A Comprehensive Clinical Assessment of the LumiraDx International Normalized Ratio (INR) Assay for Point-of-Care Monitoring in Anticoagulation Therapy

Riffat Munir, Elise Schapkaitz, Lara Noble, Sakina Loonat, Melanie McCree, Nazeer Ali, Barry Jacobson, Wendy Susan Stevens, Lesley Erica Scott, Gerald J Kost · 2024

Read A Comprehensive Clinical Assessment of the LumiraDx International Normalized Ratio (INR) Assay for Point-of-Care Monitoring in Anticoagulation Therapy by Riffat Munir, Elise Schapkaitz, Lara Noble, Sakina Loonat, Melanie McCree, Nazeer Ali, Barry Jacobson, Wendy Susan Stevens, Lesley Erica Scott, Gerald J Kost (2024) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/pmc-11639838.txt.

Description

Riffat Munir and co-authors test a bedside device that measures blood thinning levels for patients on anticoagulant drugs, at two Johannesburg hospitals in a 2024 study. The standard check needs a central lab, which can delay dose choices, especially far from big centers. A quick finger-prick test at the point of care could close that gap.

The device matched lab results well overall, with tight precision on control samples and a small average bias. Readings drifted upward at high values, though, so the authors advise confirming very high bedside results in the lab. They judge the test fit for faster, patient-centered monitoring in clinics like theirs.

Catalog record

CREATORRiffat Munir, Elise Schapkaitz, Lara Noble, Sakina Loonat, Melanie McCree, Nazeer Ali, Barry Jacobson, Wendy Susan Stevens, Lesley Erica Scott, Gerald J Kost
YEAR2024
COLLECTIONPAPERS
FORMATTXT
FILE SIZE33 KB
VERSION1.0
LICENSECC BY 4.0
RIGHTSCC BY
SOURCEEurope PMC

Source: https://europepmc.org/articles/PMC11639838