Transcatheter aortic valve canting creates cusp-specific thrombogenic microenvironments through altered sinus hemodynamics
Thangam Natarajan, Jae Hyun Kim, Christian Salgado, Akshita Jha, Cole A. Baker, Stephanie L. Sellers, Joseph E. Aslan, Monica T. Hinds, Ajit P. Yoganathan, Lakshmi Prasad Dasi · 2026
Read Transcatheter aortic valve canting creates cusp-specific thrombogenic microenvironments through altered sinus hemodynamics by Thangam Natarajan, Jae Hyun Kim, Christian Salgado, Akshita Jha, Cole A. Baker, Stephanie L. Sellers, Joseph E. Aslan, Monica T. Hinds, Ajit P. Yoganathan, Lakshmi Prasad Dasi (2026) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/PMC13524589.txt.
Description
Natarajan and colleagues published a 2026 study on how a tilted heart valve stirs clot risk. Replacement aortic valves delivered by catheter can land slightly canted. Leaflet clots have emerged as a worry after such implants in patients. The team tested three tilt angles in a patient-shaped aortic root model.
Canting reshaped swirl and washout in the valve's pockets unevenly. The tightest pocket held the stalest blood in every tilted setup. One tilt direction left 92 percent of one pocket nearly still. Clots in blood-loop tests formed exactly on those poorly washed leaflets. Overall valve function looked alike, hiding the local clot risk well. Deployment angle deserves the same care as valve size choice. Blood-loop tests used real whole blood flowing past the valve leaflets.
Catalog record
| CREATOR | Thangam Natarajan, Jae Hyun Kim, Christian Salgado, Akshita Jha, Cole A. Baker, Stephanie L. Sellers, Joseph E. Aslan, Monica T. Hinds, Ajit P. Yoganathan, Lakshmi Prasad Dasi |
|---|---|
| YEAR | 2026 |
| COLLECTION | PAPERS |
| FORMAT | TXT |
| FILE SIZE | 47 KB |
| LICENSE | https://creativecommons.org/licenses/by/4.0 |
| RIGHTS | CC BY |
| SOURCE | Europe PMC |