Balanced-armature-based, electromagnetic actuator for round window stimulation of the inner ear with static pre-load

Maren S. Prediger, Eileen Müller, Anatoly Glukhovskoy, Konrad Bethmann, András Bernát Berta, Marc C. Wurz, Hannes Maier · 2025

Read Balanced-armature-based, electromagnetic actuator for round window stimulation of the inner ear with static pre-load by Maren S. Prediger, Eileen Müller, Anatoly Glukhovskoy, Konrad Bethmann, András Bernát Berta, Marc C. Wurz, Hannes Maier (2025) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/pmc-12748139.txt.

Description

Maren Prediger and co-authors built a tiny electromagnetic actuator to drive the inner ear through its round window, in a 2025 paper. Some hearing loss stems from a failing middle ear, and implants can bypass it by shaking the round window membrane directly. Existing drivers are too bulky or too weak at low pitches. Their balanced-armature design, smaller than a grain of rice in width, anchors to bone yet stays compact.

Bench tests swept a wide frequency range, and prototypes then ran in human temporal bones. Output matched devices already in clinics, passing one hundred decibel equivalents at high pitches. Making the parts reliably proved the weak point, with low yields and failures to fix.

Catalog record

CREATORMaren S. Prediger, Eileen Müller, Anatoly Glukhovskoy, Konrad Bethmann, András Bernát Berta, Marc C. Wurz, Hannes Maier
YEAR2025
COLLECTIONPAPERS
FORMATTXT
FILE SIZE40 KB
VERSION1.0
LICENSECC BY 4.0
RIGHTSCC BY
SOURCEEurope PMC

Source: https://europepmc.org/articles/PMC12748139

Browse the full library · About the Tower of Babel