Charge Collection Efficiency in Air-Vented Plane-Parallel Ionisation Chambers at Ultra-High Dose Rates

A Self-Consistent Garfield++ Monte Carlo Model Including Space-Charge Effects and Ion Recombination · 2026

Read Charge Collection Efficiency in Air-Vented Plane-Parallel Ionisation Chambers at Ultra-High Dose Rates by A Self-Consistent Garfield++ Monte Carlo Model Including Space-Charge Effects and Ion Recombination (2026) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/arxiv-2606-24384.txt.

Introduction

Ortega and colleagues model whether radiation dose meters stay truthful at the extreme dose rates of FLASH radiotherapy. FLASH delivery, tens of grays per second, spares normal tissue while controlling tumours. The ionisation chambers that verify dose were designed for gentler beams, though. In air-vented plane-parallel chambers, ions recombine before collection, and the charge cloud itself distorts the field. Both grow vicious at ultra-high dose rates.

A self-consistent Monte Carlo model, built with the Garfield++ toolkit, simulates charge collection efficiency. Space-charge effects and ion recombination are included together, rather than corrected for each in isolation. Self-consistency matters. The drifting charges reshape the very field that moves them. The 2026 model, developed at the Université libre de Bruxelles, gives FLASH researchers a physics-based way to predict chamber response. There, the classical recombination corrections, derived for conventional dose rates, can no longer be trusted.

Catalog record

CREATORA Self-Consistent Garfield++ Monte Carlo Model Including Space-Charge Effects and Ion Recombination
YEAR2026
COLLECTIONPAPERS
FORMATTXT
FILE SIZE50 KB
LICENSEhttp://creativecommons.org/licenses/by/4.0/
RIGHTSCC BY
SOURCEarXiv

Source: https://arxiv.org/abs/2606.24384

Browse the full library · About the Tower of Babel

Full text

The opening of the work, from the library's hosted text:

