Factory-Calibrated Continuous Glucose Monitoring Systems in Type 1 Diabetes: Accuracy during In-Clinic Exercise and Home Use

Sissel Banner Lundemose, Christian Laugesen, Ajenthen Gayathri Ranjan, Kirsten Nørgaard · 2023

Read Factory-Calibrated Continuous Glucose Monitoring Systems in Type 1 Diabetes: Accuracy during In-Clinic Exercise and Home Use by Sissel Banner Lundemose, Christian Laugesen, Ajenthen Gayathri Ranjan, Kirsten Nørgaard (2023) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/pmc-10675113.txt.

Description

Sissel Banner Lundemose and co-authors check whether exercise throws off the accuracy of factory-set glucose sensors in type 1 diabetes. Thirteen adults wore all three leading European systems at once. Each took part in a one-hour clinic exercise session, then three days of home use, with blood tests as the referee.

During exercise, the three systems' average error did not differ clearly. One system, Freestyle Libre 2, lagged at tracking the direction of glucose change during exercise. At home it also showed the largest error of the three. The other two systems, Dexcom G6 and Guardian 4, held up better overall in this small head-to-head.

Catalog record

CREATORSissel Banner Lundemose, Christian Laugesen, Ajenthen Gayathri Ranjan, Kirsten Nørgaard
YEAR2023
COLLECTIONPAPERS
FORMATTXT
FILE SIZE19 KB
VERSION1.0
LICENSECC BY 4.0
RIGHTSCC BY
SOURCEEurope PMC

Source: https://europepmc.org/articles/PMC10675113

Browse the full library · About the Tower of Babel