Feasibility study of the positronium lifetime imaging with the Biograph Vision Quadra and J-PET tomographs
SPSzymon Parzych, SNSzymon Niedźwiecki, EYBErmias Yitayew Beyene, NCNeha Chug, MCMaurizio Conti, CCCatalina Curceanu, ECEryk Czerwiński, MDManish Das, KVEKavya Valsan Eliyan, JHJakub Hajduga, SJSharareh Jalali, KKKrzysztof Kacprzak, TKTevfik Kaplanoglu, ŁKŁukasz Kapłon, KKKamila Kasperska, AKAleksan · 2026
Read Feasibility study of the positronium lifetime imaging with the Biograph Vision Quadra and J-PET tomographs by SPSzymon Parzych, SNSzymon Niedźwiecki, EYBErmias Yitayew Beyene, NCNeha Chug, MCMaurizio Conti, CCCatalina Curceanu, ECEryk Czerwiński, MDManish Das, KVEKavya Valsan Eliyan, JHJakub Hajduga, SJSharareh Jalali, KKKrzysztof Kacprzak, TKTevfik Kaplanoglu, ŁKŁukasz Kapłon, KKKamila Kasperska, AKAleksan (2026) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/arxiv-2601-03172.txt.
Introduction
Szymon Parzych and forty-one co-authors test, in simulation, whether two PET scanners can perform positronium lifetime imaging, a proposed new diagnostic technique. Positronium, an electron and positron bound together, has a lifetime that may serve as a biomarker. But imaging it needs very sensitive scanners. They must register annihilation photons and a prompt gamma at once. The study compares the long axial field-of-view Biograph Vision Quadra with the total-body J-PET scanner.
Using GATE simulation software, with background radiation added for the Quadra, the team first finds the best energy window for the prompt gamma. They then calculate organ-wise sensitivity for four radioisotopes and compare scandium sensitivity across the Quadra, the total-body J-PET, and a modular J-PET prototype. The results map out which hardware and isotope combinations make the technique feasible.
Catalog record
| CREATOR | SPSzymon Parzych, SNSzymon Niedźwiecki, EYBErmias Yitayew Beyene, NCNeha Chug, MCMaurizio Conti, CCCatalina Curceanu, ECEryk Czerwiński, MDManish Das, KVEKavya Valsan Eliyan, JHJakub Hajduga, SJSharareh Jalali, KKKrzysztof Kacprzak, TKTevfik Kaplanoglu, ŁKŁukasz Kapłon, KKKamila Kasperska, AKAleksan |
|---|---|
| YEAR | 2026 |
| COLLECTION | PAPERS |
| FORMAT | TXT |
| FILE SIZE | 38 KB |
| VERSION | 1.0 |
| LICENSE | http://creativecommons.org/licenses/by/4.0/ |
| RIGHTS | CC BY |
| SOURCE | arXiv |
Source: https://arxiv.org/abs/2601-03172
Browse the full library · About the Tower of Babel
Full text
The opening of the work, from the library's hosted text:
Feasibility study of the positronium lifetime imaging with the Biograph Vision Quadra and J-PET tomographs SPSzymon Parzych, SNSzymon Niedźwiecki, EYBErmias Yitayew Beyene, NCNeha Chug, MCMaurizio Conti, CCCatalina Curceanu, ECEryk Czerwiński, MDManish Das, KVEKavya Valsan Eliyan, JHJakub Hajduga, SJSharareh Jalali, KKKrzysztof Kacprzak, TKTevfik Kaplanoglu, ŁKŁukasz Kapłon, KKKamila Kasperska, AKAleksander Khreptak, GKGrzegorz Korcyl, TKTomasz Kozik, DKDeepak Kumar, AKVAnoop Kunimmal Venadan, KKKarol Kubat, ELEdward Lisowski, FLFilip Lisowski, JMSJustyna Medrala-Sowa, WMWiktor Mryka, SMSimbarashe Moyo, PPPiyush Pandey, EPElena Perez del Rio, APAlessio Porcelli, BRBartłomiej Rachwał, MRMartin Rädler, ARAxel Rominger, SSSushil Sharma, KSKuangyu Shi, MSMagdalena Skurzok, WMSWilliam M. Steinberger, TSTomasz Szumlak, PTPooja Tanty, KTAKeyvan Tayefi Ardebili, STSatyam Tiwari, EŁSEwa Ł. Stepień, PMPaweł Moskal Abstract \parttitle Background After its first ex-vivo and in-vivo demonstration, Positronium Lifetime Imaging (PLI) has received considerable interest as a potential new diagnostic biomarker. High sensitivity Positron Emission Tomography (PET) systems are needed for PLI since it requires simultaneous registration of annihilation photons and prompt gamma. In this simulation-based study, a feasibility of PLI with the long axial field-of-view Biograph Vision Quadra (Quadra) and the Total Body J-PET scanner was investigated. \parttitle Methods The study was performed using the GATE software. Background radiation, present within the Quadra tomograph, was added to the simulation. First, the optimal placement of the energy window for the registration of the prompt gamma was investigated. Next, the organ-wise sensitivity of Quadra was calculated for the 68 Ga, 44 Sc, 22 Na and 124 I radioisotopes. Finally, the sensitivity for the scandium isotope was compared to the sensitivities obtainable with the Total Body J-PET scanner, as well as with the modular J-PET prototype. \parttitle Results The PLI sensitivities for the Quadra with the background radiation are estimated to 9.22(3), 10.46(4), 5.91(3), and 15.39(4) cps/kBq for the 44 Sc, 68 Ga, 22 Na and 124 I radioisotopes, respectively. The highest sensitivity was obtained when the energy window for the deexcitation photon is adjacent to the energy window for the annihilation photons. The determined PLI sensitivities with Quadra and the Total Body J-PET are in the order of sensitivities