Mitigating Iatrogenic Air Entrainment During Pulsed Field Ablation

Ting-Yung Chang, Chin-Yu Lin, Pei-Yi Wang, Hsin-Lei Huang, Kuo-Hsiang Lu · 2026

Read Mitigating Iatrogenic Air Entrainment During Pulsed Field Ablation by Ting-Yung Chang, Chin-Yu Lin, Pei-Yi Wang, Hsin-Lei Huang, Kuo-Hsiang Lu (2026) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/pmc-13349500.txt.

Description

Ting-Yung Chang and co-authors tested in 2026 a plain device, the Adapter Sink, to stop air bubbles from entering the bloodstream during pulsed field treatment for atrial fibrillation. Systemic air embolism is a vital risk during left atrial work, most of all with frequent catheter swaps through small sheaths.

Using a water-tank model with a commercial pulsed field system and guiding sheath, the team weighed three entry ways across ten trials each. One was standard entry with no suction. One was entry with manual suction. One was entry through the sunk Adapter Sink. Standard entry always brought small bubbles. Manual suction cut them markedly. Sunk entry wiped them out. The cheap adapter gives a workable safety fix for a routine but risky procedural step.