolLED pE-300 ultra). c.Launch the computer, camera, and microscope and allow the system to thermally equilibrate before starting image acquisition. Launch the computer, camera, and microscope and allow the system to thermally equilibrate before starting image acquisition. Note: Detailed specifications of the imaging setup, including the microscope brand and model, objective properties, and piezoelectric stage, are provided in Figure 1. Imaging was performed on an inverted microscope (Eclipse Ti-U, Nikon Instruments) using a Hamamatsu ORCA-Flash4.0 sCMOS camera, where one pixel corresponds to 0.16 μm with the 40× air objective. Images were acquired with 1×1 binning, an exposure time of 2 ms, and Micro-manager software (version 1.4.22). 24.Initialize automation software.a.Launch MATLAB code for piezoelectric stage control.b.Our MATLAB code opens and controls the Micro-manager software, which in turn controls the camera. Initialize automation software.a.Launch MATLAB code for piezoelectric stage control.b.Our MATLAB code opens and controls the Micro-manager software, which in turn controls the camera. a.Launch MATLAB code for piezoelectric stage control. Launch MATLAB code for piezoelectric stage control. b.Our MATLAB code opens and controls the Micro-manager software, which in turn controls the camera. Our MATLAB code opens and controls the Micro-manager software, which in turn controls the camera. 25.Set up image analysis.a.Define a line crossing the microindenter shaft on the live microscopy image.b.Record a reference intensity profile for further live image cross-correlation analysis. Set up image analysis.a.Define a line crossing the microindenter shaft on the live microscopy image.b.Record a reference intensity profile for further live image cross-correlation analysis. a.Define a line crossing the microindenter shaft on the live microscopy image. Define a line crossing the microindenter shaft on the live microscopy image. b.Record a reference intensity profile for further live image cross-correlation analysis. Record a reference intensity profile for further live image cross-correlation analysis. 26.Automated feedback control, where an algorithm in MATLAB continuously analyzes images and adjusts the piezoelectric stage position through a feedback loop.a.The parasite is moved at constant velocity (∼0.5 μm/s) toward the microindenter tip.b.Once contact is established, the applied force increases until it reaches a predefined threshold (typically 36–50 μN for Eimeria oocysts).c.After reaching the maximum defined force, the oocyst is retracted at constant velocity. Automated feedback control, where an algorithm in MATLAB continuously analyzes images and adjusts the piezoelectric stage position through a feedback loop.a.The parasite is moved at constant velocity (∼0.5 μm/s) toward the microindenter tip.b.Once contact is established, the applied force increases until it reaches a predefined threshold (typically 36–50 μN for Eimeria oocysts).c.After reaching the maximum defined force, the oocyst is retracted at constant velocity. a.The parasite is moved at constant velocity (∼0.5 μm/s) toward the microindenter tip. The parasite is moved at constant velocity (∼0.5 μm/s) toward the microindenter tip. b.Once contact is established, the applied force increases until it reaches a predefined threshold (typically 36–50 μN for Eimeria oocysts). Once contact is established, the applied force increases until it reaches a predefined threshold (typically 36–50 μN for Eimeria oocysts). c.After reaching the maximum defined force, the oocyst is retracted at constant velocity. After reaching the maximum defined force, the oocyst is retracted at constant velocity. Note: The position of the tip of the microindenter is measured each time the camera acquires a new image, at an approximate frequency of 400 Hz. The oocyst is pushed against the indenter, so that indenter deflection is the difference between the tip’s updated position and its position under no force measured at the start of the experiment. The force applied to the oocyst is the product of the microindenter deflection and the microindenter bending stiffness k, this being applied during the data processing steps. Note: After retracting the micropipette, discard the tested oocyst by either (1) gently tapping the micropipette to release it at the edge of the dish (where no other oocysts are present) or (2) moving the pipette to a peripheral area, stopping aspiration, and gently expelling the oocyst. Retrieve a new oocyst from the center of the dish for the next measurement. 27.Configure image acquisition and indentation parameters in MATLAB.a.Set up camera setting: 1 frame per second.b.Configure indentation speed: set piezoelectric stage translation speed to 5 μm/s.c.Define deflection limits: maximum deflection depth (i.e., piezoelectric motion) is set to 5–7 μm. Configure image acquisition and indentation parameters in MATLAB.a.Set up camera setting: 1 frame per second.b.Configure indentation speed: set piezoelectric stage translation speed to 5 μm/s.c.Define deflection limits: maximum deflection depth (i.e., piezoelectric motion) is set to 5–7 μm. a.Set up camera setting: 1 frame per second. Set up camera setting: 1 frame per second. b.Configure indentation speed: set piezoelectric stage translation speed to 5 μm/s. Configure indentation speed: set piezoelectric stage translation