Performance of a Non-Invasive System for Monitoring Blood Glucose Levels Based on Near-Infrared Spectroscopy Technology (Glucube®)
Fernando Gómez-Peralta, Luis Gabriel Luque Romero, Antonio Puppo-Moreno, Jesús Riesgo, Georg Fischer · 2024
Read Performance of a Non-Invasive System for Monitoring Blood Glucose Levels Based on Near-Infrared Spectroscopy Technology (Glucube®) by Fernando Gómez-Peralta, Luis Gabriel Luque Romero, Antonio Puppo-Moreno, Jesús Riesgo, Georg Fischer (2024) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/pmc-11645037.txt.
Description
Fernando Gomez-Peralta and co-authors test a needle free glucose monitor in 2024 that reads blood sugar through near infrared light. Frequent checks are vital in diabetes, and finger pricks are a daily burden.
People with type 1 diabetes, type 2 diabetes, prediabetes, and normal sugar control used the device for a week beside standard finger prick meters. More than nineteen hundred paired readings were gathered from one hundred five people. The average error sat near twenty percent, and nearly all readings fell in the safe zones of a standard error grid. Skill with the device improved over the week, and error fell in a sub study.
Catalog record
| CREATOR | Fernando Gómez-Peralta, Luis Gabriel Luque Romero, Antonio Puppo-Moreno, Jesús Riesgo, Georg Fischer |
|---|---|
| YEAR | 2024 |
| COLLECTION | PAPERS |
| FORMAT | TXT |
| FILE SIZE | 25 KB |
| VERSION | 1.0 |
| LICENSE | CC BY 4.0 |
| RIGHTS | CC BY |
| SOURCE | Europe PMC |