Read ISCEV guide to visual electrodiagnostic procedures by Anthony G Robson, Josefin Nilsson, Shiying Li, Subhadra Jalali, Anne B Fulton, Alma Patrizia Tormene, Graham E Holder, Scott E Brodie in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/europmc-pmc5811581.txt.
Introduction
Robson and co-authors guide readers through visual electrodiagnostic tests. These painless tests record the eye's electric replies to light. Each test probes a different cell type or layer. Results help doctors pin down where vision fails.
The guide comes from ISCEV, the field's world society. It covers the full-field and pattern electroretinogram. It also covers the multifocal test, the electrooculogram, and evoked potentials. Basic principles come first, then common symptoms and case examples. Suggested work-ups show which test fits which complaint. Readers get a primer on the standard tests of eye function.
Catalog record
| CREATOR | Anthony G Robson, Josefin Nilsson, Shiying Li, Subhadra Jalali, Anne B Fulton, Alma Patrizia Tormene, Graham E Holder, Scott E Brodie |
|---|---|
| COLLECTION | PAPERS |
| FORMAT | TXT |
| FILE SIZE | 64 KB |
| LICENSE | CC BY 4.0 |
| RIGHTS | CC BY |
| SOURCE | Europe PMC |
Browse the full library · About the Tower of Babel
Full text
The opening of the work, from the library's hosted text:
ISCEV guide to visual electrodiagnostic procedures By Anthony G Robson, Josefin Nilsson, Shiying Li, Subhadra Jalali, Anne B Fulton, Alma Patrizia Tormene, Graham E Holder, Scott E Brodie Published in: Documenta Ophthalmologica. Advances in Ophthalmology DOI: 10.1007/s10633-017-9621-y ABSTRACT Abstract Clinical electrophysiological testing of the visual system incorporates a range of noninvasive tests and provides an objective indication of function relating to different locations and cell types within the visual system. This document developed by the International Society for Clinical Electrophysiology of Vision provides an introduction to standard visual electrodiagnostic procedures in widespread use including the full-field electroretinogram (ERG), the pattern electroretinogram (pattern ERG or PERG), the multifocal electroretinogram (multifocal ERG or mfERG), the electrooculogram (EOG) and the cortical-derived visual evoked potential (VEP). The guideline outlines the basic principles of testing. Common clinical presentations and symptoms are described with illustrative examples and suggested investigation strategies. Keywords: ISCEV standards, Clinical electrophysiology, Electrooculogram (EOG), Electroretinogram (ERG), Pattern ERG, Multifocal ERG (mfERG), Visual evoked potential (VEP), Optic neuropathy, Maculopathy, Retinopathy INTRODUCTION Clinical electrophysiological testing of the visual system incorporates a range of tests based upon the recording of electrical potentials evoked by visual stimuli, using electrodes situated on the surface of the eyes, the peri-orbital skin or scalp. The tests are noninvasive and provide an objective indication of function relating to different locations and cell types within the visual system. This document developed by the International Society for Clinical Electrophysiology of Vision (ISCEV) provides an introduction to standard visual electrodiagnostic procedures in widespread use and describes the common clinical indications for which these tests are applicable. Detailed specifications for each procedure may be found in the appropriate ISCEV standards [1–5]. The basic principles of electrodiagnostic testing are outlined in this document, but the document is not intended to be prescriptive or to address every clinical scenario and is not a mandate for specific procedures on individual patients. Clinical electrophysiological testing has the greatest utility when performed in conjunction with clinical assessment by specialist eye care professionals. Clinical context is essential to enable appropriate clinical management. This guideline describes the basic methods and underlying principles of testing for each of the standard tests including the full-field flash electroretinogram (ERG), the pattern electroretinogram (pattern ERG or PERG), the multifocal electroretinogram (mfERG), the electrooculogram (EOG) and the cortical-derived visual evoked potential (VEP). The principal focus is to place these tests in clinical context. Common clinical presentations and symptoms are described with illustrative examples and suggested investigation strategies. THE ELECTROPHYSIOLOGICAL TESTS ISCEV publishes and regularly updates standards for clinical tests of the visual system. The most recent publications are listed on the ISCEV Web site www.iscev.org/standards and are freely accessible. In addition to these basic tests, extended protocols may support differential diagnosis or functional monitoring. Below