Read Use of TriClip steerable guide catheter for MitraClip: from bench to case by Fuk Kei Fong, Leo Kar Lok Lai, Kevin Ka Ho Kam, Alex Pui Wai Lee, Gilbert H L Tang, Kent Chak-yu So (2026) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/pmc-13582084.txt.
Description
Fuk Kei Fong and colleagues tell of a 77-year-old frail man with bad leaks in both the mitral and tricuspid heart valves who had keyhole valve repair. The MitraClip tool had to be pushed deep after a high puncture through the wall between the atria, which hurt control and grip on the leaflets. The team then used the TriClip guide tube off-label for the mitral job.
A turn of the tube's L-knob cut the height above the mitral valve and gave back fine steering, so the clip landed well. Mitral leak fell from bad to mild-moderate, and tricuspid leak fell from bad to mild. Bench tests show the TriClip tube's short bent tip needs 5 millimetres less puncture height, a real help when the puncture site is not ideal.