Performance Evaluation of Dry Chemical Colorimetry for Plasma Ammonia Detection on the VITROS 350 Chemistry System
Jia Liu, Lin Wang, Yue Tang, Peiran Li, Lili Zhao, Wenwen Yuan, Bo Li, Boan Li · 2026
Read Performance Evaluation of Dry Chemical Colorimetry for Plasma Ammonia Detection on the VITROS 350 Chemistry System by Jia Liu, Lin Wang, Yue Tang, Peiran Li, Lili Zhao, Wenwen Yuan, Bo Li, Boan Li (2026) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/pmc-13662657.txt.
Description
Jia Liu and co-authors check how well a lab machine measures ammonia in blood plasma. High ammonia can signal severe liver trouble, so labs need results they can trust. Their 2026 study tests the VITROS 350 system, which uses dry chemical colorimetry. The checks follow rules set by lab standards groups in the United States and China.
The team tested precision, accuracy, range, and carryover between samples. All met the set limits and the maker's claims. Results also matched a larger VITROS system across five labs. One limit stood out: very high hemoglobin in a sample can skew the reading. The paper gives labs a firm base for using the machine in daily work.
Catalog record
| CREATOR | Jia Liu, Lin Wang, Yue Tang, Peiran Li, Lili Zhao, Wenwen Yuan, Bo Li, Boan Li |
|---|---|
| YEAR | 2026 |
| COLLECTION | PAPERS |
| FORMAT | TXT |
| FILE SIZE | 32 KB |
| VERSION | 1.0 |
| LICENSE | CC BY 4.0 |
| RIGHTS | CC BY |
| SOURCE | Europe PMC |