Automated Robotic Needle Puncture for Percutaneous Dilatational Tracheostomy

Yuan Tang, Bruno V. Adorno, Brendan A. McGrath, Andrew Weightman · 2026

Read Automated Robotic Needle Puncture for Percutaneous Dilatational Tracheostomy by Yuan Tang, Bruno V. Adorno, Brendan A. McGrath, Andrew Weightman (2026) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/arxiv-2602-22952.txt.

Description

Yuan Tang and three co-authors take over the riskiest step of bedside tracheostomy. The windpipe procedure serves ICU patients on long breathing help, and its needle puncture can cause major bleeding or back wall piercing when off-target. Manual puncture without guidance today errs by about five mm and thirty degrees.

A speed-ruled robotic arm now drives needle entry, guided by two magnet-based sensors, one at the needle tip and one inside the trachea. A self-tuning bounded driver tunes shaky motion values live while dodging bumps with body and nearby tissues. Tests checked the driver, then four hundred trials on a dummy showed median place error under two mm and midline drift about four degrees, hinting at feasible robotic puncture.

Catalog record

CREATORYuan Tang, Bruno V. Adorno, Brendan A. McGrath, Andrew Weightman
YEAR2026
COLLECTIONPAPERS
FORMATTXT
FILE SIZE56 KB
VERSION1.0
LICENSEhttp://creativecommons.org/licenses/by/4.0/
RIGHTSCC BY
SOURCEarXiv

Source: https://arxiv.org/abs/2602-22952

Browse the full library · About the Tower of Babel