Human Cardiac Measurements with Diamond Magnetometers

Muhib Omar, Magnus Benke, Shaowen Zhang, Jixing Zhang, Michael Kuebler, Pouya Sharbati, Ara Rahimpour, Arno Gück, Maryna Kapitonova, Devyani Kadam, Carlos René Izquierdo Geiser, Jens Haller, Arno Trautmann, Katharina Jag-Lauber, Robert Rölver, Thanh-Duc Nguyen, Leonardo Gizzi, Michelle Schweizer, Me · 2026

Read Human Cardiac Measurements with Diamond Magnetometers by Muhib Omar, Magnus Benke, Shaowen Zhang, Jixing Zhang, Michael Kuebler, Pouya Sharbati, Ara Rahimpour, Arno Gück, Maryna Kapitonova, Devyani Kadam, Carlos René Izquierdo Geiser, Jens Haller, Arno Trautmann, Katharina Jag-Lauber, Robert Rölver, Thanh-Duc Nguyen, Leonardo Gizzi, Michelle Schweizer, Me (2026) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/arxiv-2601-18843.txt.

Introduction

Muhib Omar and many colleagues read the heart's faint field with tools made from flaws in a gem. The gem is diamond, and its flaws can feel a field with great care. The read is done with no touch and no cut: the tool sits by the chest and picks up the heart's beat as a field. Such a read could one day sit by a bed or in a small ward.

The team ran the read in three forms, in rooms with shields from stray fields and in rooms with none. Each trace took a mean over a huge count of beats to pull the faint mark from the hiss. Clear traces in plain rooms would be a big step: shield rooms are dear and rare. The work shows the gem tools can catch the heart's field in the wild of a real ward.

Catalog record

CREATORMuhib Omar, Magnus Benke, Shaowen Zhang, Jixing Zhang, Michael Kuebler, Pouya Sharbati, Ara Rahimpour, Arno Gück, Maryna Kapitonova, Devyani Kadam, Carlos René Izquierdo Geiser, Jens Haller, Arno Trautmann, Katharina Jag-Lauber, Robert Rölver, Thanh-Duc Nguyen, Leonardo Gizzi, Michelle Schweizer, Me
YEAR2026
COLLECTIONPAPERS
FORMATTXT
FILE SIZE35 KB
VERSION1.0
LICENSEhttp://creativecommons.org/licenses/by/4.0/
RIGHTSCC BY
SOURCEarXiv

Source: https://arxiv.org/abs/2601-18843

Browse the full library · About the Tower of Babel

Full text

The opening of the work, from the library's hosted text:

