Increasing efficiency in the ACR Digital Mammography QC Manual's DBT Z‐resolution evaluation

Ryan F Fisher, Nicholas B Bevins, Katie W Hulme, Colin Schaeffer · 2026

Read Increasing efficiency in the ACR Digital Mammography QC Manual's DBT Z‐resolution evaluation by Ryan F Fisher, Nicholas B Bevins, Katie W Hulme, Colin Schaeffer (2026) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/pmc-13179838.txt.

Description

Ryan F Fisher, Nicholas B Bevins, Katie W Hulme, and Colin Schaeffer report a two thousand twenty-six study on faster mammography quality checks. The standard manual asks staff to measure many small regions across phantom image slices. That work can mean dozens of measurements per unit.

The team reanalyzed hundreds of surveys from many mammography units. They compared the full multi-speck score with scores from one or two specks alone. Two-speck scores stayed closer to the full result than single-speck ones. Readers in imaging quality control get evidence on trimming a tedious test without drifting far from the manual's baseline.

Catalog record

CREATORRyan F Fisher, Nicholas B Bevins, Katie W Hulme, Colin Schaeffer
YEAR2026
COLLECTIONPAPERS
FORMATTXT
FILE SIZE37 KB
VERSION1.0
LICENSECC BY 4.0
RIGHTSCC BY
SOURCEEurope PMC

Source: https://europepmc.org/articles/PMC13179838