Exchange rates of second generation Microcuff® pediatric endotracheal tubes in children weighing more than 3 kg

Ilka Schmidt-Deubig, Michael Kemper, Pedro D Wendel-Garcia, Markus Weiss, Jörg Thomas, Christian Peter Both, Achim Schmitz · 2024

Read Exchange rates of second generation Microcuff® pediatric endotracheal tubes in children weighing more than 3 kg by Ilka Schmidt-Deubig, Michael Kemper, Pedro D Wendel-Garcia, Markus Weiss, Jörg Thomas, Christian Peter Both, Achim Schmitz (2024) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/pmc-11615024.txt.

Description

Ilka Schmidt-Deubig and co-authors audit breathing tube swaps in children in a 2024 paper. Cuffed tubes seal better and should need fewer exchanges than uncuffed ones. A second-generation pediatric tube with an adapted size chart was their test case.

Zurich records covered 14,188 intubations in children over three kilograms from 2009 to 2015. Most oral placements followed the maker's size chart. When the chart size was used, the need to swap the tube was a fraction of a percent. Off-chart choices carried a far higher exchange risk, above twenty percent in the reported figure. Correct sizing, the audit argues, is where the cuffed tube's advantage is won or lost.

Catalog record

CREATORIlka Schmidt-Deubig, Michael Kemper, Pedro D Wendel-Garcia, Markus Weiss, Jörg Thomas, Christian Peter Both, Achim Schmitz
YEAR2024
COLLECTIONPAPERS
FORMATTXT
FILE SIZE26 KB
VERSION1.0
LICENSECC BY 4.0
RIGHTSCC BY
SOURCEEurope PMC

Source: https://europepmc.org/articles/PMC11615024

Browse the full library · About the Tower of Babel