Ala®sil chemical characterization and toxicity evaluation: an example of the need for the Medical Device Regulation 2017/745

Cristina Andrés-Iglesias, Ivan Fernandez-Bueno, Salvador Pastor-Idoate, Rosa M. Coco-Martin, J. Carlos Pastor · 2024

Read Ala®sil chemical characterization and toxicity evaluation: an example of the need for the Medical Device Regulation 2017/745 by Cristina Andrés-Iglesias, Ivan Fernandez-Bueno, Salvador Pastor-Idoate, Rosa M. Coco-Martin, J. Carlos Pastor (2024) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/pmc-10811781.txt.

Description

Cristina Andrés-Iglesias and co-authors examine in 2024 a silicone eye infusion fluid that was sold in Europe and withdrawn in 2016. The fluid, Ala®sil, was used during eye surgery under older, looser device rules. The team asks whether it could pass today's stricter regulation.

Lab analysis found the fluid is a low-weight polymer mixed with small impurity molecules. In cell tests it sped up retinal cell death, though tissue form stayed normal. The only clinical report, five patients, rated it unhelpful in most cases. The authors argue the case shows why the newer rules demand solid safety proof before sale.

Catalog record

CREATORCristina Andrés-Iglesias, Ivan Fernandez-Bueno, Salvador Pastor-Idoate, Rosa M. Coco-Martin, J. Carlos Pastor
YEAR2024
COLLECTIONPAPERS
FORMATTXT
FILE SIZE37 KB
VERSION1.0
LICENSECC BY 4.0
RIGHTSCC BY
SOURCEEurope PMC