Read Recurrent femoral stem fractures in Dorr A femurs: Lessons learned and a call for alternative strategies by Carlos M Lucero, Juan Bautista Luco, Agustin Albani Forneris, Martin A Buttaro (2025) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/pmc-12400381.txt.
Description
Carlos Lucero and three co-authors report a rare case of hip implant stems breaking on both sides in a patient with a particular thigh bone shape. Their report came out in 2025. The shape, called Dorr A, has a narrow canal, which changes how stress concentrates in an implant stem.
The first failure involved a cemented revision stem placed with a cement-within-cement technique after a bone cut along the thigh. A second revision used a different bone window and an uncemented stem fixed lower down, and it succeeded. The same approach later fixed the other side, also with a good result. The authors argue that standard cemented stems can be oversized in narrow canals. They advocate individualized implant choices, including shorter or modular stems, for this anatomy.