SAFIR-I: first NEMA NU 4-2008-based performance characterization
Pascal Bebié, Werner Lustermann, Jan Debus, Christian Ritzer, Günther Dissertori, Bruno Weber · 2023
Read SAFIR-I: first NEMA NU 4-2008-based performance characterization by Pascal Bebié, Werner Lustermann, Jan Debus, Christian Ritzer, Günther Dissertori, Bruno Weber (2023) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/pmc-10716090.txt.
Description
Pascal Bebié and co-authors benchmark a new PET scanner insert for small animals, in work from 2023. The device, SAFIR-I, slides into an MRI scanner so both imaging types can run at once in mice and rats, even at high tracer doses.
Following the standard NEMA protocol for such scanners, the team measured spatial resolution, count-rate performance, sensitivity, and image quality in mouse- and rat-sized phantoms. Count-rate peaks were not even reached at the highest doses tested, and calibration of image values proved accurate. The authors conclude that the design goals were met, especially for fast tracer studies at high activity, though room for improvement remains.
Catalog record
| CREATOR | Pascal Bebié, Werner Lustermann, Jan Debus, Christian Ritzer, Günther Dissertori, Bruno Weber |
|---|---|
| YEAR | 2023 |
| COLLECTION | PAPERS |
| FORMAT | TXT |
| FILE SIZE | 112 KB |
| VERSION | 1.0 |
| LICENSE | CC BY 4.0 |
| RIGHTS | CC BY |
| SOURCE | Europe PMC |