Comparison of the TEMPO binocular perimeter and Humphrey field analyzer

Takashi Nishida, Robert N. Weinreb, Juan Arias, Cristiana Vasile, Sasan Moghimi · 2023

Read Comparison of the TEMPO binocular perimeter and Humphrey field analyzer by Takashi Nishida, Robert N. Weinreb, Juan Arias, Cristiana Vasile, Sasan Moghimi (2023) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/pmc-10692178.txt.

Description

Takashi Nishida and co-authors compare, in a 2023 study, a new two-eyed visual field tester against the standard machine used to track glaucoma. The Humphrey analyzer tests one eye at a time and anchors clinic practice. The TEMPO device tests both eyes together with a quicker estimation routine. Seven hundred forty eyes, spanning healthy, suspect, and glaucomatous, took both tests on the same day in random order.

Headline scores agreed closely between the machines. The new tester's readings tied slightly more strongly to nerve fiber layer thickness measured by tomography, a sign of validity. It also finished much faster, taking a bit over four minutes against more than seven. The authors say the binocular approach merits study for long-term monitoring, where visit burden shapes whether patients keep coming.

Catalog record

CREATORTakashi Nishida, Robert N. Weinreb, Juan Arias, Cristiana Vasile, Sasan Moghimi
YEAR2023
COLLECTIONPAPERS
FORMATTXT
FILE SIZE27 KB
VERSION1.0
LICENSECC BY 4.0
RIGHTSCC BY
SOURCEEurope PMC

Source: https://europepmc.org/articles/PMC10692178