of standard PET imaging with the short axial field-of-view ( ∼ \sim 20 cm) PET scanners. \parttitle Conclusion The organ-wise PLI sensitivity of Quadra has been computed for the 68 Ga, 44 Sc, 22 Na and 124 I radioisotopes. A sensitivity gain by a factor of 150 was estimated relative to the modular J-PET system previously used for the first in-vivo PLI. keywords \startlocaldefs \endlocaldefs {fmbox} \dochead Research {artnotes} \note [id=n1]… {abstractbox} 1 Background The standard procedure of Positron Emission Tomography (PET) provides a metabolic image. Traditional Standardized Uptake Value (SUV) parameter quantifies the uptake of a given marker in tissues, enabling to identify regions of pathological metabolism [ 3 , 64 , 7 ] . Nevertheless, current PET systems cannot yet take advantage of measuring the lifetime of the metastable positronium atom formed in the intra-molecular voids and the information that may be obtained from it [ 76 , 2 ] . For this purpose, a new PET-based imaging method has been conceived [ 54 ] – Positronium Lifetime Imaging (PLI) – to provide complementary information about the imaged tissue. In particular, mean lifetime or formation probability [ 42 , 41 , 53 , 48 ] constitute additional and qualitatively different contrast mechanisms from SUV [ 44 ] . This novel type of imaging is recently gaining interest in both laboratory and hospital research [ 41 , 70 ] with its many challenges and new opportunities, e.g. analysis of positronium lifetime spectra [ 21 , 65 ] , PLI algorithms [ 61 , 66 , 16 , 24 , 26 , 25 ] , double tracer imaging [ 59 , 22 , 75 , 14 ] . Although the requirement of three photon coincidence reduces sensitivity significantly, it may be resolved by another current trend towards the construction of long axial field-of-view PET scanners [ 38 , 6 , 4 , 60 , 19 , 33 , 51 ] . The first applications of both ex-vivo and in-vivo PLI were demonstrated by the Jagiellonian-PET (J-PET) collaboration [ 54 , 41 , 40 ] . However, the first-ever clinical PLI image of the brain presented by the J-PET collaboration suffers from the low statistics due to the utilized radionuclide ( 68 Ga) and low scanner’s sensitivity [ 41 ] . Firstly, the pharmaceutical labeled with 68 Ga radionuclide has a low branching ratio (1.2% [ 12 ] ) for \beta^{+} decay followed by a deexcitation through the emission of a prompt gamma [ 20 ] . The administration of dedicated radiopharmaceuticals, such as 44 Sc radionuclide, can enhance statistics by about 76 times [ 51 , 67 , 37 ] . Secondly, the patient was imaged with the prototype of the J-PET tomograph - single ring only modular J-PET with 50 cm axial field-of-view (AFOV) constructed with organic scintillators [ 41 , 71 , 72 ] . Its sensitivity for metabolic and positronium imaging amounts to 3.4 and 0.06 cps/kBq [ 41 ] . Therefore, in this study we investigate a crystal-based system with higher interaction probability of photons. Most of the clinically certified PET tomographs are acquiring data in the form of sinograms or list-mode pairs, which limits raw data to only double coincidences. However, in order to perform positronium imaging, a specific type of triple coincidence is required. The type of acquisition, which allows for the registration of all interactions is referred to as triggerless mode [ 34 , 50 ] or singles mode [ 70 ] . Apart from the J-PET system, which works in the triggerless data acquisition mode, the PennPET Explorer in Philadelphia, USA [ 19 , 33 ] as well as the recently upgraded Biograph Vision Quadra (hereinafter referred to as Quadra) tomograph from Siemens Healthineers [ 70 ] have the possibility of singles mode acquisition. In this simulation-based study, a feasibility of positronium lifetime imaging with Quadra was investigated and compared to the J-PET scanners. We focus on PLI sensitivity, which has been identified as a major issue in the images previously presented by the J-PET collaboration and is expected to significantly improve with the crystal-based and Total-Body tomographs. Four isotopes were explored in this work: 68 Ga, 44 Sc, 22 Na and 124 I [ 51 , 40 , 67 , 37 , 36 , 74 ] . The decay scheme of these radionuclides are shown in Figure 1 . The isotope of gallium was chosen due to its common use in PET imaging. The scandium radioisotope is considered due to its feasibility for positronium imaging [ 20 ] . Despite the present lack of clinical imaging applications, the 22 Na isotope was chosen due to its exploitation in basic research [ 8 ] . Finally, the iodine isotope was chosen for its diagnosis and treatment capabilities of thyroid cancer [ 58 ] . [width=0.7]Isotopes.png Figure 1: Schematic of selected decay modes relevant to PET imaging for the four investigated radioisotopes: 44 Sc, 68 Ga, 22 Na and 124 I. Positron emission is denoted by e + , whereas EC indicates the contribution from electron capture [ 12 , 10 , 11 , 13 ] . 2 Methods 2.1 Positronium imaging The pharmaceuticals administered