Comment on Maga et al. The Effect of Selective Occlusal Adjustment on the Disclusion Time Reduction and Symmetry of Occlusal Contacts of the Own Dentition Using Digital Occlusion Analysis in Patients with Temporomandibular Disorders. J. Clin. Med. 2025, 14, 7007

Robert B. Kerstein · 2026

Read Comment on Maga et al. The Effect of Selective Occlusal Adjustment on the Disclusion Time Reduction and Symmetry of Occlusal Contacts of the Own Dentition Using Digital Occlusion Analysis in Patients with Temporomandibular Disorders. J. Clin. Med. 2025, 14, 7007 by Robert B. Kerstein (2026) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/PMC13301486.txt.

Description

A widely used bite-force scanner can flatter its own readings if the operator skips one setup step, Robert B. Kerstein warns in a 2026 comment on a study by Maga and colleagues. The target paper used T-Scan digital analysis to show that selective occlusal adjustment shortened disclusion time and balanced contacts in patients with temporomandibular disorders.

Kerstein argues the team never set per-patient sensor sensitivity before recording, so maximum bites oversaturated the sensors. Their figures show clusters of pink, saturated cells, which inflate force percentages and fake balanced distributions, he writes. Proper calibration keeps only one to three pink cells at full bite. The comment urges readers to discount the reported gains until recording technique is fixed.

Catalog record

CREATORRobert B. Kerstein
YEAR2026
COLLECTIONPAPERS
FORMATTXT
FILE SIZE19767
LICENSEhttps://creativecommons.org/licenses/by/4.0
RIGHTSCC BY
SOURCEEurope PMC

Source: https://europepmc.org/article/MED/PMC13301486

Browse the full library · About the Tower of Babel