Read Phantom Quantification of Magnetoencephalography Source Imaging Distortion Caused by Deep Brain Stimulation by Saar Kariv, Jeong Woo Choi, Amy L. Proskovec, Mahak Virlley, Tyrell Pruitt, Nader Pouratian, Elizabeth M. Davenport (2026) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/pmc-13297442.txt.

Description

Saar Kariv and colleagues measure how much deep brain stimulation warps magnetic brain imaging, using a head phantom, in a 2026 study. Magnetoencephalography maps brain activity from faint magnetic fields, and patients with implanted stimulators form a hard case for it. Past work looked at sensor noise, not at errors in locating the brain sources themselves.

Known test sources were placed at set distances from a stimulator in the phantom, across strengths and electrode layouts. Location and angle errors stayed modest and grew as the source neared the electrode, while the stimulation on or off made little difference overall. Head-position signals stayed steady throughout. The setup gives labs a yardstick for distortion from this and other implants.

Catalog record

CREATORSaar Kariv, Jeong Woo Choi, Amy L. Proskovec, Mahak Virlley, Tyrell Pruitt, Nader Pouratian, Elizabeth M. Davenport
YEAR2026
COLLECTIONPAPERS
FORMATTXT
FILE SIZE25 KB
VERSION1.0
LICENSECC BY 4.0
RIGHTSCC BY
SOURCEEurope PMC

Source: https://europepmc.org/articles/PMC13297442

Browse the full library · About the Tower of Babel