Comparison of Intraocular Pressure Measurements Using Three Different Methods (Goldmann Applanation Tonometry (GAT), Corvis ST, and iCare) Following Penetrating Keratoplasty

Tatiana Tziola, Argyrios Tzamalis, Spyridon Koronis, Panagiotis Garitsis, Ioannis Tsinopoulos, Nikolaos Ziakas, Brent Siesky · 2024

Read Comparison of Intraocular Pressure Measurements Using Three Different Methods (Goldmann Applanation Tonometry (GAT), Corvis ST, and iCare) Following Penetrating Keratoplasty by Tatiana Tziola, Argyrios Tzamalis, Spyridon Koronis, Panagiotis Garitsis, Ioannis Tsinopoulos, Nikolaos Ziakas, Brent Siesky (2024) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/pmc-11642321.txt.

Description

Tatiana Tziola and co-authors compare three ways to measure pressure inside the eye in a 2024 study of patients who had a cornea transplant. After such surgery, pressure checks guide care, but it is not clear whether the newer devices agree with the long-standing standard test.

The team measured twenty-five eyes from twenty-five patients with the standard Goldmann test, a handheld iCare device, and a Corvis ST device that also gives a corrected value. The corrected Corvis value agreed most closely with the standard. The iCare readings ran a bit high and varied more when many stitches remained in the eye. A thicker cornea linked to lower corrected readings, while other traits showed no clear effect.

Catalog record

CREATORTatiana Tziola, Argyrios Tzamalis, Spyridon Koronis, Panagiotis Garitsis, Ioannis Tsinopoulos, Nikolaos Ziakas, Brent Siesky
YEAR2024
COLLECTIONPAPERS
FORMATTXT
FILE SIZE30 KB
VERSION1.0
LICENSECC BY 4.0
RIGHTSCC BY
SOURCEEurope PMC

Source: https://europepmc.org/articles/PMC11642321

Browse the full library · About the Tower of Babel