Biocompatibility of a novel directly printed orthodontic aligner at variable material thicknesses

Maximilian Bleilöb, Claudia Welte-Jzyk, Vanessa Knode, Björn Ludwig, Julia Bulski, Christina Erbe · 2026

Read Biocompatibility of a novel directly printed orthodontic aligner at variable material thicknesses by Maximilian Bleilöb, Claudia Welte-Jzyk, Vanessa Knode, Björn Ludwig, Julia Bulski, Christina Erbe (2026) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/PMC13421538.txt.

Description

Maximilian Bleilöb and five colleagues test, in a 2026 study, whether directly 3D-printed teeth aligners are safe against gum tissue at thicknesses well beyond the standard. Printed aligners let orthodontists make appliances in-office and vary thickness by region, yet thicker parts might cure less completely under the standard twenty-minute ultraviolet post-cure.

Discs from half a millimetre to six millimetres thick were printed in a dental resin through the manufacturer's closed workflow. Dimensions matched their targets closely. In a standard cell-viability test on human gum fibroblasts, every thickness stayed above the seventy percent threshold that defines non-cytotoxic material, with no meaningful difference from the standard thickness. The authors conclude the standard cure suffices up to six millimetres.

Catalog record

CREATORMaximilian Bleilöb, Claudia Welte-Jzyk, Vanessa Knode, Björn Ludwig, Julia Bulski, Christina Erbe
YEAR2026
COLLECTIONPAPERS
FORMATTXT
FILE SIZE26 KB
LICENSEhttps://creativecommons.org/licenses/by/4.0
RIGHTSCC BY
SOURCEEurope PMC

Source: https://europepmc.org/article/MED/PMC13421538

Browse the full library · About the Tower of Babel