Massive tricuspid regurgitation caused by iatrogenic papillary muscle rupture during leadless pacemaker retrieval: a case report

Hikaru Tanemura, Noriaki Moriyama, Masato Murakami, Shigeru Saito · 2026

Read Massive tricuspid regurgitation caused by iatrogenic papillary muscle rupture during leadless pacemaker retrieval: a case report by Hikaru Tanemura, Noriaki Moriyama, Masato Murakami, Shigeru Saito (2026) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/pmc-13403278.txt.

Description

Tanemura and co-authors report, in 2026, a severe complication of leadless pacemaker implantation. During retrieval of the device in an 86-year-old man treated for heart block and syncope, the heart's papillary muscle was caught between the delivery sheath and the pacemaker. It tore away, causing massive leakage through the tricuspid valve. Leadless devices avoid many lead-related problems, but effects on the tricuspid valve were known; this rupture route was new.

Resistance during removal was the warning sign of the entrapment, a lesson the authors stress for operators. The patient then developed worsening heart failure needing non-invasive ventilation and intravenous furosemide at 20 milligrams twice daily, and imaging showed sharply reduced left ventricular function with an ejection fraction of 26.7 percent in follow-up studies. He received a transvenous pacemaker, recovered on medication without valve surgery, and transferred to rehabilitation, a benign ending the authors document.

Catalog record

CREATORHikaru Tanemura, Noriaki Moriyama, Masato Murakami, Shigeru Saito
YEAR2026
COLLECTIONPAPERS
FORMATTXT
FILE SIZE10 KB
VERSION1.0
LICENSECC BY 4.0
RIGHTSCC BY
SOURCEEurope PMC

Source: https://europepmc.org/articles/PMC13403278