Using nitric oxide (NO) for gas sterilization at room temperature: A novel sterilization modality
Weilue He, Julie Osborne, Megan C Frost · 2026
Read Using nitric oxide (NO) for gas sterilization at room temperature: A novel sterilization modality by Weilue He, Julie Osborne, Megan C Frost (2026) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/plos-pone-0355735.txt.
Description
Weilue He, Julie Osborne, and Megan Frost test a new way to sterilize medical equipment in their 2026 paper. Hospitals and device makers now rely heavily on ethylene oxide gas, which works well but is highly toxic to people and the environment. Their candidate replacement is nitric oxide gas, used at ordinary room temperature and normal air pressure.
The team exposed hardy bacterial spores, the standard test for sterilization, to nitric oxide and tracked how fast they died off. The gas killed the test spores in a steady, predictable pattern and also inactivated two bacteria that commonly cause infections. The dose that workers or users might encounter stayed well below the level classed as immediately dangerous, pointing to a safer, greener option.
Catalog record
| CREATOR | Weilue He, Julie Osborne, Megan C Frost |
|---|---|
| YEAR | 2026 |
| COLLECTION | PAPERS |
| FORMAT | TXT |
| FILE SIZE | 34276 |
| LICENSE | CC_BY |
| RIGHTS | CC BY |
| SOURCE | PLOS |