speed to 5 μm/s. c.Define deflection limits: maximum deflection depth (i.e., piezoelectric motion) is set to 5–7 μm. Define deflection limits: maximum deflection depth (i.e., piezoelectric motion) is set to 5–7 μm. Note: Adjust exposure time and resolution to balance image quality and acquisition speed. Ensure that the frame rate is high enough to capture the indentation process without motion blur. CRITICAL: The maximum deflection depth should be carefully selected to prevent sample damage or indenter tip breakage. Start with a lower depth and gradually increase if needed. Note: For more dynamic processes or to capture rapid deformations, increase the frame rate to 10–20 frames per second, adjusting exposure time accordingly. Expected outcomes The microindenter (also called microneedle herein) deflection vs. time obtained using this protocol are converted to force measurements by multiplying the microindenter deflection by the calibrated bending stiffness of the microindenter. Parasite indentation is calculated by subtracting the microindenter deflection from the piezoelectric stage displacement. If an infinitely stiff object were brought into contact with the microindenter, the piezoelectric displacement would be entirely transmitted to microindenter deflection. However, when a deformable object such as a parasite is pressed against the microindenter tip, the piezoelectric displacement is partitioned between parasite indentation and microindenter deflection. As a result, the parasite indentation is obtained by a straightforward subtraction of the microindenter deflection to the piezoelectric displacement. This relationship between piezoelectric displacement, microindenter deflection, and target indentation is further explained in Husson.1 As illustrated in Figure 3, we obtain in fine a series of images capturing real-time rupture events for each wall component, synchronized with corresponding force/indentation vs. time curves that provide a comprehensive visualization of force dynamics throughout the indentation process. In panel A, the force–time curve shows the gradual buildup of force as the oocyst is indented, punctuated by distinct rupture events corresponding to mechanical failure of structural components. Panels B–G provide synchronized microscopy snapshots at key time points. The sequence begins before contact (B), followed by proximal wall contact (C) and its rupture (D), after which the microindenter continues until distal wall contact (E) and rupture (F). Finally, the glass–glass contact between the microindenter tip and the inner holding pipette wall is reached (G), corresponding to the sharp linear increase in force. Panel H summarizes proximal wall rupture forces across multiple measurements, highlighting the reproducibility and variability of the observed mechanical thresholds. Force–time curve showing the mechanical response of the Eimeria oocyst wall under microindentation (A) The force (μN) is plotted against time (s), with key events marked by colored dots. Black dots indicate apparent (image determined) contact points: (C) first contact with the proximal wall (t = 12 s) and (E) contact with the distal wall (t = 60 s). Red dots denote rupture events: (D) rupture of the proximal wall (t = 34 s) and (F) rupture of the distal wall (t = 78 s). (G) The orange dot represents the peak force at glass-glass contact (t = 87 s). (B–G) Representative brightfield images corresponding to each event highlight structural deformation and rupture of the E. tenella oocyst wall at each stage. (H) Boxplot showing the distribution of proximal wall rupture forces for E. acervulina (μN). Boxes represent the interquartile range, the horizontal line indicates the median, whiskers show the 1.5× interquartile range, and each dot corresponds to a single oocyst rupture. Scale bar: 10 μm. Glass-glass contact cannot be determined with sufficient accuracy through visual inspection alone. However, this event is clearly identifiable in the deflection vs. time curve. When the needle tip contacts the inner wall of the holding micropipette, friction causes the tip to interlock with the micropipette wall. Consequently, the microindenter tip suddenly begins moving at the constant translation velocity of the piezoelectric stage (0.5 μm/s). Glass-glass contact typically occurs after the distal wall of the parasite has been pierced and therefore does not interfere with our measurements. The algorithm controlling piezoelectric translation is programmed to halt when a maximum deflection, set in the interface of the software, is reached, preventing microindenter breakage. Measuring both force and indentation vs. time enables precise quantification of indentation and rupture forces for both proximal and distal walls of the oocyst after a meticulous calibration of the microneedle bending stiffness. These force curves reveal critical mechanical behaviors, beginning with initial contact between the microindenter and oocyst wall, progressing through elastic deformation of the structure and yield points, culminating in distinct rupture events marked by sudden force drops. The synchronized presentation of visual time points and force curves allows us to correlate specific structural events with quantitative force measurements, offering nuanced insights into the oocyst wall mechanics. By tracking the differential strengths between proximal and distal