Early Catastrophic Failure of Highly-Cross-Linked Polyethylene Liner Due to Acetabular Shell Screw Placement: A Case Report

David Dallas-Orr, Gregory J Harbison, Samuel K Simister, Shannon Tse, John Meehan · 2026

Read Early Catastrophic Failure of Highly-Cross-Linked Polyethylene Liner Due to Acetabular Shell Screw Placement: A Case Report by David Dallas-Orr, Gregory J Harbison, Samuel K Simister, Shannon Tse, John Meehan (2026) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/pmc-13633924.txt.

Description

David Dallas-Orr, Gregory J Harbison, Samuel K Simister, Shannon Tse, and John Meehan report a hip liner shattered for a preventable reason. Cross-linked polyethylene liners resist wear but forgive concentrated stress less, and failures follow trauma. Their patient, a sixty-two-year-old woman, arrived twenty months after hip replacement with groin and hip pain, a squeaking joint, and a hip unable to bend past ninety degrees.

Scans showed the ball sitting off-center in the socket and the locking ring displaced. At revision, the liner lay in fragments amid metal debris. The cause stood out: an off-center socket screw poked through the shell's inner surface where the liner's back side had worn. The team removed the screw and exchanged the liner. Eight years later she walks without difficulty or pain. The authors urge surgeons to confirm screws sit flush before liner placement.

Catalog record

CREATORDavid Dallas-Orr, Gregory J Harbison, Samuel K Simister, Shannon Tse, John Meehan
YEAR2026
COLLECTIONPAPERS
FORMATTXT
FILE SIZE12 KB
VERSION1.0
LICENSECC BY 4.0
RIGHTSCC BY
SOURCEEurope PMC

Source: https://europepmc.org/articles/PMC13633924