Charge Collection Efficiency in Air-Vented Plane-Parallel Ionisation Chambers at Ultra-High Dose Rates A Self-Consistent Garfield++ Monte Carlo Model Including Space-Charge Effects and Ion Recombination Abstract 1 1 footnotetext: Corresponding author: [email protected] 1 Introduction Since the discovery of what is now referred to as the FLASH effect by [ 12 ] in [ 12 ] , research on ultra-high dose rate (UHDR) radiation ( \geq 40\text{\,}\mathrm{Gy}\text{\,}{\mathrm{s}}^{-1} ) has grown rapidly. The reduced toxicity observed in normal tissues during UHDR irradiation, combined with equivalent or even enhanced tumour control, offers promising perspectives for radiotherapy. Ionisation chambers (ICs) are the reference dosimeters recommended by the International Atomic Energy Agency (IAEA) [ 18 ] for the determination of absorbed dose to water. In particular, plane-parallel ionisation chambers (PPICs) are widely used in radiotherapy systems for beam monitoring during patient treatment. To accurately determine the absorbed dose, several correction factors must be applied during IC calibration. One of these factors, k_{s} , accounts for incomplete charge collection caused by ion recombination. Under conventional radiotherapy conditions, ion recombination corrections are commonly performed using the model originally developed by [ 6 ] in [ 6 ] and later extended by [ 5 ] in [ 5 ] to account for the fraction of electrons escaping attachment to electronegative molecules, referred to as the free electron fraction (FEF). However, it is now well established [ 21 , 13 , 24 , 15 ] that these analytical models are no longer valid under the UHDR conditions required for FLASH radiotherapy. In particular, electric-field distortions induced by space charge can no longer be neglected. Several studies [ 20 , 1 , 7 ] have shown that these distortions can locally exceed the nominal applied electric field by several times. To better understand recombination effects in PPICs under UHDR pulsed beams, we extended the Garfield++ [ 26 ] framework and developed a self-consistent Monte Carlo simulation coupling particle transport, electron attachment, ion–ion recombination, and space-charge effects. The flexibility of Garfield++ enables simulations for arbitrary gas mixtures, atmospheric conditions, IC geometries, incident particle types, and beam spatial and temporal structures, allowing comprehensive investigations of recombination phenomena under a wide range of irradiation conditions. 2 Physical background and literature review This section reviews the physical mechanisms governing charge transport and recombination in air-filled ionisation chambers. Particular emphasis is placed on electron attachment, ion transport, and ion–ion recombination processes. 2.1 Electron swarm parameters and attachment processes The dominant mechanism responsible for the formation of negative ions in air is the attachment of electrons to oxygen molecules. Unlike noble gases, often used in gaseous detectors to avoid such phenomena, oxygen requires only a relatively low collision energy to capture free electrons. Two different attachment mechanisms exist: two-body and three-body attachment. Two-body attachment occurs through a dissociative process involving the oxygen molecule and requires a minimum dissociation energy threshold of 5.61\text{\,}\mathrm{eV} [ 11 ] : {}\mathrm{e}{\vphantom{\mathrm{X}}}^{\mathrm{-}}{}+{}{}\mathrm{O}{\vphantom{\mathrm{X}}}_{\smash[t]{\mathrm{2}}}{}\mathrel{\mathop{\makebox[0.0pt]{\rightarrow\displaystyle}}\limits}{}{}\mathrm{O}{\vphantom{\mathrm{X}}}^{\mathrm{-}}{}+{}{}\mathrm{O}. (1) Water molecules present in humid air can also undergo dissociation and form negative ions following electron attachment, although their dissociation energy is higher, around 6.32\text{\,}\mathrm{eV} : {}\mathrm{e}{\vphantom{\mathrm{X}}}^{\mathrm{-}}{}+{}{}{}\mathrm{H}{\vphantom{\mathrm{X}}}_{\smash[t]{\mathrm{2}}}\mathrm{O}{}\mathrel{\mathop{\makebox[0.0pt]{\rightarrow\displaystyle}}\limits}{}{}{}{}\mathrm{OH}{\vphantom{\mathrm{X}}}^{\mathrm{-}}\,\mathrm{/}\,\mathrm{O}{\vphantom{\mathrm{X}}}^{\mathrm{-}}\,\mathrm{/}\,\mathrm{H}{\vphantom{\mathrm{X}}}^{\mathrm{-}}. (2) At low electron energies, three-body attachment becomes dominant and proceeds through a two-step process. In this mechanism, a third body, generally \text{O}{\vphantom{\text{X}}}_{\smash[t]{\text{2}}} or \text{H}{\vphantom{\text{X}}}_{\smash[t]{\text{2}}}\text{O} , is required to de-excite the metastable oxygen anion [ 11 , 4 , 28 ] : UsingMagboltz [ 2 ] , wesimulatedelectrontransportinairunderareducedelectricfieldof 20\text{\,}\mathrm{Td} ( 1\text{\,}\mathrm{Td} = {10}^{-21}\text{\,}\mathrm{V}\text{\,}{\mathrm{m}}^{2} ).Thesimulationyieldedameanelectroncollisionenergyof$1.14\text{\,}\mathrm{eV}$(\lx@cref{creftype\lx@tilde refnum}{fig:electron_energies}),indicatingthatonlythethree-bodyattachmentprocessisenergeticallyaccessibleundertheseconditions.Thisresultisconsistentwiththefindingsreportedby\cite[cite]{[\@@bibref{}{deurquijo_2024}{}{}]}.