walls and examining rate-dependent mechanical properties, the protocol provides an unprecedented understanding of cellular structural integrity. See Methods video S1 for a time-synchronized demonstration of mechanical indentation of the oocyst wall. Methods video S1. Video captures the mechanical indentation of an Eimeria tenella oocyst wall using a microindenter, related to “expected outcomes” section The left panel shows a time-synchronized bright-field microscopic view of the indentation process, with a scale bar indicating 10 μm. The right panel presents a force-time graph, where the rupture force (μN) is plotted against time (seconds). The force increases as the probe indents the oocyst wall, with sudden drops indicating rupture events. The correlation between the video and the graph provides insights into the biomechanical properties of the oocyst wall. Quantification and statistical analysis For data acquisition, analysis and post-processing, we chose to use open source software and programming languages. We use the last version of Micro-manager (version 1.4.22) and Fiji (version 2.16.0), this latter being a meta-package of ImageJ, with specialized plugins available either on Fiji/ImageJ website, programmer’s website or GitHub repository. For Python, we use the latest release of Anaconda Python distribution (version 3.10.18), which contains all the needed packages for data science applications, together with the Jupyter notebook system utilized in this protocol. The complete analysis code is openly available on GitHub (https://github.com/phpuech/FindPeaks). Installation can be performed directly with the provided requirements.txt files to recreate the Python environment. A Python Jupyter notebook process has been designed to analyze time-series data from the experimentally obtained tab-separated values files, specifically focusing on deflection measurements over time. As illustrated in Figure 4, the pipeline begins by importing the time-resolved microindenter deflection and piezo displacement data (tab-separated format) and computing indentation over time by subtracting deflection from piezo displacement. To reduce noise and improve peak detection, it applies a mean filtering technique through the mean_filter_adaptive function, smoothing the deflection data. Once the data is processed, the function identifies peaks using the find_peaks method from the scipy.signal module. The method allows for adjustable parameters, such as peak prominence, which can be varied based on the shape of the peaks, and height constraints, where the maximum height is determined by the highest value recorded before the retraction of the microindenter. Following detection, rupture forces are computed by multiplying the calibrated microindenter deflection by its bending stiffness, and the resulting forces and rupture positions are stored in a pandas DataFrame, which also includes metadata such as the file path and folder name. For visualization, the code generates matplotlib/seaborn plot displaying the original deflection data, the smoothed signal, and the identified peaks, which are marked as red dots. To enhance readability, the plot is refined using seaborn.despine(), and appropriate labels and titles are added. The final plot is saved as a high-resolution.png file in a specified output folder. Ultimately, the code returns the DataFrame containing peak information and saves it as a.xlsx file, for further processing with ad-hoc plotting and statistical analysis secondary Jupyter-based code. Of note, we mainly use non parametric statistical tests since the data distribution is often observed to be non-Gaussian. Sketch summarizing the data processing strategy The most characteristic elements of the in-house developed automated Python code (smoothing, peak detection) are given. The obtained forces and positions of rupture events are further stored and used to build summary plots. In the boxplot, boxes represent the interquartile range, the horizontal line indicates the median, and whiskers correspond to 1.5× the interquartile range. Differences between conditions are explored using (non-parametric) statistical tests (see text). Limitations This microindentation technique, while precise, has several limitations. For Eimeria oocyst parasites, temperature control is less critical as they survive at lower temperatures. However, when working with other cell types that require physiological conditions, maintaining 37°C might be challenging. The inefficient heating of the large open dish and the resulting temperature gradients can cause unwanted objects motions. Temperature control represents an additional experimental challenge that would require substantial modifications to the current setup. Implementation would necessitate choosing between a simple heating plate approach versus full system thermalization, each presenting distinct complications. A heating plate would create temperature gradients and potential thermal drift, while complete system thermalization would require protecting sensitive electronics from heat exposure and potentially moisture if humidity is also controlled. Additionally, temperature elevation increases the risk of bubble formation in the tubing linking the micropipette holder and the reservoir, which would block the applied aspiration. Thermal expansion of system components could also introduce positional drift that would