Deviation between programmed and actual capsulotomy diameter in femtosecond laser–assisted cataract surgery: a clinical study of influencing factors and predictive modeling

Zhao Tan, Fei Yuan · 2025

Read Deviation between programmed and actual capsulotomy diameter in femtosecond laser–assisted cataract surgery: a clinical study of influencing factors and predictive modeling by Zhao Tan, Fei Yuan (2025) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/pmc-12837011.txt.

Description

In laser cataract surgery, the machine is told how wide to cut the eye's lens capsule, yet the real opening often ends up bigger. Zhao Tan and Fei Yuan measure that gap in forty-seven eyes and hunt for its causes. They read the true opening size from surgery videos and compare it with the programmed size for each eye.

The real openings averaged about 6.0 millimeters against 5.67 programmed, a gap near a third of a millimeter. In most eyes the gap passed a fifth of a millimeter. Age stood out: older eyes drifted less, and the link survived adjustment for other eye measures. A formula using age and the programmed size predicts the real size. The authors offer it to help surgeons plan a more exact cut.

Catalog record

CREATORZhao Tan, Fei Yuan
YEAR2025
COLLECTIONPAPERS
FORMATTXT
FILE SIZE28 KB
VERSION1.0
LICENSECC BY 4.0
RIGHTSCC BY
SOURCEEurope PMC

Source: https://europepmc.org/articles/PMC12837011

Browse the full library · About the Tower of Babel