Comparative analysis of hemoglobin a1c measurement methods: inter-device agreement and clinical impact
Ergin Taşkin, Ebubekir Bakan, Dilek Taşkin, İrem Arslantürk · 2026
Read Comparative analysis of hemoglobin a1c measurement methods: inter-device agreement and clinical impact by Ergin Taşkin, Ebubekir Bakan, Dilek Taşkin, İrem Arslantürk (2026) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/pmc-13478365.txt.
Description
Taşkin and colleagues compared three laboratory methods for measuring HbA1c in 2026, the blood marker that summarizes average sugar control in diabetes and guides treatment changes. Small reading differences can flip clinical decisions, so agreement between machines matters. Fifty patient blood samples were run on all three systems: a chromatography instrument, a capillary electrophoresis instrument, and a general autoanalyzer using an antibody-based method.
Electrophoresis and chromatography agreed almost perfectly, with a correlation of 0.998 and an average bias of only 0.07 points between them. The autoanalyzer correlated strongly too, at 0.997, but its bias was larger, at 0.52 points against electrophoresis, and the gap widened at higher sugar-control values. The authors warn that autoanalyzer readings near decision thresholds may mislead, while noting the small sample of 50 means the warning needs confirmation in larger work.
Catalog record
| CREATOR | Ergin Taşkin, Ebubekir Bakan, Dilek Taşkin, İrem Arslantürk |
|---|---|
| YEAR | 2026 |
| COLLECTION | PAPERS |
| FORMAT | TXT |
| FILE SIZE | 25 KB |
| VERSION | 1.0 |
| LICENSE | CC BY 4.0 |
| RIGHTS | CC BY |
| SOURCE | Europe PMC |