\par\begin{figure}\centering\includegraphics[width]{electron_energy.pdf} \@@toccaption{{\lx@tag[ ]{{1}}{Electron collision energy at $20\text{\,}\mathrm{Td}$ in humid air with $2\text{\,}\mathrm{\%}$ {}{}$\text{H}{\vphantom{\text{X}}}_{\smash[t]{\text{2}}}\text{O}$.}}}\@@caption{{\lx@tag[: ]{{\small Figure 1}}{\small Electron collision energy at $20\text{\,}\mathrm{Td}$ in humid air with $2\text{\,}\mathrm{\%}$ {}{}$\text{H}{\vphantom{\text{X}}}_{\smash[t]{\text{2}}}\text{O}$.}}} \@add@centering\end{figure}\par TovalidatethetransportparametersobtainedwithMagboltz,wesimulatedelectrontransportindryairandhumidaircontaining$2\text{\,}\mathrm{\%}${}{}\mathrm{H}{\vphantom{\mathrm{X}}}_{\smash[t]{\mathrm{2}}}\mathrm{O}.Simulationswereperformedat$293.15\text{\,}\mathrm{K}$and$760\text{\,}\mathrm{Torr}$,usingagascompositionof$76.54\text{\,}\mathrm{\%}${}\mathrm{N}{\vphantom{\mathrm{X}}}_{\smash[t]{\mathrm{2}}},$20.54\text{\,}\mathrm{\%}${}\mathrm{O}{\vphantom{\mathrm{X}}}_{\smash[t]{\mathrm{2}}},$0.92\text{\,}\mathrm{\%}${}\mathrm{Ar}and$2\text{\,}\mathrm{\%}${}{}\mathrm{H}{\vphantom{\mathrm{X}}}_{\smash[t]{\mathrm{2}}}\mathrm{O}forhumidair.Theresultingelectronmobility μ _e andelectronattachmentcoefficient η werecomparedwithexperimentalmeasurementsreportedintheliterature\penalty\ (\lx@cref{creftype\lx@tilde refnum}{fig:attachment_and_mobility}).Thecomparisonincludesmeasurementsfrom\cite[cite]{[\@@bibref{}{deurquijo_2024}{}{}]}and\cite[cite]{[\@@bibref{}{davies_1985}{}{}]}fordryairandhumidaircontaining$2\text{\,}\mathrm{\%}${}{}\mathrm{H}{\vphantom{\mathrm{X}}}_{\smash[t]{\mathrm{2}}}\mathrm{O},aswellasmeasurementsfrom\cite[cite]{[\@@bibref{}{hochhaeuser_1994}{}{}]}forhumidaircontaining$1.2\text{\,}\mathrm{\%}${}{}\mathrm{H}{\vphantom{\mathrm{X}}}_{\smash[t]{\mathrm{2}}}\mathrm{O}.\par\begin{figure*}\centering\begin{subfigure}[165.59853pt]\includegraphics[width]{mobility_dry-air.pdf} \@@toccaption{{\lx@tag[ ]{{(a)}}{Electron mobility in dry air.}}}\@@caption{{\lx@tag[ ]{{\small(a)}}{\small Electron mobility in dry air.}}}\end{subfigure}\hfill\centering\begin{subfigure}[165.59853pt]\includegraphics[width]{attachment_dry-air.pdf} \@@toccaption{{\lx@tag[ ]{{(b)}}{Electron attachment coefficient in dry air.}}}\@@caption{{\lx@tag[ ]{{\small(b)}}{\small Electron attachment coefficient in dry air.}}}\end{subfigure}\begin{subfigure}[165.59853pt]\includegraphics[width]{mobility_humid-air.pdf} \@@toccaption{{\lx@tag[ ]{{(c)}}{Electron mobility in humid air ($1.2\text{\,}\mathrm{\%}$ {}{}$\text{H}{\vphantom{\text{X}}}_{\smash[t]{\text{2}}}\text{O}$ and $2\text{\,}\mathrm{\%}$ {}{}$\text{H}{\vphantom{\text{X}}}_{\smash[t]{\text{2}}}\text{O}$).}}}\@@caption{{\lx@tag[ ]{{\small(c)}}{\small Electron mobility in humid air ($1.2\text{\,}\mathrm{\%}$ {}{}$\text{H}{\vphantom{\text{X}}}_{\smash[t]{\text{2}}}\text{O}$ and $2\text{\,}\mathrm{\%}$ {}{}$\text{H}{\vphantom{\text{X}}}_{\smash[t]{\text{2}}}\text{O}$).}}}\end{subfigure}\begin{subfigure}[165.59853pt]\includegraphics[width]{attachment_humid-air.pdf} \@@toccaption{{\lx@tag[ ]{{(d)}}{Electron attachment coefficient in humid air ($1.2\text{\,}\mathrm{\%}$ {}{}$\text{H}{\vphantom{\text{X}}}_{\smash[t]{\text{2}}}\text{O}$ and $2\text{\,}\mathrm{\%}$ {}{}$\text{H}{\vphantom{\text{X}}}_{\smash[t]{\text{2}}}\text{O}$).}}}\@@caption{{\lx@tag[ ]{{\small(d)}}{\small Electron attachment coefficient in humid air ($1.2\text{\,}\mathrm{\%}$ {}{}$\text{H}{\vphantom{\text{X}}}_{\smash[t]{\text{2}}}\text{O}$ and $2\text{\,}\mathrm{\%}$ {}{}$\text{H}{\vphantom{\text{X}}}_{\smash[t]{\text{2}}}\text{O}$).}}}\end{subfigure}\@@toccaption{{\lx@tag[ ]{{2}}{Comparison of the electron attachment coefficient and electron mobility in dry and humid air obtained from Magboltz simulations with experimental measurements reported in the literature. Experimental data correspond to dry air and humid air with $2\text{\,}\mathrm{\%}$ {}{}$\text{H}{\vphantom{\text{X}}}_{\smash[t]{\text{2}}}\text{O}$\penalty\ \cite[cite]{[\@@bibref{}{deurquijo_2024, davies_1985}{}{}]}, while data from \cite[cite]{[\@@bibref{}{hochhaeuser_1994}{}{}]} correspond to humid air with $1.2\text{\,}\mathrm{\%}$ {}{}$\text{H}{\vphantom{\text{X}}}_{\smash[t]{\text{2}}}\text{O}$.