Comparison of three rebound tonometers in rabbits
Chenghai Wang, Yunhao Su, Fushan Shi, Ji Fu · 2025
Read Comparison of three rebound tonometers in rabbits by Chenghai Wang, Yunhao Su, Fushan Shi, Ji Fu (2025) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/pmc-12749333.txt.
Description
Chenghai Wang and co-authors present a 2025 comparison of three devices that measure pressure inside the eye. Rebound tonometers tap the cornea with a tiny probe and read the bounce; they are quick and need no drops. Three brands were tested head to head in rabbits.
Pressure in ten rabbit eyes was set by a saline column from about 11 to 86 millimeters of mercury. Each device took readings across that range, giving 298 data sets in all. Average readings from the three agreed closely, and all tracked the true pressure well. Small trends differed: one read slightly low at high pressure, another crossed over near the low twenties. The new iFalcon device proved stable. The authors advise care when switching brands mid-study.