Agreement of keratometry measurements for toric IOL planning in routine clinical practice: KR-800, IOLMaster 500, and Pentacam HR

Yeoeun Lim, Moonwon Hwang · 2026

Read Agreement of keratometry measurements for toric IOL planning in routine clinical practice: KR-800, IOLMaster 500, and Pentacam HR by Yeoeun Lim, Moonwon Hwang (2026) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/plos-pone-0356995.txt.

Description

Yeoeun Lim and Moonwon Hwang compared three machines that measure the cornea's curve in everyday clinic conditions. They used 227 eyes of cataract surgery candidates. Their 2026 study matters because toric lens implants are chosen from exactly these measurements. Toric lenses correct astigmatism during cataract surgery. Prior device comparisons were run under tidy research conditions.

On average the devices agreed well. For individual patients they often did not. Predicted leftover astigmatism from device disagreements passed the meaningful line in nearly two thirds of patients. Among genuine toric candidates, up to a third had axis readings differing by an amount known to spoil correction. One device also nominated fewer candidates. The authors advise confirming measurements before committing a lens.

Catalog record

CREATORYeoeun Lim, Moonwon Hwang
YEAR2026
COLLECTIONPAPERS
FORMATTXT
FILE SIZE44764
LICENSECC_BY
RIGHTSCC BY
SOURCEPLOS

Source: https://doi.org/10.1371/journal.pone.0356995

Browse the full library · About the Tower of Babel