An Experimental Measurement Method to Characterize and Apply Platinum Silicon Material for a Biomechanical Replica of the Thoracic Aorta

Mario Alberto Grave-Capistrán, Francesco Lamonaca, Giuseppe Carbone, Christopher René Torres-SanMiguel · 2026

Read An Experimental Measurement Method to Characterize and Apply Platinum Silicon Material for a Biomechanical Replica of the Thoracic Aorta by Mario Alberto Grave-Capistrán, Francesco Lamonaca, Giuseppe Carbone, Christopher René Torres-SanMiguel (2026) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/pmc-13113696.txt.

Description

Mario Alberto Grave-Capistrán and co-authors test, in a 2026 study, whether an extra-soft platinum silicone can mimic the chest aorta for medical models. Silicone is widely used in such replicas, but its stretch must match real tissue. The team molded samples and ran standard pull and squeeze tests, tracking surface strain with cameras.

The material stretched more than ten times its length before breaking, far beyond most rubbers. Its stiffness values landed in the range of a healthy chest aorta. The team then brushed the silicone into a full arch replica with close to real length and vessel widths. Such replicas can serve surgical practice and device testing without animal tissue.

Catalog record

CREATORMario Alberto Grave-Capistrán, Francesco Lamonaca, Giuseppe Carbone, Christopher René Torres-SanMiguel
YEAR2026
COLLECTIONPAPERS
FORMATTXT
FILE SIZE23 KB
VERSION1.0
LICENSECC BY 4.0
RIGHTSCC BY
SOURCEEurope PMC