Reliability of uGMRT Band-4 Polarimetry: Results from a Quadrature Hybrid Polarizer Bypass Experiment
\sn Arpan \gn Pal 1,2 and \sn Sanjeet \gn Rai 3 and \sn Ganesh \gn Kumbhar 3 · 2026
Read Reliability of uGMRT Band-4 Polarimetry: Results from a Quadrature Hybrid Polarizer Bypass Experiment by \sn Arpan \gn Pal 1,2 and \sn Sanjeet \gn Rai 3 and \sn Ganesh \gn Kumbhar 3 (2026) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/arxiv-2604-06399.txt.
Introduction
Pal, Rai, and Kumbhar report a calibration instability in new Giant Metrewave Radio Telescope Band 4. Tests showed cross-hand phase varies by source, breaking calibration sharing. Checks traced the fault to the Quadrature Hybrid polarizer. Skipping it in seven dishes cut leakage from 10 to 15 percent to 2 to 5 percent, and leakage to under a fifth of a percent, calming phases. For DA 240, the fixed system recovered the expected 25 degree rotation, which the polarizer system failed to match. The results blame the Quadrature Hybrid polarizer and support linear feeds.
Catalog record
| CREATOR | \sn Arpan \gn Pal 1,2 and \sn Sanjeet \gn Rai 3 and \sn Ganesh \gn Kumbhar 3 |
|---|---|
| YEAR | 2026 |
| COLLECTION | PAPERS |
| FORMAT | TXT |
| FILE SIZE | 47623 |
| LICENSE | http://creativecommons.org/licenses/by/4.0/ |
| RIGHTS | CC BY |
| SOURCE | arXiv |
Source: https://arxiv.org/abs/2604.06399
Browse the full library · About the Tower of Babel
Full text
The opening of the work, from the library's hosted text:
Reliability of uGMRT Band-4 Polarimetry: Results from a Quadrature Hybrid Polarizer Bypass Experiment \sn Arpan \gn Pal 1,2 and \sn Sanjeet \gn Rai 3 and \sn Ganesh \gn Kumbhar 3 Abstract Polarimetric observations at sub-GHz frequencies offer unique access to the magnetized universe through Faraday rotation and depolarization studies, but achieving reliable polarization calibration at these frequencies remains challenging. We report the identification and resolution of a systematic polarization calibration instability in the upgraded Giant Metrewave Radio Telescope (uGMRT) Band 4 (550–750 MHz). Through diagnostic observations of multiple calibrators, we discovered that the cross-hand phase response varies with the fractional polarization of the observed source, violating the fundamental assumption of calibration transferability in radio interferometry. Systematic engineering tests traced this behaviour to the Quadrature Hybrid (QH) polarizer in the frontend signal chain. We conducted a controlled experiment in which the QH was bypassed in seven antennas, converting them to linear polarization feeds. The bypassed system shows dramatically improved performance: instrumental leakage reduced from 10–15% to 2–5%, residual leakage after calibration reduced from ∼ \sim 0.5% to less than 0.2 % 0.2\% , and stable cross-hand phases independent of source polarization. For the polarized source DA 240 (RM = = 3.3 rad m -2 ), the QH-bypassed system accurately recovers the expected 25 ∘ 25^{\circ} polarization angle rotation across the band, which the with QH system fails to reproduce. These results establish that the QH polarizer is the dominant source of polarimetric instability in uGMRT Band 4 and demonstrate that its removal enables reliable sub-GHz polarimetry. We recommend the linear feed configuration for science cases requiring accurate polarization angle and rotation measure measurements. \lefttitle Reliable Polarimetry with uGMRT Band 4 \righttitle Pal et al. \jnlPage 120 \jnlDoiYr 2026 \doival 10.1017/pasa.xxxx.xx \articletitt Research Paper \corresp A. Pal, Email: [email protected] \citeauth Pal A, Rai S and Kumbhar G, Reliability of uGMRT Band-4 Polarimetry: Results from a Quadrature Hybrid Polarizer Bypass Experiment. Publications of the Astronomical Society of Australia 00 , 1–20 \history (Received xx xx xxxx; revised xx xx xxxx; accepted xx xx xxxx) keywords Low-frequency Radio Polarimetry, Polarization Calibration, Radio Interferometry 1 Introduction Radio telescopes measure the properties of the incoming electromagnetic radiation using feeds sensitive to two orthogonal polarization states. This makes radio interferometers inherently suitable for polarimetric observations, by measuring the full Stokes parameters ( I , Q , U , V ) that completely characterize the polarization state of the radiation ( 21 ) . The response of a radio telescope to incoming radiation can be elegantly described using the Jones matrix formalism ( 12 ; 17 ; 18 ) . The electric field of an incoming electromagnetic wave can be represented as a two-component complex vector \mathbf{e} . The measured voltage \mathbf{v} at the output of a feed is related to the true sky electric field through a 2\times 2 complex Jones matrix \mathbf{J} : \mathbf{v}=\mathbf{J}\mathbf{e} (1) The total Jones matrix of an antenna can be decomposed into a product of matrices, each representing a distinct instrumental or propagation effect. For an interferometer baseline