Accuracy of Portable Ultrasound Measurement of the Bladder Volume Using the Manual Fanning Scan and Traditional Motorized Scan Methods
Hyun Ju Jeong, Seung-June Oh · 2025
Read Accuracy of Portable Ultrasound Measurement of the Bladder Volume Using the Manual Fanning Scan and Traditional Motorized Scan Methods by Hyun Ju Jeong, Seung-June Oh (2025) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/pmc-12784057.txt.
Description
Hyun Ju Jeong and Seung-June Oh compare two ways of measuring bladder volume with a portable ultrasound scanner in a 2025 study. One scanner sweeps the probe by motor, while the other asks the user to fan the probe by hand. The patients were adults with lower urinary tract symptoms seen at a urology clinic in late 2024.
Data came from thirty-nine patients, most of them men, with an average age near sixty-five. Both scans gave mean volumes very close to the true volume, found by adding voided urine to what a tube drained after voiding. The motor scan tracked the true value a bit more closely. Still, the authors find the hand-fanned scan about as accurate in practice.