Case Report: Electrode perforation with normal pacemaker parameters leading to hemorrhagic shock: a rare case report of the “penetration-exit-re-entry” phenomenon
Zuoan Qin, Jian Yang, Li Luo, Shi Chen, Fei Yang · 2026
Read Case Report: Electrode perforation with normal pacemaker parameters leading to hemorrhagic shock: a rare case report of the “penetration-exit-re-entry” phenomenon by Zuoan Qin, Jian Yang, Li Luo, Shi Chen, Fei Yang (2026) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/pmc-13581696.txt.
Description
Zuoan Qin and four co-authors present a rare pacemaker case in two thousand twenty-six. A seventy-year-old woman got a dual-chamber pacemaker for a complete heart block. All checks on the leads during surgery read normal, which usually signals a safe placement.
Right after surgery, she bled heavily into her chest and went into shock, with low oxygen. A scope exam found the lead had pierced her subclavian vein, left the vessel, and re-entered another vein. That path caused steady internal bleeding. Emergency drainage, vein repair, and transfusion saved her, and she recovered fully. At one month, the pacemaker worked well. The authors warn that normal readings cannot rule out this hidden injury.
Catalog record
| CREATOR | Zuoan Qin, Jian Yang, Li Luo, Shi Chen, Fei Yang |
|---|---|
| YEAR | 2026 |
| COLLECTION | PAPERS |
| FORMAT | TXT |
| FILE SIZE | 25 KB |
| VERSION | 1.0 |
| LICENSE | CC BY 4.0 |
| RIGHTS | CC BY |
| SOURCE | Europe PMC |