Continuous monitoring of endotracheal tube obstructions using naturally occurring pressure and flow oscillations
Ben Fabry, Christian Kuster, Roland C. E. Francis · 2026
Read Continuous monitoring of endotracheal tube obstructions using naturally occurring pressure and flow oscillations by Ben Fabry, Christian Kuster, Roland C. E. Francis (2026) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/PMC13624231.txt.
Description
Ben Fabry, Christian Kuster, and Roland C Francis present a way to watch a breathing tube clog up in real time, using only the pressure and flow signals a ventilator measures. Tubes narrow as mucus and biofilm build inside them, adding breathing effort and risking pressure build-up, yet during assisted spontaneous breathing no check catches the change early.
The method filters the natural wiggles in airway pressure and flow, and strips out the slow push of the patient's own muscles and lung stretch. It then solves the motion equations to recover the tube's resistance, split into steady and flow-dependent parts. Across ventilation modes, tube sizes, filters, and breathing patterns, the recovered values stayed within ten percent of the truth with small bias. The estimates update breath by breath, so a rising blockage shows as it forms. The authors see uses in automatic compensation and safer weaning.
Catalog record
| CREATOR | Ben Fabry, Christian Kuster, Roland C. E. Francis |
|---|---|
| YEAR | 2026 |
| COLLECTION | PAPERS |
| FORMAT | TXT |
| FILE SIZE | 25 KB |
| LICENSE | https://creativecommons.org/licenses/by/4.0 |
| RIGHTS | CC BY |
| SOURCE | Europe PMC |