Valve-in-Valve TAVR for Degenerated Perceval Valves

Mouna Kodali, Rajesh Gutpa, Ahmad Mahdi, Omar Sajdeya, Ashish Yadav, George V Moukarbel · 2026

Read Valve-in-Valve TAVR for Degenerated Perceval Valves by Mouna Kodali, Rajesh Gutpa, Ahmad Mahdi, Omar Sajdeya, Ashish Yadav, George V Moukarbel (2026) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/pmc-13464266.txt.

Description

Mouna Kodali and colleagues report two 2026 cases of valve-in-valve replacement inside failed sutureless Perceval valves, whose ring design complicates repeat implants. The Perceval frame sits above the annulus, lacks X-ray markers, and can trap delivery gear when wires cross off-centre. Both patients received balloon-expandable Sapien 3 Ultra valves for degenerated implants.

In the first case, standard crossing trapped the delivery system, demanding balloon shifts and snaring to free it. In the second, a forward route through the septum with a wire loop gave precise central crossing and a smooth implant. Both patients recovered well with symptoms eased, and the authors judge the forward route safer and faster when standard crossing fails.