Ablation Removal of Transport-Blocking Defects in Surface-Electrode Ion Traps Preprint: AIP/123-QED

T. Maddock, P. Rahimi, M. Aylett, R. Barcan, S. Weidt, W.K. Hensinger · 2026

Read Ablation Removal of Transport-Blocking Defects in Surface-Electrode Ion Traps Preprint: AIP/123-QED by T. Maddock, P. Rahimi, M. Aylett, R. Barcan, S. Weidt, W.K. Hensinger (2026) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/arxiv-2605-06312.txt.

Introduction

Toby Maddock and five co-authors remove a transport blocking defect from an ion trap without opening the vacuum system. Surface electrode ion traps shuttle ions along paths for quantum computing, and a single defect can block a route entirely. The usual fix means venting and rebaking the apparatus, a slow process that costs heavy downtime, above all for cold systems that need long bakes after any opening.

Their remedy fires a pulsed ablation laser, of a type many ion trap labs already own, at the defect in place. After ablation, shuttling across the once blocked region succeeded at near unity rates. Micromotion, a key quality measure for trapped ions, stayed within acceptable limits. The technique restores transport pathways quickly and suits shuttling focused experiments well.

Catalog record

CREATORT. Maddock, P. Rahimi, M. Aylett, R. Barcan, S. Weidt, W.K. Hensinger
YEAR2026
COLLECTIONPAPERS
FORMATTXT
FILE SIZE18 KB
LICENSEhttp://creativecommons.org/licenses/by/4.0/
RIGHTSCC BY
SOURCEarXiv

Source: https://arxiv.org/abs/2605.06312

Browse the full library · About the Tower of Babel

Full text

The opening of the work, from the library's hosted text:

