Clinical evaluation of a new fully automated biometer integrating optical low-coherence reflectometry and Placido-disk topography
Yihong Chen, Xinning Yang, Zheng Li, Bowen Tan, Kexin Li, Giacomo Savini, Domenico Schiano-Lomoriello, Xingtao Zhou, Chak Seng Lei, Jinhai Huang · 2026
Read Clinical evaluation of a new fully automated biometer integrating optical low-coherence reflectometry and Placido-disk topography by Yihong Chen, Xinning Yang, Zheng Li, Bowen Tan, Kexin Li, Giacomo Savini, Domenico Schiano-Lomoriello, Xingtao Zhou, Chak Seng Lei, Jinhai Huang (2026) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/pmc-13224415.txt.
Description
Yihong Chen and colleagues test a 2026 eye-measuring device, the AL550, designed to size up eyes before lens surgery. Accurate sizing of eye length and corneal shape decides which lens a patient gets. The new machine combines two optical techniques in one automated unit, and the team tested it against a widely used standard device.
Healthy volunteers numbered 112, each measured many times by two operators. Repeat readings agreed closely, with scores above 0.939 on every test. Agreement with the standard machine was strong for most values, though corneal diameter showed a wider gap. Overall, the device delivered stable, steady readings suitable for clinical use.
Catalog record
| CREATOR | Yihong Chen, Xinning Yang, Zheng Li, Bowen Tan, Kexin Li, Giacomo Savini, Domenico Schiano-Lomoriello, Xingtao Zhou, Chak Seng Lei, Jinhai Huang |
|---|---|
| YEAR | 2026 |
| COLLECTION | PAPERS |
| FORMAT | TXT |
| FILE SIZE | 35 KB |
| VERSION | 1.0 |
| LICENSE | CC BY 4.0 |
| RIGHTS | CC BY |
| SOURCE | Europe PMC |