Piezoelectric shaping of radical mastoid cavity and dissection of ossified cochlea prior to cochlear implantation
Samiah S Al-Angari, Marc Bloching, Katharina Stölzel, Stefan Lyutenski · 2026
Read Piezoelectric shaping of radical mastoid cavity and dissection of ossified cochlea prior to cochlear implantation by Samiah S Al-Angari, Marc Bloching, Katharina Stölzel, Stefan Lyutenski (2026) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/pmc-13476123.txt.
Description
Samiah Al-Angari and co-authors present a 2026 case on hearing implant surgery in a hard ear cavity. Ears that had radical mastoid surgery pose a risk that the implant wire slips out later. The patient was a 72 year old woman with deep hearing loss in her left ear and a past open cavity surgery.
The team used a piezoelectric tool, which cuts bone with fine vibration, to shape the cavity in one stage. Angled tips let them clear bone near the facial nerve and carve a slit with an overhang to lock the lead. Ossified parts of the inner ear channel were cleared for a full insertion. At eight months, no sign of wire drift showed in clinic or electric tests.
Catalog record
| CREATOR | Samiah S Al-Angari, Marc Bloching, Katharina Stölzel, Stefan Lyutenski |
|---|---|
| YEAR | 2026 |
| COLLECTION | PAPERS |
| FORMAT | TXT |
| FILE SIZE | 11 KB |
| VERSION | 1.0 |
| LICENSE | CC BY 4.0 |
| RIGHTS | CC BY |
| SOURCE | Europe PMC |