Authors’ response: balloon size and interpretation of the REBOARREST trial

Jostein Rødseth Brede, Andreas Jørstad Krüger · 2026

Read Authors’ response: balloon size and interpretation of the REBOARREST trial by Jostein Rødseth Brede, Andreas Jørstad Krüger (2026) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/pmc-13505167.txt.

Description

Jostein Rødseth Brede and Andreas Jørstad Krüger answered critics of their cardiac arrest trial in a 2026 response. The REBOARREST trial tested a balloon catheter that blocks the aorta from inside during resuscitation for out-of-hospital cardiac arrest. A comment questioned balloon sizing and trial conduct. The authors reply point by point, starting with approvals: all ethics, institutional, and device permissions were in place before the trial began.

They concede one limit. Patients excluded for missing deferred consent may skew the sample, since consent failure tracks with early death; the rule came from ethics committees, not the team. On sizing, they set the cited thirty-millimeter aorta estimate against pooled data nearer twenty-four to twenty-six millimeters.

Catalog record

CREATORJostein Rødseth Brede, Andreas Jørstad Krüger
YEAR2026
COLLECTIONPAPERS
FORMATTXT
FILE SIZE7 KB
VERSION1.0
LICENSECC BY 4.0
RIGHTSCC BY
SOURCEEurope PMC