is a brief description of normal waveforms resulting from the ISCEV standard tests and the physiologic implications of abnormal responses. Users should consult the relevant standard or extended protocol for detailed testing protocols. The full-field ERG The ISCEV standard full-field ERGs (Fig. 1a) are global responses of the retina to brief flashes of light and provide an assessment of generalized retinal function under light- and dark-adapted conditions. A ganzfeld (German for “whole field”) stimulator, which provides a uniformly illuminated field, is used to deliver a range of flash stimuli that evenly illuminate the maximal area of retina. The ERGs are recorded with electrodes in contact with the cornea or conjunctiva or with skin electrodes attached to the lower eyelids. Several types of corneal electrode may be used including contact lens, fiber, jet and gold foil electrodes. The pupils are dilated to maximize retinal illumination and to minimize inter-subject and inter-visit variability. Reliable interpretation of recordings requires comparison with electrode-specific and age-matched normative data. The normal test–retest variability of ERG parameters is also an important consideration if used to monitor disease progression or the safety or efficacy of treatments. [Figure: Representative full-field and pattern ERGs in a normal subject (a), in a case of macular dystrophy (b), cone-rod dystrophy (c), rod-cone dystrophy with relative sparing of macular function (d), complete CSNB (e), incomplete CSNB (f) and birdshot retinochoroidopathy (BRC) before treatment (g) and after treatment illustrating full recovery of the ERG and PERG (h). Recordings showed a high degree of inter-ocular symmetry except in BRC (data from other eye are not shown). Note there is a 20-ms pre-stimulus delay in all single flash ERG recordings. Two responses for each stimulus condition are superimposed to illustrate reproducibility. Broken lines replace blink artefacts occurring after the ERGs, for clarity] The ISCEV standard protocol includes dark-adapted (DA) recordings after 20-min dark adaptation to flash strengths of 0.01, 3.0 and 10.0 cd s m−2 (DA 0.01; DA 3.0; DA 10.0). The weak flash (DA 0.01) ERG arises in the inner retinal rod bipolar cells and is the only standard test that selectively monitors rod system function. Abnormality of the DA 0.01 ERG can be caused by either rod photoreceptor dysfunction or selective dysfunction occurring post-phototransduction or at the level of the inner retinal rod bipolar cells. The DA 3.0 (standard flash) and DA 10.0 (strong flash) ERGs have input from both rod and cone systems, but the DA rod system contribution dominates in a normal retina. Approximately the first 8 ms of the cornea-negative a-wave reflects rod hyperpolarizations, and as the a-wave in the DA 10.0 ERG is of shorter peak time and larger than in the DA 3.0 ERG, it provides a better measure of rod photoreceptor function. The subsequent cornea-positive b-wave arises largely in the rod On-bipolar cells and reflects function that is post-phototransduction. Thus, the DA strong flash ERG enables localization of dysfunction to the rod photoreceptors (a-wave reduction and concomitant b-wave reduction) or to a level that is post-phototransduction or inner retinal (sparing of the a-wave with b-wave reduction). The DA oscillatory potentials (OPs) are small high-frequency components normally visible on the rising limb of the DA 3.0 and DA 10.0 ERG b-waves and are thought to reflect amacrine cell signaling. Reduction in the OPs is often associated with other ERG abnormalities but may occur selectively in some disorders. The cone system contribution to both DA ERG a- and b-waves is minor in a normal retina but can be of greater significance in patients with disease primarily or exclusively affecting the rod system. Standard light-adapted (LA) ERGs provide two measures of generalized cone system function; both are obtained to a flash strength of 3.0 cd s m−2, after a standard period of 10-min light adaptation in the Ganzfeld with a constant background luminance of 30 cd m−2. A 30 Hz flash stimulus, superimposed on the background, is used to elicit the LA 30 Hz flicker ERG, generated largely by post-receptoral retinal structures. The single flash cone (LA 3.0) ERG consists mainly of a- and b-waves. The LA 3.0 ERG a-wave arises in the cone photoreceptors and Off-bipolar cells; the b-wave is dominated by a combination of cone On- and Off-bipolar cell activity, and a reduced b/a ratio suggests cone system dysfunction that is post-phototransduction or post-receptoral. The full-field ERG enables the distinction between generalized outer and inner retinal dysfunction and predominant rod or cone system dysfunction. Symptoms and/or clinical signs may suggest a