compromise the precision required for microindentation measurements. The current protocol prioritizes mechanical reproducibility at ambient temperature, though future developments could address these environmental control challenges. The setup is sensitive to environmental factors like vibrations and air currents. The use of a damping table or any vibration isolation system is highly recommended. The proximity of vents or aspiration systems as for air conditioning may introduce mechanical noise. The integrity of the microindenter is critical for accurate force measurements, and any damage can lead to calibration errors. Standard microindenter calibration is based on using a commercial force probe. In case of limited budget, an alternative method consists of using the weight of small pieces of wire or paper as known forces. In a previous study,7 we determined the density of a lightweight paper by measuring the mass of pieces whose surface area was measured under a microscope. We then calibrated reference microindenters by measuring their vertical deflection under known weights of paper pieces. This alternative technique, however, is less accurate and much more time-consuming than using a commercial force probe. The protocol presented here is optimized for specific coccidian parasites, and adaptation may be required for other microorganisms and cell types, especially regarding their diameter and the force applied to pierce them. When manipulating non-ovoidal parasites, several key parameters require adjustment. Researchers should adapt the pipette diameter to target locally spherical regions of the parasite to ensure optimal sealing and prevent leakage that would reduce aspiration efficiency. The approach geometry is equally critical aligning the holding micropipette axis with the indentation tip minimizes lateral torque that could cause the parasite to rotate or even escape from the pipette during manipulation. Regarding aspiration pressure, the holding force follows the relationship F=ΔP×π×Rpipette2, where ΔP is the pressure difference and Rpipette is the internal pipette radius. Therefore, when using pipettes of different diameters, the aspiration pressure must be adjusted accordingly to ΔP=F/(π×Rpipette2) to maintain equivalent holding force. This calculation assumes a tight seal between the parasite and pipette opening; any leakage will reduce the effective aspiration force below the theoretical maximum. For irregularly shaped parasites, researchers may empirically optimize these parameters through preliminary trials, as the contact geometry will influence both sealing efficiency and may also impact the mechanical response during indentation. Along with this, adjustment of the indenter’s bending stiffness may be necessary when working with particularly fragile or mechanically stiff cells. This can be achieved by selecting indenters with appropriate geometries to either decrease or increase stiffness as required. Finally, the technique measures only one oocyst at a time, requiring multiple measurements for statistical significance (typically 25–30 per experimental condition), potentially introducing sample-to-sample variability. Moreover, the manual nature of oocyst selection and positioning limits experimental throughput. Troubleshooting Problem 1 Glass enters inside the capillary during trimming, leading to a broken opening and aspiration system leakage (related to Step 2). Potential solution Ensure the cut is as straight as possible from all directions. This will help maintain the integrity of the capillary and prevent unwanted glass intrusion. If the section of the capillary becomes too close to the desired one or is larger, pull another needle and restart the process. Problem 2 The pipette bends irregularly or breaks during heating (related to Step 3, 6, and 7). Potential solution Reduce power or heating duration. Proceed with shorter, more gentle, and more numerous heating steps. If issues persist, restart with a new capillary. This approach allows for more controlled shaping of the pipette. Problem 3 Errors in calibration of the microindenter, specifically in Kref (reference microindenter bending stiffness), propagate to the calculated Kmi (related to Step 12). Potential solution Ensure accurate knowledge of the microindenter’s bending stiffness. Regularly calibrate the force transducer used to calibrate reference microindenters according to the manufacturer’s specifications. This will ensure and maintain the accuracy of force measurements in the entire experiment set. Problem 4 Bubbles get stuck in the micropipette (in particular in very thin and/or the curved regions), syringe, 4-way valve, or meniscus of the pipette (related to Step 14, 15, and 16). Potential solution Gently flick the syringe with your fingernail to dislodge bubbles. If passing to a positive pressure by raising the water reservoir does not stop aspiration, it’s likely due to a bubble in the micropipette holder or tubing. In such cases, consider replacing the micropipette or cutting its back a few millimeters with a diamond tip pen. Problem 5 Objects, biological or not, such as preparation remnants, adhere to the side walls of the pipette (related to Step 19). Potential solution Increase BSA concentration in the buffer to 2%–3%. This enhances surface passivation and reduces unwanted adhesion. Be careful then with air bubbles, which are a