Quantitative validation of data-driven motion correction for brain PET using phantom with motion generator system
Yuto Kamitaka, Muneyuki Sakata, Keiichi Oda, Akie Katsuki, Hirofumi Kawakami, Kei Wagatsuma, Masato Kobayashi, Kenji Ishii · 2025
Read Quantitative validation of data-driven motion correction for brain PET using phantom with motion generator system by Yuto Kamitaka, Muneyuki Sakata, Keiichi Oda, Akie Katsuki, Hirofumi Kawakami, Kei Wagatsuma, Masato Kobayashi, Kenji Ishii (2025) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/pmc-12824033.txt.
Description
Kamitaka and co-authors validate a data driven method that corrects for head motion in brain PET scans, using a phantom head on a motion machine. Motion blurs scans and skews their numbers, and patient data cannot supply exact truth for testing. Their 2025 study fills a standard brain phantom with a radioactive tracer at two levels and moves it in set ways.
The machine reproduced the planned moves faithfully, including tilts and shifts. Motion estimates stayed stable at the higher tracer level but wobbled at the lower level, above all with short estimation windows. Longer windows understated steady motion. Correction improved image contrast and uniformity in all cases. Short windows fixed fast motion best but added noise when tracer was low.
Catalog record
| CREATOR | Yuto Kamitaka, Muneyuki Sakata, Keiichi Oda, Akie Katsuki, Hirofumi Kawakami, Kei Wagatsuma, Masato Kobayashi, Kenji Ishii |
|---|---|
| YEAR | 2025 |
| COLLECTION | PAPERS |
| FORMAT | TXT |
| FILE SIZE | 35 KB |
| VERSION | 1.0 |
| LICENSE | CC BY 4.0 |
| RIGHTS | CC BY |
| SOURCE | Europe PMC |