Read Accuracy of the Imageless Mode of the ROSA Robotic System for Targeted Resection Thickness in Total Knee Arthroplasty: A Prospective, Single Surgeon Case‐Series Study by Zakareya Gamie, Eustathios Kenanidis, Georgios Douvlis, Nikolaos Milonakis, Alexander Maslaris, Eleftherios Tsiridis (2024) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/pmc-11666924.txt.

Description

Zakareya Gamie and five co-authors wrote a 2024 study on the accuracy of a robotic system for knee replacement surgery. The ROSA robot helps surgeons cut bone to planned thicknesses. This study tested its imageless mode, which works without pre-operative scans, in patients with primary knee osteoarthritis.

One surgeon treated 44 patients between June 2023 and January 2024. Caliper checks compared planned and actual cuts at the thigh bone, shin bone, and back of the kneecap joint. Average gaps ran from about 0.67 to 1.1 millimeters across the six sites. Some gaps reached statistical significance but stayed small in absolute terms. The authors conclude the system hits planned resection thickness with high accuracy. Results apply only to this mode and patient group.

Catalog record

CREATORZakareya Gamie, Eustathios Kenanidis, Georgios Douvlis, Nikolaos Milonakis, Alexander Maslaris, Eleftherios Tsiridis
YEAR2024
COLLECTIONPAPERS
FORMATTXT
FILE SIZE30 KB
VERSION1.0
LICENSECC BY 4.0
RIGHTSCC BY
SOURCEEurope PMC

Source: https://europepmc.org/articles/PMC11666924

Browse the full library · About the Tower of Babel