bit more stabilized hence prone to form. Problem 6 The oocyst wall and microindenter are not in the same focus during acquisition, causing the microindenter to deflect or slide on the wall (related to Step 26). Potential solution Monitor the focus before acquisition to ensure both objects are aligned. Make slight focus adjustments during acquisition if necessary. This ensures accurate interaction between the microindenter and the sample. Problem 7 Poor indenter deflection signal quality due to setup vibrations (related to Step 25). Potential solution Avoid or reduce vibrations for better signal quality by using damping tables and not touching the setup during measurements, especially when using soft microindenters and high-magnification objectives. Avoid tension in wires, particularly in heavy camera cables. Minimize objects with fans present or touching on the insulating table. Consider enclosing the setup in an acoustic insulation case, though this may reduce accessibility. Avoid proximity to air conditioning and open windows in the room. Problem 8 The notebook cannot read the input file, or the columns “time” or “deflection” are not detected (related to the “expected outcomes” section). Potential solution Verify that the file path is correct and that the file exists. Ensure the tab-separated format is consistent. Check the file headers for typos, extra spaces, or mismatched names, and rename columns as needed. Problem 9 Peaks are flattened or missed after smoothing, spurious peaks appear due to noise, or true peaks are not detected during peak identification (related to the “expected outcomes” section). Potential solution Adjust the smoothing window and modify find_peaks parameters such as prominence, height, width, and distance. Always visually inspect the plots to confirm that detected peaks correspond to true events and not artifacts, especially when peak values are low. Resource availability Lead contact Further information and requests for resources and reagents should be directed to and will be fulfilled by the lead contact, Pierre-Henri Puech (pierre-henri.puech@univ-amu.fr). Technical contact Technical questions on executing this protocol should be directed to and will be answered by the technical contact, Julien Husson (julien.husson@polytechnique.edu) and Jana El Husseiny (jana.elhusseiny@univ-amu.fr). Materials availability This study did not generate new unique reagents. Data and code availability •Data sets used here can be obtained from the authors upon request. Data sets used here can be obtained from the authors upon request. •Software can be obtained from online resources (for Fiji/ImageJ, anaconda python, see text). Python codes can be found online on Zenodo (https://doi.org/10.5281/zenodo.17467539). Software can be obtained from online resources (for Fiji/ImageJ, anaconda python, see text). Python codes can be found online on Zenodo (https://doi.org/10.5281/zenodo.17467539). Acknowledgments We sincerely thank Anne Silvestre and Laura Sedano (INRAE Nouzilly) for providing Eimeria tenella parasites and Jean-Michel Répérant (ANSES Ploufragan) for supplying Eimeria acervulina parasites. We also extend our thanks to Rémy Torro for his valuable assistance with coding. This work was supported by PhD funding to J.E.H. from the ANR project “BreakingTheWall” (ANR-22-CE35-0008). Author contributions J.E.H. performed the experimental work, analyzed the data, and wrote and reviewed the protocol. P.-H.P. designed the research study, analyzed the data, and reviewed the protocol. A.D. designed the research study, contributed materials and funding, and reviewed the protocol. J.H. designed the research study, developed the initial setup, contributed materials, performed experimental work, analyzed the data, and reviewed the protocol. Declaration of interests The authors declare no competing interests. Footnotes Supplemental information can be found online at https://doi.org/10.1016/j.xpro.2025.104234. Contributor Information Jana El Husseiny, Email: jana.elhusseiny@univ-amu.fr. Pierre-Henri Puech, Email: pierre-henri.puech@univ-amu.fr. Aurélien Dumètre, Email: aurelien.dumetre@univ-amu.fr. Julien Husson, Email: julien.husson@polytechnique.edu. References Associated Data Supplementary Materials Methods video S1. Video captures the mechanical indentation of an Eimeria tenella oocyst wall using a microindenter, related to “expected outcomes” section The left panel shows a time-synchronized bright-field microscopic view of the indentation process, with a scale bar indicating 10 μm. The right panel presents a force-time graph, where the rupture force (μN) is plotted against time (seconds). The force increases as the probe indents the oocyst wall, with sudden drops indicating rupture events. The correlation between the video and the graph provides insights into the biomechanical properties of the oocyst wall. Data Availability Statement •Data sets used here can be obtained from the authors upon request. Data sets used here can be obtained from the authors upon request. •Software can be obtained from online resources (for Fiji/ImageJ, anaconda python, see text). Python codes can be found online on Zenodo (https://doi.org/10.5281/zenodo.17467539). Software can be obtained from online resources (for Fiji/ImageJ, anaconda python, see text). Python codes can be found online on Zenodo (https://doi.org/10.5281/zenodo.17467539).