A Portable Electrical Response Verification System for Blood Glucose Meters Based on Dynamic Microcurrent Reproduction
Wenshu Yue, Yifan Li, Jinhui Cai · 2026
Read A Portable Electrical Response Verification System for Blood Glucose Meters Based on Dynamic Microcurrent Reproduction by Wenshu Yue, Yifan Li, Jinhui Cai (2026) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/pmc-13611197.txt.
Description
Wenshu Yue, Yifan Li, and Jinhui Cai build a portable device that verifies blood glucose meters electrically. It needs no test strips or control solutions. Standard checks test the meter, strip, and reagent as one whole. So they can not isolate the meter itself with a controlled stimulus. Their system instead reproduces dynamic microcurrents. It uses a calibrated programmable current source.
Real strip responses at two levels were distilled into mean waveforms. Each became a code table of a hundred points. A microcontroller then reproduced each waveform on cue. The reproductions matched the originals almost perfectly. Peak current varied only a little. Readings on the matching commercial meter differed from the original readings by less than about two percent. The results support controlled electrical checks of strip-based meters. They also support short-term repeatability checks. Such a device could split instrument drift from strip chemistry problems.
Catalog record
| CREATOR | Wenshu Yue, Yifan Li, Jinhui Cai |
|---|---|
| YEAR | 2026 |
| COLLECTION | PAPERS |
| FORMAT | TXT |
| FILE SIZE | 46 KB |
| VERSION | 1.0 |
| LICENSE | CC BY 4.0 |
| RIGHTS | CC BY |
| SOURCE | Europe PMC |