Human Cardiac Measurements with Diamond Magnetometers Muhib Omar, Magnus Benke, Shaowen Zhang, Jixing Zhang, Michael Kuebler, Pouya Sharbati, Ara Rahimpour, Arno Gück, Maryna Kapitonova, Devyani Kadam, Carlos René Izquierdo Geiser, Jens Haller, Arno Trautmann, Katharina Jag-Lauber, Robert Rölver, Thanh-Duc Nguyen, Leonardo Gizzi, Michelle Schweizer, Mena Abdelsayed, Ingo Wickenbrock, Andrew M. Edmonds, Matthew Markham, Peter A. Koss, Oliver Schnell, Ulrich G. Hofmann, Tonio Ball, Jürgen Beck, Dmitry Budker, Joerg Wrachtrup, Arne Wickenbrock Abstract We demonstrate direct, non-invasive and non-contact detection of human cardiac magnetic signals using quantum sensors based on nitrogen-vacancy (NV) centers in diamond. Three configurations were employed recording magnetocardiography (MCG) signals in various shielded and unshielded environments. The signals were averaged over a few hundreds up to several thousands of heart beats to detect the MCG traces. The compact room-temperature NV sensors exhibit sensitivities of 6–26 pT/ Hz \sqrt{\text{Hz}} with active sensing volumes below 0.5 mm 3 \text{mm}^{3} , defining the performance level of the demonstrated MCG measurements. While the present signals are obtained by averaging, this performance already indicates a clear path toward single-shot MCG sensing. To move beyond shielded environments toward practical clinical use, strong noise suppression is required. To this end, we implement NV-based gradiometry and achieve efficient common-mode noise rejection, enabled by the intrinsically small sensing volume of NV sensors. Together, these multi-platform results obtained across diverse magnetic environments provide a solid foundation for translating quantum sensors into human medical diagnostics such as MCG and magnetoencephalography (MEG). 1 Introduction Weak magnetic fields generated by neuronal and muscular activity, especially cardiac magnetic fields, carry rich physiological information. Magnetocardiography (MCG) can be used to diagnose conditions such as arrhythmias and coronary artery disease 23 ; 10 . Electrocardiography (ECG) and MCG are both based on cardiac electrical activity, with the former recording electric potential differences and the latter recording magnetic fields, reflecting the same physiological processes of depolarization and repolarization. ECG signals originated from action potentials in the heart and can be recorded on the body surface through electrodes. This technique is mature, low-cost, and widely available, making it the clinical gold standard for routine diagnosis. However, ECG signals are strongly influenced by the conductivity of body tissues, resulting in limited spatial resolution while the necessity to contact the skin precludes application when such contact is not possible (e.g. burn victims). In contrast, MCG signals arise from the magnetic fields generated by cardiac currents and can be detected by highly sensitive magnetometers. As a non-contact technique, MCG signals are minimally affected by tissue conductivity, enabling the detection of subtle abnormalities that ECG may miss and supporting precise localization of pathological sources 23 . Historically, MCG relied on SQUIDs and later on optically-pumped magnetometers based on vapor cells (OPMs). Other magnetometers, such as fluxgate sensors and magnetoresistive sensors, have also been demonstrated to detect MCG signals 29 ; 17 . Although MCG offers high clinical potential, the cardiac magnetic signals are inherently weak and the measurements are highly susceptible to environmental magnetic noise, necessitating stringent magnetic shielding. In addition, the high cost of equipment and the environmental dependence of present-day sensors have limited its widespread clinical adoption. Additionally, the ECG and MCG techniques are complementary (see Table 1 ), and their integration may substantially improve early diagnosis and precise localization of cardiac diseases. Current MCG technologies are predominantly based on superconducting quantum interference devices (SQUIDs), see for example 21 , which require cryogenic cooling and thus hinder widespread clinical deployment. In unshielded environments OPMs have been implemented as an alternative 41 . But following this approach requires facing the challenges due to heading errors, dead zones, and high operating temperatures that complicate contact-based measurements 18 ; 38 . Therefore, development of miniaturized, room-temperature MCG technologies capable of providing full vector information in unshielded environments remains desirable. One such pathway is based on nitrogen-vacancy (NV) centers in diamond. During the past two decades, NV sensors have been widely used to measure various physical quantities, including temperature 32 , strain 20 , electric fields 8 , magnetic fields 6 , and rotation 33 . Compared with OPMs, NV magnetometers feature a much smaller sensing volume, simplifying gradiometer based, vector-resolved MCG measurements under unshielded conditions. Their comparatively high volume sensitivity 5 makes them credible competitors to OPMs when combined with flux concentrators 11 . The comparably lower sensitivity may then be an advantage, since the intrinsic noise of the flux concentrator might not become a limiting factor 14 . They also offer fast initialization, excellent biocompatibility, and stable operation over a broad temperature range showing great potential for biomedical sensing. NV magnetometers have been demonstrated in animal studies for both invasive 3 and non-invasive MCG detection using flux concentrators 43 . In this study, we detected human cardiac magnetic signals using NV magnetometers independently developed by three groups, Johannes Gutenberg-University Mainz (JGU), University of Stuttgart (ZAQuant) and the quantum technology start-up Q.ANT GmbH (Q.ANT). These results represent a critical step toward the real-world biomedical application of quantum magnetometers and demonstrate that NV diamond magnetometry has now reached a stage where meaningful biomagnetic signals can be measured under various experimental conditions. This progress allows to quantify the needed improvement in signal-to-noise-ratio (SNR) and positions NV sensing as being competitive with existing technologies while promising to offer improved accessibility and practicality for both clinical and home use. 2 Results This section describes the three integrated NV magnetometers and the corresponding MCG measurement procedures, covering both zero-bias and finite bias-field configurations operated in shielded, partially shielded, and unshielded environments. It further reports a cross-validation of the NV-based MCG signals using a commercial OPM sensor array, and concludes with gradiometric noise-suppression results that demonstrate the feasibility of NV magnetometry in realistic, noisy, and unshielded settings. 2.1 MCG with NV-based magnetometers Figure 1 : Column (a) JGU: a.1): Diamond sensor head dimensions. a.2): Amplitude spectral density with a baseline of 13 pT/ \sqrt{\mathrm{Hz}} , measured inside the shielded room of the Helmholtz Institute Mainz (JGU). a.3): ECG trace and diamond MCG signal (12000 averages) with indications of the main components of the ECG, P wave, QRS complex and T wave. Column (b) ZAQuant: b.1): Sensor schematic. b.2): Measured amplitude spectral densities inside the partially shielded environment, showing a noise baseline of 7 pT/ \sqrt{\mathrm{Hz}} . b.3): ECG trace and averaged diamond MCG signal (300 averages). Column (c) Q.ANT: a.1): Sensor design. a.2): Amplitude spectral density with a baseline of 26 pT/ \sqrt{\mathrm{Hz}} ; filters were applied during acquisition. 3̇): ECG trace and diamond MCG signal (2000 averages). We first describe the JGU MCG measurements conducted at the Helmholtz Institute Mainz within its multilayer magnetically shielded room using the zero-bias-field NV magnetometer. Afterwards, the MCG measurements realized with bias-field NV magnetometers at the University of Stuttgart and at Q.ANT GmbH are introduced. Schematics of the sensors, their noise characteristics and the MCG measurements for all sensors can be seen in Fig. 1 . To record human cardiac magnetic signals in Mainz, we employed a fiberized NV diamond magnetometer designed for the use as a transportable endoscope. The sensor is operated without a magnetic bias field. The sensing protocol employing an oscillating magnetic field to recover magnetic sensitivity without a dc bias magnet is described in 45 ;

Read the complete text