Zero-current chronopotentiometry for wired biosensors
Andrea Nonis, Polyxeni Damala, Eric Bakker · 2025
Read Zero-current chronopotentiometry for wired biosensors by Andrea Nonis, Polyxeni Damala, Eric Bakker (2025) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/pmc-12708822.txt.
Description
Glucose monitors that run day and night usually read current, which drains power and leans on a reference electrode that can drift. Andrea Nonis, Polyxeni Damala, and Eric Bakker present a readout that draws no current at all, in a 2025 paper. Their wired enzyme sensor is read by timing instead of by current flow.
After a short oxidation step, the electrode is left open and its voltage drifts as the enzyme steadily turns over glucose and resets the sensor's redox probe. When the oxidized probe runs out, the voltage trace bends at a clear transition time. The inverse of that time tracks glucose in a straight line across the full range found in the body. Because the exact voltage does not matter, cheap reference electrodes suffice. The method suits long-wear sensors meant to sip power.
Catalog record
| CREATOR | Andrea Nonis, Polyxeni Damala, Eric Bakker |
|---|---|
| YEAR | 2025 |
| COLLECTION | PAPERS |
| FORMAT | TXT |
| FILE SIZE | 26 KB |
| VERSION | 1.0 |
| LICENSE | CC BY 4.0 |
| RIGHTS | CC BY |
| SOURCE | Europe PMC |