}}}\@@caption{{\lx@tag[: ]{{\small Figure 2}}{\small Comparison of the electron attachment coefficient and electron mobility in dry and humid air obtained from Magboltz simulations with experimental measurements reported in the literature. Experimental data correspond to dry air and humid air with $2\text{\,}\mathrm{\%}$ {}{}$\text{H}{\vphantom{\text{X}}}_{\smash[t]{\text{2}}}\text{O}$\penalty\ \cite[cite]{[\@@bibref{}{deurquijo_2024, davies_1985}{}{}]}, while data from \cite[cite]{[\@@bibref{}{hochhaeuser_1994}{}{}]} correspond to humid air with $1.2\text{\,}\mathrm{\%}$ {}{}$\text{H}{\vphantom{\text{X}}}_{\smash[t]{\text{2}}}\text{O}$.}}} \@add@centering\@add@centering\end{figure*}\par Smalldiscrepanciesareobservedintheattachmentcoefficient,mainlyatlowreducedelectricfields.However,thesedifferencesoccurwelloutsidetheoperatingrangeconsideredinthiswork.Typicalionisationchambersoperateatelectricfieldsaround$3\text{\,}\mathrm{kV}\text{\,}{\mathrm{cm}}^{-1}$,correspondingtoareducedelectricfieldofapproximately$12\text{\,}\mathrm{Td}$inair.\par 2.2 Ion swarm parameters and recombination Following the ionisation of the incident particle, the dominant positive ions formed in air are \text{N}{\vphantom{\text{X}}}_{\smash[t]{\text{2}}}{\vphantom{\text{X}}}^{\text{+}} and \text{O}{\vphantom{\text{X}}}_{\smash[t]{\text{2}}}{\vphantom{\text{X}}}^{\text{+}} . As discussed in Section 2.1 , free electrons rapidly attach to electronegative molecules in air, forming mainly \text{O}{\vphantom{\text{X}}}_{\smash[t]{\text{2}}}{\vphantom{\text{X}}}^{\text{\hskip 0.90417pt--\hskip 0.90417pt}} anions. In air, the attachment time depends on pressure, temperature, humidity, and electric field strength. In PPICs, the characteristic attachment time is expressed as \tau=\frac{1}{\eta\,v_{e}}, (3d) where v_{e} denotes the electron drift velocity. Under typical operating conditions, \tau ranges from approximately 1\text{\,}\mathrm{ns} to 100\text{\,}\mathrm{ns} . Very few experimental measurements of ion mobility in air are available in the literature. Ion mobility is known to depend not only on the electric field [ 10 , 31 , 36 ] , but also on atmospheric conditions such as pressure, temperature, and humidity [ 34 , 35 ] . During the drift phase, hydration and clustering processes involving water molecules can lead to the formation of heavier ions [ 14 ] . These species reduce the average mobility of both cations and anions. [ 35 ] reported ion mobility values ranging from 1.5\text{\times}{10}^{-4}3.3\text{\times}{10}^{-4}\text{\,}{\mathrm{m}}^{2}\text{\,}{\mathrm{V}}^{-1}\text{\,}{\mathrm{s}}^{-1} for anions and from 1.4\text{\times}{10}^{-4}2.0\text{\times}{10}^{-4}\text{\,}{\mathrm{m}}^{2}\text{\,}{\mathrm{V}}^{-1}\text{\,}{\mathrm{s}}^{-1} for cations. A similarly large spread exists for the ion–ion recombination coefficient \alpha , with reported values mainly ranging between 0.9\text{\times}{10}^{-12}1.6\text{\times}{10}^{-12}\text{\,}{\mathrm{m}}^{3}\text{\,}{\mathrm{s}}^{-1} [ 6 , 22 , 16 ] . Furthermore, [ 14 ] demonstrated a strong dependence of \alpha on atmospheric conditions such as temperature and relative humidity, reporting values ranging from 0.93\text{\times}{10}^{-12}1.5\text{\times}{10}^{-12}\text{\,}{\mathrm{m}}^{3}\text{\,}{\mathrm{s}}^{-1} for relative humidities between 070\text{\,}\mathrm{\%} . Consequently, the large uncertainties associated with ion mobility and ion–ion recombination coefficients constitute one of the dominant sources of uncertainty in the physical parameters of the present simulation. 3 Methods In this study, Garfield++ was used to simulate PPICs and model the physical processes occurring under UHDR irradiation conditions. To achieve this, several existing Garfield++ classes and methods were extended to account for space-charge effects and ion–ion recombination. 3.1 Garfield++ simulation toolkit Garfield++ [ 26 ] is an object-oriented simulation toolkit dedicated to the study of charged particle transport and interactions in gases and semiconductors. Originally developed by Rob Veenhof and now actively maintained at CERN, Garfield++ combines Monte Carlo transport techniques with electric and magnetic field maps to simulate charged particle transport in detector systems. The framework interfaces with several external packages, including ROOT [ 8 ] for visualisation, Magboltz [ 2 ] for the calculation of electron transport properties in gas mixtures, and Heed [ 30 ] for the simulation of primary ionisation produced by charged particles. A typical Garfield++ simulation follows the structure

Read the complete text