Ablation Removal of Transport-Blocking Defects in Surface-Electrode Ion Traps Preprint: AIP/123-QED T. Maddock, P. Rahimi, M. Aylett, R. Barcan, S. Weidt, W.K. Hensinger Abstract We demonstrate in situ removal of a transport-blocking defect on a surface-electrode ion trap device using a Q-switched Nd:YAG 532 nm pulsed ablation laser. This approach eliminates the need to vent and rebake the vacuum system, providing a low-overhead defect-remediation technique well suited for ion-shuttling architectures where system modifications typically incur substantial downtime - particularly in shuttling focussed experiments operating at temperatures that necessitate bakes. Additionally, the hardware used is readily available in many ion trap laboratories, making this solution attractive to experiments operating in such regimes. Following ablation, we observe near-unit shuttling success rates across the previously obstructed region and measure micromotion levels that remain within acceptable limits. This technique enables rapid, reliable restoration of transport pathways without interruption to experimental operation. Reliable ion transport is an essential requirement for many proposed architectures for trapped ion quantum information processing. Surface-electrode traps increasingly incorporate complex junctions, long shuttling pathways, and multi-zone architectures to move ions between memory, interaction, and detection regions 13 ; 16 ; 17 ; 26 ; 14 ; 22 ; 8 ; 20 . In such architectures, the integrity of transport links directly impacts computational fidelity, system reliability and information throughput: high-fidelity shuttling enables fast algorithmic execution 5 , multiplexed operations 4 , and modular integration across separate trap chips or cryogenic stages 1 ; 19 . Even a single obstruction can halt all downstream operations, making uninterrupted transport channels essential for large-scale designs. Surface-trap defects can arise through multiple mechanisms; most notable are particulate contamination due to neutral atom flux exposed to the chip during ion loading 29 ; 6 ; 12 , debris from electrode edges, and redeposition from laser-driven photoionization or ablation sources 21 ; 23 . Debris 10s of micrometres in size can introduce strong local perturbations to the pseudopotential, generating excess micromotion, induce static offsets via beam charge up and completely block transport through narrow junctions. Typical remediation of such problems introduce long experimental downtime, owing to long bakeout times, which becomes increasingly impractical as experiments scale into multi-module or cryogenic architectures 27 ; 18 . Minimising vent–bake cycles is therefore a major engineering priority. These constraints have motivated interest in in situ laser-cleaning techniques. Prior work has demonstrated ultraviolet pulsed-laser cleaning to remove contaminants and reduce anomalous heating 2 ; 12 , continuous-wave laser cleaning to modify adsorbates 10 , and laser desorption methods for electrode surface recovery 9 . While these studies show that targeted laser illumination can modify electrode surfaces without breaking vacuum, we note that these techniques have been used to remove adsorbed atomic and molecular layers from surfaces. By contrast, far less attention has been devoted to the removal of large, discrete, transport-blocking particulates, despite their severe operational consequences. This gap motivates the development of fast, reliable, in situ laser-based methods capable of restoring blocked transport pathways without interrupting system operation. Figure 1: a) Side-view image of the transport-blocking defect on the ion trap chip. The inset shows a magnified view of the defect region. b) EMCCD pixel data corresponding to a region of interest (ROI) centred on the defect as a guide laser illuminates it, and scanned perpendicular to the chip surface. The measurement is consistent with height estimates based on high resolution images, indicating a defect height of approximately 65 µm. In this study, a two-module quantum processor experiment was impeded by a transport blocking defect. The location proved to be particularly egregious due to its position between the gate zone of one of the chip modules and the second module, disallowing shuttling thermometry measurements. Highlighted in Fig. 1 , analysis of high-resolution images and reconstruction using a scanned guide laser allowed us to estimate the dimensions of the defect at approximately 65\,\mu\mathrm{m}\times 40\,\mu\mathrm{m} , confirming that it was large enough to block ion transport and detected during routing electron microscope detection, which implies that the origin of the contaminant was following the bake out procedure. To address this, we implemented an in situ removal technique using a Yb-doped 1.5 ns pulsed laser, carefully aligned and focused to the defect. This approach avoids venting the vacuum system or performing a full bakeout, eliminating the long downtime and alignment risks associated with conventional cleaning methods. The method enabled complete restoration of transport through the previously blocked region while minimizing the risk of collateral damage to nearby electrodes. The ion trap assembly is comprised of two spatially disparate surface ion trap modules 1 , cryogenically cooled to typical operating temperatures of 47 - 50 K. Ablation is performed using a 532nm, 1.5ns pulsed laser, with a maximum available pulse energy of 2mJ. As indicated in Fig. 2 , precise alignment to the microchip prior to ablation pulses necessitated the use of a 532nm continuous wave guide laser, with a comparatively lower peak fluence. Figure 2: The beam path taken by both the guide laser and the ablation beam. The telescope serves to guarantee precise overlap of the ablation and guide beam, ensure sufficient beam quality to minimise risk to other areas of the chip, and to expand the beam prior to the final lens Figure 3: Schematic diagram of the UHV system housing the surface ion trap, together with the imaging and laser ablation setup. The combined guide and ablation laser beams are delivered through a side window into the UHV chamber. Both beams operate at the same wavelength (532 nm) to enable reliable alignment and to avoid chromatic aberrations that could lead to systematic misalignment. Beam alignment is performed in the XZ plane using a micrometer translation stage, while focusing is achieved with a lens of focal length F=150 . Figure 4: Images of the ion trap chip acquired through the side windows a)-c) and the top windows d)-f) of the vacuum chamber. Panels a) and d) show an overview of the chip region containing the defect, while panels b), c), e), and f) provide magnified views of the defect location and the chip surface after defect removal. Electrodes are numbered sequentially starting from the central gap. The defect is located at the edge of electrode 7. The electrode width is indicated for scale. The ablation and guide lasers are combined using a polarising beam splitter, both of which are then passed through a beam expanding telescope with a 50um pin hole to ensure both beams are sufficiently Gaussian. Beam expansion ensured a small spot size following the final alignment to the defect. Both beams are carefully overlapped prior to precise chip alignment. The combined beams are then steered upwards to a motorised translation stage wherein beams are focussed to the defect using an adjustable 150mm lens. The motorised stage enables \mu m steps in vertical translation, allowing for sufficient precision for pulse targetting, given the 65 \mu m approximate height of the defect. To reduce risk to neighboring electrodes, and electrode wirebonds, the beam was expanded to a large beam waist of approximately 0.856mm; the lens prior to the vacuum chamber was adjusted to focus the beam to exactly the defect’s location, such that the beam’s fluence rapidly declined at all

Read the complete text