formed between antennas i and j , the measured visibility matrix \mathbf{V}_{ij}^{\mathrm{obs}} is related to the true sky coherency matrix \mathbf{V}_{ij}^{\mathrm{sky}} by: \mathbf{V}_{ij}^{\mathrm{obs}}=\mathbf{J}_{i}\mathbf{V}_{ij}^{\mathrm{sky}}\mathbf{J}_{j}^{\dagger} (2) where the dagger denotes the conjugate transpose. The calibration of a radio interferometer involves solving these Jones matrices that describe the instrumental response by observing appropriate calibrator sources. The standard calibration chain includes residual delay calibration, time and frequency dependent gain corrections using known calibrator sources. These steps primarily address the parallel-hand correlations and assume that the Jones matrix is approximately diagonal, which is generally sufficient for total intensity observations. For full polarization calibration, additional steps are required to determine the off-diagonal elements of the Jones matrix. These steps include estimating the instrumental leakage terms by observing an unpolarized source, or a source with known polarization properties, or a polarized source over a wide range of parallactic angles ( 12 ; 17 ; 16 ) . Another step corrects the differential delay between the two polarization channels, which introduces a linear phase slope across frequency in the cross-hand correlations ( 17 ; 16 ) . This delay is measured using a polarized calibrator. The final step calibrates the absolute polarization angle using a source with a well-determined polarization orientation ( 17 ; 16 ) . The calibration procedure relies on the principle that the instrumental response remains stable between calibrator and target observations or varies smoothly enough to be reliably interpolated. The upgraded Giant Metrewave Radio Telescope (uGMRT) is a 30-element interferometer array with 45-meter dishes, operating across four frequency bands from 120 MHz to 1460 MHz ( 11 ) . Band 4 of the uGMRT covers from 550 to 950 MHz. uGMRT is the most sensitive radio interferometer in the Northern hemisphere to be operated at these frequencies. Sub-GHz polarimetry provides unique insights into Faraday rotation ( 5 ; 3 ) , depolarization mechanisms ( 19 ) , and the large-scale magnetic Universe ( 1 ) . Nevertheless, achieving reliable polarization calibration at these frequencies has long been a significant challenge. During regular polarimetric observations with uGMRT Band 4, we encountered a systematic issue that hindered accurate polarization calibration. The calibrated data exhibited persistent phase ramps in the cross-hand visibilities, which could not be corrected using standard polarization calibration procedures, including leakage correction, cross-hand delay estimation, and polarization angle calibration. The cross-hand phase (RL/LR) displayed a non-linear response, and residual phase ramps persisted regardless of the choice of calibrator sources. An engineering experiment by 14 revealed that the uGMRT signal chain introduces a variable cross-phase response depending on the fractional polarization. This behavior undermines the usual assumption that calibration can be transferred between sources, leading to significant inaccuracies in the polarization calibration. These instabilities are particularly problematic because they mimic astrophysical effects such as Faraday rotation and depolarization, making it difficult to disentangle true source polarization from instrumental artifacts. In this paper, we present a systematic investigation of this polarization calibration problem and demonstrate that the Quadrature Hybrid (QH) polarizer in the uGMRT Band 4 signal chain is the primary source of the instability. We describe a controlled experiment in which the QH polarizer was bypassed in a subset of antennas, converting them from circular to linear polarization feeds, and show that this modification results in stable cross-hand phases, dramatically reduced leakage amplitudes, and accurate recovery of polarization angles and rotation measures. The paper is structured as follows. Section 2 presents the architecture of the uGMRT signal chain. Section 3 discusses the polarization calibration challenges encountered with the standard system and details the engineering investigations that identified the QH polarizer as the source of the problem. Section 4 describes a controlled astronomical experiment conducted with bypassed polarizers. Section 5 outlines the data analysis methodology applied to both circular and linear feed configurations. Section 6 reports the comparative results obtained from the two systems. Section 7 presents a recent, simplified noise-source experiment comparing the same systems. Finally, Section 8 summarizes our conclusions and provides recommendations for achieving reliable sub-GHz polarimetry with the uGMRT. Figure 1 : Block diagram of the uGMRT dual-channel RF signal chain for a single antenna. CH1 and CH2 are the polarization inputs from the feed. The